Skeletal landmarks for TKR implantations: Evaluation of their accuracy using EOS imaging acquisition system - 05/03/09

Doi : 10.1016/j.otsr.2008.05.001 
B. Schlatterer a, , I. Suedhoff b, X. Bonnet b, Y. Catonne c, M. Maestro a, W. Skalli b
a Monaco Sportsmedicine and Surgery Institute, 11, avenue d’Ostende, 98000 Monaco, Monaco 
b Biomecanics Laboratory ENSAM – CNRS, 151, boulevard de l’Hôpital, 75013 Paris cedex, France 
c Department of Orthopaedic Surgery and Traumatology, Pitié-Salpêtrière Hospital, 47–83, boulevard de l’Hôpital, 75651 Paris cedex 13, France 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Summary

Introduction

Lower extremity alignment remains one essential objective during total knee replacement. Implants positioning analysis requires selecting reliable skeletal landmarks. Our objective was to in vivo evaluate the precision of the implemented skeletal landmarks. This evaluation was based on multiple three-dimensional (3D) computer reconstructions of the lower extremity derived from an EOS® biplanar low-dose X-ray system acquisition. A 3D angle measurement protocol was used.

Hypothesis

Currently defined landmarks carry a tolerable uncertainty margin, which can still probably be further improved.

Material and methods

Nine lower extremity 3D computer reconstructions were obtained from an EOS protocol based on seven simultaneous A–P and lateral views performed in standing position. A database was established by four operators; finally, building up a total of 99 in vivo 3D reconstructions of these nine lower extremities. Specific algorithms were used for such 3D reconstructions of lower extremities based on bone points and pre-identified contours on X-ray. Four femoral landmarks and four tibial landmarks were thus defined. For each bone and each landmark studied, a mean landmark for the 11 consecutive series elements was established. The deviation from each constructed landmark to the corresponding mean landmark was calculated based on the anteroposterior (x), longitudinal (y) and mediolateral axes (z), in translation (Tx, Ty, Tz) and in rotation (Rx, Ry, Rz). Uncertainty was estimated by the 95% confidence interval (95% CI).

Results

The landmarks located at the middle of the segment joining the center of each posterior condyle and at the barycenter of the plateaux showed a greater reliability; these landmarks uncertainty (95% CI) of Tx, Ty, Tz was less than 1, 0.5, 1.5 mm for the femur and 1.5, 0.6, 0.6 mm for the tibia, respectively. The femoral landmarks using the center or posterior edge of the posterior condyles to define the mediolateral axis were retained; for rotations Rx, Ry, and Rz, uncertainty remained less than 0.3, 4, and 0.5°. All of the tibial landmarks had a comparable reliability in rotation, 95% of the Rx and Rz deviations were under 0.5 and 1.3°, respectively, with a mean error less than 1°. For the tibial rotation Ry, the mean error was greater (4°), with uncertainty (95% CI) at 11.2°. All tibial translations showed a mean error of 1 mm. The 3D implantation angles were measured on two patients using preoperative 3D skeletal reconstructions and 3D geometric models of the implants repositioned on postoperative EOS® knee X-rays.

Discussion

The posterior condyles are rarely involved in the arthritic wear process, making them an anatomic landmark of choice in the analysis of the femoral component positioning. The femoral landmarks using the posterior condyles were sufficiently reliable for clinical use. However, the posterior contours of the tibial plateaux were less precise. The knees should be staggered from an anteroposterior perspective on the EOS® lateral images so that they can be visualized separately. The anatomic zones on which the skeletal landmarks are based are usually removed by the bone cuts, making it preferable to save the preoperative computer reconstructions to analyze the postimplantation 3D reconstruction.

Conclusion

The lower extremity skeletal landmarks precision relates to the quality of the corresponding 3D reconstructions. Except for tibial rotation, all the translation and rotation parameters were estimated within a mean error margin inferior to 1.2 mm and 1.3°, respectively. Making the reconstruction algorithms more robust would render certain anatomic zones even more precise. Biplanar low-dose EOS® X-ray system is a tool of the future to generate 3D knee X-rays that can improve the evaluation and follow-up of total knee arthroplasty patients.

Level of evidence: level IV, diagnostic study.

Le texte complet de cet article est disponible en PDF.

Keywords : Biplanar low-dose EOS® X-ray 3D imaging, Total knee replacement, Alignment


Plan


© 2008  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 95 - N° 1

P. 2-11 - février 2009 Retour au numéro
Article précédent Article précédent
  • Expanding into the future the OTSR is born
  • J.-M. Thomine
| Article suivant Article suivant
  • Cemented all polyethylene tibial insert unicompartimental knee arthroplasty: A long term follow-up study
  • S. Lustig, J.-L. Paillot, E. Servien, J. Henry, T. Ait Si Selmi, P. Neyret

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.