S'abonner

Complications following rotating hinge Endo-Modell (Link®) knee arthroplasty - 19/10/09

Doi : 10.1016/j.otsr.2009.07.013 
B. Guenoun a, , L. Latargez a, M. Freslon a, G. Defossez b, N. Salas a, L.-E. Gayet a
a Department of Orthopaedic Surgery and Traumatology, Poitiers Teaching Hospital Center, 2, rue de la Milétrie, BP 577, 86021 Poitiers cedex, France 
b Public Health Academic Institute, Poitiers Teaching Hospital Center, France 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Sous presse. Épreuves corrigées par l'auteur. Disponible en ligne depuis le Monday 19 October 2009
Cet article a été publié dans un numéro de la revue, cliquez ici pour y accéder

Summary

Background

Rotating hinge knee prostheses are indicated in revisions especially when major ligament laxity or substantial AP deformities are present. These situations make ligament balancing difficult with less constrained design implants. Despite its use for nearly 50 years, this type of prosthesis continues to have a poor reputation due to a high complication rate.

Hypothesis

Complications are frequent after this type of arthroplasty and the complication rate is similar in primary or revision arthroplasties. The objective of this study is to report the medium-term results of these implants and determine the eventual predictive factors of complications in order to refine operative indications.

Material and methods

In this retrospective study of patients operated on between 1998 and 2006, 85 Endo-Modell (Link®) rotating hinge knee prostheses had been used in 61 females and 24 males. The mean age at surgery was 72.4 years (range, 32–92 years). Fifty-two arthroplasties were primary and 33 were revisions either for loosening (24) or deep infections (9). The mean follow-up was 36 months±22 (range, 0–75 months).

Results

Complications were observed in 24 patients (28.2%): nine deep infections, four patellar complications, and three cases of aseptic loosening. No significant difference was found between the primary arthroplasties and the revisions regarding all complication types. A significant relation was established between the occurrence of a complication and presence of several associated comorbidity factors (obesity, heart disease, diabetes, etc.).

Discussion

The use of this type of implant carries a high risk of complications, higher than the one pertaining to unconstrained design prostheses; this fact is noted irrespective of the surgical indication and other comparison elements. The leading criteria to poor functional results appear to be the indication (gonarthrosis with substantial ligament laxity at primary surgery) and the number of associated comorbidities. These prostheses should therefore be restricted to selected indications, notably in view of the fact that less constrained prostheses give superior outcomes.

Level of evidence

Level IV. Retrospective therapeutic study.

Le texte complet de cet article est disponible en PDF.

Keywords : Total knee arthroplasty, Complication, Hinge Prosthesis, Failure


Plan


© 2009  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.