Determination of HbA1c concentrations in patients with acute myocardial infarction: comparison of the DCA 2000 device with the HPLC method - 17/02/08

Doi : DM-06-2005-31-3-1262-3636-101019-200514511 

S Hadjadj [1],

F Duengler [1],

M Barriere [2],

G Mauco [2],

D Coisne [3],

F Warnier [4],

P Sosner [3],

F Torremocha [1],

D Herpin [3],

R Marechaud [1]

Ver las filiaciones

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Détermination de la concentration d'HbA1c chez des patients présentant un infarctus myocardique aigu: comparaison du dispositif DCA 2000 et de la méthode HPLC

État de la question

Des études récentes suggèrent que l'HbA1c est un important prédicteur de l'état glycométabolique des patients hospitalisés pour infarctus myocardique aigu (IMA).

Objectifs

Nous avions pour objectif de comparer les résultats de concentrations d'HbA1c obtenues par 2 méthodes différentes chez des patients avec IMA.

Design de l'étude et méthodes

Dans une première étude, l'HbA1c a été mesurée chez tous les patients consécutivement hospitalisés pour AMI, pendant une période de 6 mois, en utilisant la méthode HPLC et le dispositif DCA 2000 dans le laboratoire de biochimie. Dans une seconde étude, l'HbA1c mesurée par DCA 2000 dans l'unité de soins intensifs a été comparée à celle déterminée par HPLC dans le laboratoire de biochimie dans un échantillon similaire de patients. Chez les patients sans antécédents personnels de diabète, ceux qui présentaient une HbA1c > 6,5 % (méthode HPLC) ont été classés comme susceptibles d'être diabétique.

Résultats

Un total de 146 patients a été inclus (119 hommes, 27 femmes; âge moyen: 63 ± 15 ans). Un antécédent personnel de diabète était connu chez 27 patients. Les HbA1c déterminées par les 2 techniques étaient en corrélation étroite (r = 0,939, P ≪ 0,0001). La moyenne des différences (analyse de Bland et Altman) était de 0,4 ± 0,3 %.

Par rapport à la méthode HPLC, le dispositif DCA 2000 a une sensibilité de 81,8 ± 11,6 et une spécificité de 99,1 ± 0,9 % dans la détection d'un possible diabète. La précision diagnostique de la méthode DCA était de 97,5 ± 1,4 %.

Dans la seconde étude, les concentrations d'HbA1c de 21 sujets supplémentaires, déterminées dans une unité de soins intensifs, n'étaient pas différentes de celles des 21 premiers patients de la première étude.

Conclusions

L'HbA1c peut être effectivement déterminée par le dispositif DCA 2000. Cette méthode est fiable et facile à mettre en place dans une unité de soins intensifs.

Background

Recent studies suggest that HbA1c is an important predictor of the glycometabolic state of patients admitted for acute myocardial infarction (AMI).

Objective

We aimed at comparing the results of HbA1c concentrations obtained by 2 different methods in patients with AMI.

Research design and methods

In a first study, HbA1c was measured in all patients consecutively hospitalized for AMI, during a 6month period using the HPLC method and the DCA 2000 device in the biochemistry laboratory. In a second study, HbA1c measured by the DCA 2000 device in the intensive care unit was compared with HbA1c determined by HPLC in the biochemistry laboratory in a similar sample of patients. In patients without personal history of diabetes, those patients with HbA1c > 6.5% (HPLC method), were classified as possible diabetes.

Results

A total of 146 patients were included (119 males, 27 females; mean age: 63 ± 15 years). Twenty-seven of the patients had a personal history of diabetes. HbA1c determined by 2 techniques were highly correlated (r = 0.939, P ≪ 0.0001). The mean of the differences (Bland and Altman analysis) was 0.4 ± 0.3%.

Compared with the HPLC method, the sensitivity of DCA 2000 device for the detection of possible diabetes was 81.8 ± 11.6 and the specificity was 99.1 ± 0.9%. The diagnostic accuracy of DCA method was 97.5 ± 1.4%.

In the second study, the HbA1c concentrations of 21 additional subjects, determined in an intensive care unit, were not different from the first 21 patients of the first study.

Conclusions

HbA1c can be effectively determined using the DCA2000 device. This method is reliable and easy to be implemented in an intensive care unit.


Mots clés : Infarctus myocardique aigu , Diabète , HbA1c , Dépistage

Keywords: Acute myocardial infarction , Diabetes , Screening , HbA1c.


Esquema



© 2005 Elsevier Masson SAS. Tous droits réservés.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 31 - N° 3

P. 290-294 - juin 2005 Regresar al número
Artículo precedente Artículo precedente
  • Epidemiology of dyslipidemia among schoolchildren in Sousse, Tunisia
  • I Harrabi, H Ghannem, R Gaha, M Hochlaf, K Limam, AS Essoussi
| Artículo siguiente Artículo siguiente
  • Combination of apolipoprotein E2 and lipoprotein lipase heterozygosity causes severe hypertriglyceridemia during pregnancy
  • S Hiéronimus, P Benlian, P Bayer, A Bongain, A Fredenrich

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.