S'abonner

Comparaison des taux d’usure et d’ostéolyse des polyéthylènes hautement réticulés entre un recuit de seconde génération et un refondu de première génération : étude cas témoin au recul minimal de 5 ans - 28/03/17

Comparison of wear rate and osteolysis between second-generation annealed and first-generation remelted highly cross-linked polyethylene in total hip arthroplasty. A case control study at a minimum of five years

Doi : 10.1016/j.rcot.2017.03.004 
R. Takada, T. Jinno , D. Koga, K. Miyatake, T. Muneta, A. Okawa
 Department of Orthopaedic Surgery, Medical Hospital, Tokyo Medical and Dental University, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8519, Japon 

Auteur correspondant.
Sous presse. Épreuves corrigées par l'auteur. Disponible en ligne depuis le Tuesday 28 March 2017
Cet article a été publié dans un numéro de la revue, cliquez ici pour y accéder

Abstract

Background

There is no previous report that directly compared wear resistance of second-generation annealed highly cross-linked polyethylene with that of first-generation remelted highly cross-linked polyethylene. We therefore performed a retrospective study at a minimum of 5-year follow-up comparing second-generation annealed and first-generation remelted highly cross-linked polyethylene in order to: (1) assess wear rates and (2) compare the incidence of osteolysis between, (3) identify the frequency of complication related to the two types of highly cross-linked polyethylene.

Hypothesis

There is a difference in the linear wear rate and the incidence of osteolysis between the two types of highly cross-linked polyethylene in total hip arthroplasty.

Materials and methods

In a single center study, we reviewed 123 primary cementless total hip arthroplasties between 2010 and 2011 that were performed with 32mm alumina ceramic on second-generation annealed (X3) or first-generation remelted (longevity) highly cross-linked polyethylene liner. There was no specific reason for the choice of the type of highly cross-linked polyethylene. There were no significant differences between the two groups in respect of gender, diagnosis, body mass index, pre- and postoperative functional and activity score, cup size, and cup orientation except the younger age in the X3 group. The mean wear rate and the incidence of osteolysis were evaluated at the latest follow-up.

Results

One hundred and nine cases followed over 5 years postoperatively (88.6% in all consecutive cases) were evaluated. X3 and longevity were used in 54 and 55 cases, respectively. The mean follow-up was 5.3 years in both groups. The mean linear wear rate of X3 and longevity group was 0.045±0.023mm/year and 0.076±0.031mm/year, respectively (P<0.001). No osteolysis was found on plain X-rays in both groups and no specific complication was related to these highly cross-linked components.

Discussion

Excellent wear resistance of both types of highly cross-linked polyethylene liner was revealed in our study. The difference of wear rate between two materials should be monitored in a longer follow-up.

Level of evidence

Level III retrospective case control study.

Le texte complet de cet article est disponible en PDF.

Keywords : X3, Longevity, PolyWare, Tapered-wedge, Cementless, Cross-linked polyethylene, Wear



 Cet article peut être consulté in extenso dans la version anglaise de la revue Orthopaedics & Traumatology: Surgery & Research sur Science Direct (sciencedirect.com) en utilisant le DOI ci-dessus.


© 2017  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.