Reliability and validity of MicroScribe-3DXL system in comparison with radiographic cephalometric system: Angular measurements - 17/04/18
Cet article a été publié dans un numéro de la revue, cliquez ici pour y accéder
Summary |
Aim |
The aim of the study was to assess the reliability and validity of cephalometric variables from MicroScribe-3DXL.
Materials and methods |
Seven cephalometric variables (facial angle, ANB, maxillary depth, U1/FH, FMA, IMPA, FMIA) were measured by a dentist in 60 Malay subjects (30 males and 30 females) with class I occlusion and balanced face. Two standard images were taken for each subject with conventional cephalometric radiography and MicroScribe-3DXL. All the images were traced and analysed. SPSS version 2.0 was used for statistical analysis with P-value was set at P<0.05.
Results |
The results revealed a significant statistic difference in four measurements (U1/FH, FMA, IMPA, FMIA) with P-value range (0.00 to 0.03). The difference in the measurements was considered clinically acceptable. The overall reliability of MicroScribe-3DXL was 92.7% and its validity was 91.8%.
Conclusion |
The MicroScribe-3DXL is reliable and valid to most of the cephalometric variables with the advantages of saving time and cost. This is a promising device to assist in diverse areas in dental practice and research.
Le texte complet de cet article est disponible en PDF.Key-words : Lateral cephalometric radiography, MicroScribe-3DXL, Radiation, Malay norms