S'abonner

Can three-dimensional patient-specific cutting guides be used to achieve optimal correction for high tibial osteotomy? Pilot study - 30/03/17

Doi : 10.1016/j.otsr.2016.11.020 
M. Munier a, M. Donnez a, b, M. Ollivier a, X. Flecher a, P. Chabrand a, J.-N. Argenson a, S. Parratte a,
a UMR CNRS 787/AMU, institut du mouvement et de l’appareil locomoteur, groupe interdisciplinaire en biomécanique, hôpital Sainte-Marguerite, centre hospitalo-universitaire Sud, boulevard de Sainte-Marguerite, 13009 Marseille, France 
b Newclip Technics, 44115 Haute-Goulaine, France 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Introduction

Treatment of medial tibiofemoral osteoarthritis with a high-tibial osteotomy (HTO) is most effective when the optimal angular correction is achieved. However, conventional instrumentation is limited when multiplanar correction is needed.

Hypothesis

Use of patient-specific cutting guides (PSCGs) for HTO provides an accurate correction (difference<2°) relative to the preoperative planning.

Materials and methods

Between February 2014 and February 2015, 10 patients (mean age: 46 years [range: 31–59]; grade 1 or 2 osteoarthritis in Ahlbäck's classification) were included prospectively in this reliability and safety study. All patients were operated using the same medial opening-wedge osteotomy technique. Preoperative planning was based on long-leg radiographs and CT scans with 3D reconstruction. The PSGCs were used to align the osteotomy cut and position the screw holes for the plate. The desired correction was achieved in the three planes when the holes on the plate were aligned with the holes drilled based on the PSCG. Preoperatively, the mean HKA angle was 171.9° (range: 166–179°), the mean proximal tibial angle was 87° (86–88°) and the mean tibial slope was 7.8° (1–22°). The postoperative correction was compared to the planned correction using 3D CT scan transformations. Intraoperative and postoperative complications were assessed at a minimum follow-up of 1 year.

Results

The procedure was successfully carried out in all patients with the PSCGs. On postoperative long-leg radiographs, the mean HKA was 182.3° (180–185°); on the CT scan, the mean tibial mechanical angle was 94° (90–98°) and the mean tibial slope was 7.1° (4–11°). In 19 out of 20 postoperative HKA and slope measurements, the difference between the planned and achieved correction was <2° based on the 3D analysis of the three planes in space; in the other case, the slope was 13° instead of the planned 10°. The intra-class correlation coefficients between the postoperative and planned parameters were 0.98 [0.92–0.99] for the HKA and 0.96 [0.79–0.99] for the tibial slope. There were no surgical site infections; one patient had a postoperative hematoma that resolved spontaneously.

Discussion

The results of this study showed that use of PSCGs in HTO procedures helps to achieve optimal correction in a safe and reliable manner.

Level of evidence

IV – Prospective cohort study.

Le texte complet de cet article est disponible en PDF.

Keywords : Medial tibiofemoral osteoarthritis, High tibial osteotomy, Opening-wedge osteotomy, Patient-specific surgical guides, 3D printing


Plan


© 2017  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 103 - N° 2

P. 245-250 - avril 2017 Retour au numéro
Article précédent Article précédent
  • Comparison of postoperative biomechanical function between anatomic double-bundle and single-bundle ACL reconstructions using calcium phosphate-hybridized tendon grafts in goats
  • H. Mutsuzaki, H. Fujie, H. Nakajima, M. Fukagawa, S. Nomura, M. Sakane
| Article suivant Article suivant
  • Influence of coronal bowing on the lower alignment and the positioning of component in navigation and conventional total knee arthroplasty
  • H. Kobayashi, Y. Akamatsu, K. Kumagai, Y. Kusayama, M. Aratake, T. Saito

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.