S'abonner

No effect of femoral offset on bone implant micromotion in an experimental model - 25/04/16

Doi : 10.1016/j.otsr.2016.01.010 
F. Amirouche , G. Solitro, A. Walia
 University of Illinois at Chicago, Department of Orthopaedics, 835, S. Wolcott avenue, Room E270, Chicago, IL 60612, USA 

Corresponding author. Tel.: +1 312 996 7161; fax: +1 312 996 9025.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Background

In total hip replacement (THR), the femoral offset (FO) is assessed preoperatively, and the surgeon must determine whether to restore, increase, or decrease the FO based on experience and the patient's clinical history. The FO is known to influence the abductor muscle strength, range of motion (ROM), gait, and hip pain after THR; however, the true effect of FO on bone implant micromotion is unclear. Therefore, we investigated to assess: (1) the muscle loading response during gait, (2) whether FO affects bone implant micromotion during gait.

Hypothesis

A variation of ±10mm from the anatomical FO affects the muscle loading forces.

Materials and methods

We modified a personalized musculoskeletal model of the lower extremity to determine the 3-dimensional contact forces at the hip joint in the presence of a stem with varying offsets during a gait cycle. A detailed finite element (FE) model was then constructed for increased, restored, and decreased FOs. The maximum load obtained during normal walking gait from the musculoskeletal model was applied to the respective FE models, and the resultant stem-bone micromotion and stress distribution were computed.

Results

Increasing the FO to +10mm decreased the peak force generated by the abductor muscles during the cycle by 15.0% and decreasing the FO to −10mm increased the von Mises stress distribution at the distal bone by 77.5% (P<0.05). A variation of the offset within 10mm of the anatomical offset showed no significant differences in micromotion (P>0.05) and peak stresses (P>0.05).

Discussion

Coupling the musculoskeletal model of the gait cycle with FE analysis provides a realistic model to understand how FO affects bone implant micromotion. We found that there was no effect of FO on bone implant micromotion; thus, a surgeon does not need to evaluate the implications of FO on micromotion and can determine a FO that best decreases the work load of abductor muscles, increases ROM, and reduces hip pain.

Level of evidence

IV, biomechanical study.

Le texte complet de cet article est disponible en PDF.

Keywords : Hip offset, Hip contact force, Micromotion, Finite element, Musculoskeletal model


Plan


© 2016  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 102 - N° 3

P. 379-385 - mai 2016 Retour au numéro
Article précédent Article précédent
  • Is determination of the hip rotation center during computer-assisted surgery influenced by the pinless femoral sensor attachment method? An in vitro preliminary study comparing osseous pin fixation
  • E. de Thomasson, S. Gouzy, F. Cloerec, L. Geais
| Article suivant Article suivant
  • Results of the Sugioka transtrochanteric rotational osteotomy for osteonecrosis: Frequency and role of a defect of the quadratus femoris muscle in osteonecrosis progression
  • T. Yamamoto, G. Motomura, K. Karasuyama, Y. Nakashima, T. Doi, Y. Iwamoto

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.