Introduction

New 3D automatic software (Heart model (HM), Philips Healthcare, Andover, Massachusetts) is now available to quantify left heart chamber size and function. Its feasibility and accuracy in children have not been reported. The aim of this study was to assess feasibility and reproducibility of HM in healthy children, and to compare LV indices obtained by this technique with classic 2D biplane method.

Methods

Sixty consecutive healthy children aged more than 5 years were included (mean age 6.5 years, mean weight 20.8kg). All infants underwent TTE (Epic 7, X5-1, Philips Healthcare, Andover, Massachusetts). Left ventricle (LV) end-systolic and end-diastolic volumes, LV ejection fraction (EF) and left atrium (LA) end-systolic volume were obtained by biplane Simpson method. 3D dataset was obtained from apical 4-chamber view and analyzed using HM software. LV indices obtained by the two methods were compared.

Results

Feasibility of HM was 80%, in 30 cases (62.5%) with contour adjustment and 18 cases (37.5%) without contour adjustment. Intraobserver variability was (0%, 4% and 8% for HM without contour adjustment, HM with contour adjustment, and biplane method respectively). Interobserver variability was (6% and 12% for HM with contour adjustment and biplane method respectively). Test- retest variability for HM was 6.5%. Correlation between HM and biplane measurements was strong (r=0.85 to 0.94). LVEF obtained by HM was lower than that obtained by biplane method (bias −8%). However, LV volumes and LA end-systolic volume were larger (bias +12 and +8 respectively) (Fig. 1).

Conclusion

HM is a promising software for assessment of left heart chamber volume and function. Its feasibility in infants aged more than 5 years is good with excellent reproducibility. Working is going on to validate this technique in children with congenital heart disease and to compare it with MRI measurement as the gold standard.

Le texte complet de cet article est disponible en PDF.

Plan


© 2017  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 10 - N° 1

P. 64 - janvier 2018 Retour au numéro
Article précédent Article précédent
  • Impact of mitral regurgitation correction (Mitraclip® vs. surgical repair) on left ventricular myocardial performance according to baseline LVEF
  • A. Hubert, E. Galli, E. Samset, C. Leclercq, E. Donal
| Article suivant Article suivant
  • Speckle tracking analysis with the two main ultrasound software in pediatrics: Comparison between QLAB by Philips Healthcare and EchoPac by General Electric Healthcare
  • P. Amedro, C. Bredy, S. Guillaumont, K. Lavastre, G. De La Villeon, T. Mura, L. Gamon, O. Cazorla, A. Lacampagne, M. Vincenti

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.