S'abonner

Non-invasive cardiac output monitor validation study in pediatric cardiac surgery patients - 18/04/17

Doi : 10.1016/j.jclinane.2017.02.001 
Christopher F. Tirotta, MD, MBA a, , Richard G. Lagueruela, MD a, Danielle Madril, MD a, Evelio Velis, MD, PhD b, Jorge Ojito, CCP a, Daniel Monroe, RN a, David Aguero, RN a, Marysory Irizarry, RN a, John McBride, CVT a, Robert L. Hannan, MD a, Redmond P. Burke, MD a
a The Heart Program, Nicklaus Children's Hospital, 3100 SW 62nd Avenue, Miami, FL 33155, USA 
b College of Health Sciences, Barry University, 11300 NE Second Avenue, Office SNHS #229, Miami Shores, FL 33161-6695, USA 

Corresponding author.

Abstract

Study objective

To validate a non-invasive cardiac output monitor in pediatric cardiac surgery patients.

Design

Prospective trial.

Setting

Operating room.

Patients

20 pediatric cardiac surgery patients ASA physical status 3 and 4.

Interventions

Aesculon noninvasive cardiac output monitor was used and compared to the cardiac output derived from the Fick equation. Oxygen consumption was measured and blood samples were taken from the arterial line and from the superior and inferior vena cava.

Measurements

Noninvasive cardiac output and cardiac index and Fick cardiac output and cardiac index. Oxygen consumption was measured by the TreyMed metabolic monitor. Blood samples were simultaneously drawn from the arterial line and from the superior and inferior vena cava purse string sites by the surgeon, prior to commencing cardiopulmonary bypass. Another data set was obtained right after termination of cardiopulmonary bypass.

Results

There was a direct, significant relationship between Fick CO/CI and NICOM CO/CI measurements. More dispersion is detected when the magnitude of the measure increases, i.e., for older and larger patients.

Conclusions

There is a strong correlation between the cardiac output values derived from the Fick equation and the Aesculon non-invasive cardiac output monitor.

Le texte complet de cet article est disponible en PDF.

Highlights

Cardiac output can be a useful variable to clinicians caring for critically ill patients.
The Aesculon monitor can provide cardiac output non-invasively.
The Aesculon cardiac output is strongly correlated with the cardiac output derived from the Fick equation.

Le texte complet de cet article est disponible en PDF.

Keywords : Cardiac output, Non-invasive monitor, Cardiac surgery, Pediatrics, Congenital heart disease, Fick equation, Oxygen consumption


Plan


© 2017  Elsevier Inc. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 38

P. 129-132 - mai 2017 Retour au numéro
Article précédent Article précédent
  • ASA physical status assignment by non-anesthesia providers: Do surgeons consistently downgrade the ASA score preoperatively?
  • Christopher Curatolo, Andrew Goldberg, David Maerz, Hung-Mo Lin, Hardikkumar Shah, Muoi Trinh
| Article suivant Article suivant
  • Dexamethasone as a ropivacaine adjuvant for ultrasound-guided interscalene brachial plexus block: A randomized, double-blinded clinical trial
  • Thiago Mamôru Sakae, Patricia Marchioro, Fabiana Schuelter-Trevisol, Daisson José Trevisol

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.