Évaluation microbiologique de la désinfection des cystoscopes souples au CHRU de Brest de janvier 2007 à décembre 2014 - 28/11/15

Microbiological evaluation of cystoscope reprocessing at Brest university hospital from January 2007 through December 2014

Doi : 10.1016/j.purol.2015.10.012 
P. Saliou a, , H. Le Bars b, G. Fournier c, d, R. Baron a, b
a Équipe opérationnelle d’hygiène, service d’hygiène hospitalière, CHRU Morvan, 2, avenue Foch, 29200 Brest, France 
b Département de microbiologie, CHRU de Brest, 29200 Brest, France 
c Service de chirurgie urologique, CHRU de Brest, 29200 Brest, France 
d Université de Bretagne occidentale, 29200 Brest, France 

Auteur correspondant.
Sous presse. Épreuves corrigées par l'auteur. Disponible en ligne depuis le Saturday 28 November 2015
Cet article a été publié dans un numéro de la revue, cliquez ici pour y accéder

Résumé

But

Les cystoscopes souples sont des appareils thermosensibles non autoclavables qui doivent subir une désinfection de haut niveau selon la classification de Spaulding. L’objectif de cette étude est de fournir des données épidémiologiques relatives à la surveillance de la qualité microbiologique des cystoscopes utilisés dans un centre hospitalo-universitaire, de présenter le taux de conformité des contrôles microbiologiques et de lister les principaux micro-organismes identifiés.

Matériel

Étude prospective de l’ensemble des résultats des prélèvements microbiologiques réalisés pendant 8ans au CHRU de Brest. Les résultats des contrôles ont été interprétés selon les recommandations ministérielles.

Résultats

Au cours de l’étude, 87 contrôles microbiologiques ont été réalisés. Le taux de prélèvements non conformes était de 19,5 % (17/87). Ce taux atteignait 24,5 % (12/49) pour les contrôles programmés. Les micro-organismes identifiés étaient présents en faible quantité et correspondaient principalement à des bactéries issues de l’environnement.

Conclusion

Le taux de non-conformité des contrôles microbiologiques réalisés sur les cystoscopes est relativement élevé (19,5 %) mais le risque infectieux semble limité. En effet, les micro-organismes retrouvés sont présents en faible quantité et ne sont pas connus pour être responsables d’infections urinaires. La sensibilisation des professionnels chargés du traitement des cystoscopes et la réalisation d’audits réguliers devraient permettre d’assurer un bon niveau de désinfection des cystoscopes.

Niveau de preuve

3.

Le texte complet de cet article est disponible en PDF.

Summary

Purpose

Flexible cystoscopes are high temperature sensitive devices that must undergo high-level disinfection according to Spaulding classification. The objective of this study is to provide epidemiological data on the monitoring of microbiological quality of cystoscopes used in a teaching hospital, in order to determine the compliance rate of disinfection and to describe the main identified microorganisms.

Material

Prospective study of all the results of microbiological samples taken for 8 years at the Brest teaching hospital. The analysis results were interpreted according to the ministerial recommendations.

Results

During the study, 87 microbiological tests were performed. The rate of non-compliant samples was 19.5% (17/87). This rate reached 24.5% (12/49) of the programmed controls. The microorganisms identified were present in small amounts, corresponding mainly to bacteria from the environment.

Conclusion

The rate of non-compliance of the microbiological tests performed on cystoscopes is relatively high (19.5%), but the infectious risk seems limited. Hidden microorganisms are present in small quantities and identified germs are not known to be responsible for urinary tract infections. Educating professionals responsible for cystoscopes reprocessing and conducting regular audits may help to ensure a good level of cystoscope disinfection.

Level of evidence

3.

Le texte complet de cet article est disponible en PDF.

Mots clés : Cystoscopes souples, Désinfection, Surveillance microbiologique, Épidémiologie

Keywords : Flexible cystoscopes, Disinfection, Microbiological monitoring, Epidemiology


Plan


© 2015  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.