S'abonner

Mechanisms of Günther Tulip filter tilting during transfemoral placement - 01/08/17

Doi : 10.1016/j.diii.2017.01.009 
Y. Matsui a, b, , M. Horikawa a, K. Ohta a, c, Y. Jahangiri Noudeh a, J.A. Kaufman a, K. Farsad a
a Dotter Interventional Institute, Oregon Health and Science University, 3181 SW Sam Jackson Park Rd, Portland, OR 97239, USA 
b Department of Radiology, Okayama University Medical School, 2-5-1 Sikata-cho, Kita-ku, Okayama 700-8558, Japan 
c Department of Radiology, Nagoya City University Graduate School of Medical Sciences, 1 Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya 467-8601, Japan 

Corresponding author. Department of Radiology, Okayama University Medical School, 2-5-1 Sikata-cho, Kita-ku, Okayama 700-8558, Japan. Tel.: +81-86-235-7313; fax: +81-86-235-7316.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Objective

The purpose of this study was to characterize the mechanisms of Günther Tulip filter (GTF) tilting during transfemoral placement in an experimental model with further validation in a clinical series.

Materials and methods

In an experimental study, 120 GTF placements in an inferior vena cava (IVC) model were performed using 6 configurations of pre-deployment filter position. The angle between the pre-deployment filter axis and IVC axis, and the proximity of the constrained filter legs to IVC wall prior to deployment were evaluated. The association of those pre-deployment factors with post-deployment filter tilting was analyzed. The association noted in the experimental study was then evaluated in a retrospective clinical series of 21 patients.

Results

In the experimental study, there was a significant association between the pre-deployment angle and post-deployment filter tilting (P<0.0001). With a low pre-deployment angle (≤5°), a significant association was noted between filter tilting and the proximity of the constrained filter legs to the far IVC wall (P=0.001). In a retrospective clinical study, a significant association between the pre-deployment angle and post-deployment filter tilting was also noted with a linear regression model (P=0.026).

Conclusion

Significant association of the pre-deployment angle with post-deployment GTF tilting was shown in both the experimental and clinical studies. The experimental study also showed that proximity of filter legs is relevant when pre-deployment angle is small. Addressing these factors may result in a lower incidence of filter tilting.

Le texte complet de cet article est disponible en PDF.

Keywords : Inferior vena cava filter, Experimental studies, Interventional imaging, Inferior vena cava


Plan


© 2017  Editions françaises de radiologie. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 98 - N° 7-8

P. 543-549 - juillet 2017 Retour au numéro
Article précédent Article précédent
  • Radiofrequency ablation of pulmonary tumors near the diaphragm
  • T. Iguchi, T. Hiraki, H. Gobara, H. Fujiwara, J. Sakurai, Y. Matsui, T. Mitsuhashi, S. Toyooka, S. Kanazawa
| Article suivant Article suivant
  • Evaluation of S100B blood level as a biomarker to avoid computed tomography in patients with mild head trauma under antithrombotic medication
  • A. David, C. Mari, F. Vignaud, D. Masson, L. Planche, E. Bord, R. Bourcier, E. Frampas, E. Batard, H. Desal

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.