Dual mobility retentive acetabular liners and wear: Surface analysis of 40 retrieved polyethylene implants - 20/01/14

Doi : 10.1016/j.otsr.2013.12.011 
P. Adam , F. Farizon, M.-H. Fessy
 Centre d’orthopédie et traumatologie, hôpital Bellevue, CHU de Saint-Étienne, 42055 Saint-Étienne cedex 2, France 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Sous presse. Épreuves corrigées par l'auteur. Disponible en ligne depuis le Monday 20 January 2014
Cet article a été publié dans un numéro de la revue, cliquez ici pour y accéder

Abstract

Purpose of the study

Dual mobility systems with retentive acetabular liners have been used in clinical practice for total hip arthroplasty since 1976. The dual mobility system preserves a wide range of motion while providing greater stability. This study measured wear on the concave and convex surfaces of 40 retrieved polyethylene liners, to evaluate the advantages of this system in relation to wear.

Material and methods

Forty polyethylene inserts that had been removed due to infection or mechanical failure after a mean 8 years were analyzed. The mean age of patients at arthroplasty was 46 years old. Macroscopic analysis was followed by surface analysis with direct measurement of changes in the curvature radii. The internal concave surface was measured in three dimensions using a 4-mm stylus (BNH 706). External convexity was measured by lateral projection. The estimated error was ± 5 μm for both measurement methods. Manufacturers’ tolerance for these implants was approximately 50 μm. Linear wear and wear volume was determined by comparing the measured dimensions with the theoretical dimensions of new liners.

Results

Macroscopically, all of the pieces studied had lost the initial machined grooves on the convex surface; 40% of the pieces showed visible wear of the retentive collar. Mean annual convex surface wear was 9 μm (SD 9μm) and 73 μm (SD 69 μm) for the concave surface. Mean total wear, which was the sum of the wear on the convex and concave surfaces was 82 μm (SD 72 μm). Wear volume was 28.9 mm3/yr for the convex surface (SD 27.6) and 25.5 for the concave surface (SD 23.2) with a mean annual total wear volume of 54.3 mm3/yr (SD 39.6).

Discussion

Total wear in the 40 dual mobility liners that had functioned in vivo was similar to that reported in metal-polyethylene bearings with 22.2mm femoral heads. The results of wear in both the convex and concave surfaces show that wear with the dual mobility system was not increased compared to conventional metal-polyethylene bearings, while providing better retention and greater stability.

Conclusion

The use of dual mobility acetabular liners is an attractive solution when a metal-polyethylene bearing is needed. The increased joint stability is not associated with increased wear.

Le texte complet de cet article est disponible en PDF.

Keywords : Hip, Prosthesis, Dual Mobility, Bipomar prosthesis


Plan


 Original article. For citation, use not the present reference but that of the original publication: Adam P, Farizon F, Fessy MH. Dual articulation retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants. Rev Chir Orthop Reparatrice Appar Mot 2005;91(7):627-36.


© 2013  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.