A new knee arthrometer, the GNRB®: Experience in ACL complete and partial tears - 03/06/09

Doi : 10.1016/j.otsr.2009.03.009 
H. Robert a, , S. Nouveau a, S. Gageot b, B. Gagnière c
a North Mayenne Hospital Orthopaedic department, 229, boulevard Paul-Lintier, 53100 Mayenne, France 
b Superior School of Informatics, Electronic and Automatism (West Esiea), 53000 Laval, France 
c Institute of Sanitary Engineering, 35042 Rennes, France 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Summary

Introduction

Clinical diagnosis of anterior cruciate ligament (ACL) tears (Lachman test and Pivot shift test in valgus and internal rotation) is reliable in case of complete ACL tear but reveals elusive in case of partial tears. Quantitative assessment of anterior tibial translation proves to be imprecise, subjective and poorly reproducible especially with the KT-1000 arthrometer. We developed the GNRB®, an alternative original anterior knee laxity measurement device. The lower limb is placed in a rigid support with the knee at 0° of rotation, the restraining power being recorded. A 0–250N thrust force is transmitted by a jack to the upper segment of the calf. This force is only applied in the absence of hamstring muscles contraction. Displacement of the anterior tibial tubercle is recorded using a sensor with a 0.1mm precision.

Hypothesis

We hypothesize that this knee laxity measurement device is more reliable and reproducible than other currently available arthrometers.

Material and methods

During a first validation study, the GNRB® was compared to the KT-1000 arthrometer, in 20 pairs of healthy knees, measurements being performed by two investigators. Variance analyses were carried out at 134 N. In a second clinical study, 21 complete ACL tears (the notch is devoid of ACL) and 24 partial ACL tears (anterior or posterior bundle tear and cicatricial ACL remnant in continuity) were tested with these arthrometers to exact a differential laxity threshold value between both knees at 250 N. Statistical analysis was subsequently performed using variance and ROC curves analysis.

Results

The GNRB® arthrometer reveals to be significantly more reproducible than the KT-1000, irrespectively of the tester’s experience level. Moreover, unlike the KT-1000, the achieved measurement is independent from the uninvolved side. Reproducibility of laximetry proves to be significantly better with the GNRB® than with the KT-1000, wherever the examiner’s experience stands and whatever the evaluated side condition could be. When differential laxity threshold value was 3mm in complete ACL tears, sensitivity was 70% and specificity 99% at 134 N. Using a 1.5mm threshold value in ACL partial tears, the arthrometer sensitivity was 80% and specificity was 87% at 134 N.

Discussion

Reproducibility of laximetry was significantly better with the GNRB® than the KT-1000 device, wherever the examiner’s experience stands and whatever the evaluated side-condition could be. The GNRB® reports various supplementary advantages compared with other available laximeters. Good control of the investigated limb position in rotation, recording of translation in the absence of hamstring muscles contraction and in direct comparison with the KT-1000: reproducibility, constant pressure, arthrometry improved accuracy and automated measurements recording. The GNRB® might be used for diagnosis of partial and complete ACL tears and during follow-up of reconstructed or not ACL tears.

Level of evidence: type II. Prospective comparative study

Le texte complet de cet article est disponible en PDF.

Keywords : Arthrometer, Anterior cruciate ligament, KT-1000, Telos, Partial anterior cruciate ligament tear, Complete anterior cruciate ligament disruption


Plan


© 2009  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 95 - N° 3

P. 171-176 - mai 2009 Retour au numéro
Article précédent Article précédent
  • Contents
| Article suivant Article suivant
  • LCS mobile-bearing total knee replacement. A 10-year’s follow-up study
  • J.-C. Vogt, C. Saarbach

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.