Is alumina-on-alumina ceramic bearings total hip replacement the right choice in patients younger than 50 years of age? A 7- to 15-year follow-up study - 04/10/10

Doi : 10.1016/j.otsr.2010.02.013 
P. Boyer a, D. Huten b, P. Loriaut a, V. Lestrat a, C. Jeanrot a, P. Massin a,
a Department of Orthopaedic Surgery, Bichat Claude-Bernard Hospital, North Paris Teaching Hospitals Group Nord, Paris Diderot University, 46, rue Henri-Huchard, 75877 Paris cedex 18, France 
b Department of Orthopaedic Surgery, Rennes Teaching Hospital Center, 16, boulevard Bulgarie, BP 90347, 35203 Rennes cedex 2, France 

Corresponding author. Tel.: +33 0 1 40 25 75 03.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Summary

Introduction

The alumina-on-alumina bearing couple in total hip replacement seems to be well adapted for young and active patients because of the absence of wear and the rarity of osteolysis. Over the long term, doubts persist as to the cementless cup fixation and on the functioning of this bearing system because of possible acoustic emissions during use.

Hypothesis

In young subjects, the ceramic-on-ceramic bearing system limits wear and osteolysis occurrences, without exposing patients to serious side effects.

Material and mehtods

We report the results, with between 7 and 15 years of follow-up, for 32mm-diameter alumina-on-alumina implants in 76 patients younger than 50 years of age (83 hips), combining cementless press-fit hemispheric cups with titanium stems, [either cemented (63 Osteal™ stems) or cementless (20 Multicône™ stems)], with particular attention paid to cup fixation and noise emissions during implant function. First-generation or Cerafit trellis™ acetabular components had a riveted titanium mesh (31 cases), whereas the most recent (Cerafit hydroxyapatite [HA]™) cups had a porous surface coated with hydroxyapatite (52 cases).

Results

Three cases of aseptic loosening of the cemented stems were observed as well as late migration of a Cerafit trellis™ cup in the 12th postoperative year. One ceramic insert broke in the eighth postoperative year. With the exception of one case, the patients, questioned retrospectively, reported no audible noise. With aseptic loosening (revised or not), the criterion for failure, the 12-year survival rate was 91±11% for the Cerafit trellis™ acetabular components and 91±16% for the cemented Osteal™ stems. The 9- and 7-year survival rates for the Cerafit HA™ cups and the Multicône™ stems, respectively, were 100%. Including all revisions for any cause, the 10-year survival rate of the entire series was 92%±11%.

Discussion

Despite the absence of wear and osteolysis, the long-term survival of these implants in young subjects should be improved. Although longer follow-up is necessary to formulate a definitive opinion, we tend to prefer cementless stem and cup fixation in ceramic-on-ceramic bearing systems.

Level of evidence

Level 4 retrospective study.

Le texte complet de cet article est disponible en PDF.

Keywords : Alumina-on-alumina total hip replacement, Ceramic-on-ceramic, Patients younger than 50 years of age, Long-term results, Squeaking


Plan


© 2010  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 96 - N° 6

P. 616-622 - octobre 2010 Retour au numéro
Article précédent Article précédent
  • Dual mobility hip arthroplasty wear measurement: Experimental accuracy assessment using radiostereometric analysis (RSA)
  • V. Pineau, B. Lebel, S. Gouzy, J.-J. Dutheil, C. Vielpeau
| Article suivant Article suivant
  • Is femoral head fracture-dislocation management improvable: A retrospective study in 110 cases
  • J. Tonetti, S. Ruatti, V. Lafontan, F. Loubignac, P. Chiron, H. Sari-Ali, P. Bonnevialle

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.