S'abonner

0090: Comparison between non-invasive coronary flow reserve, instantaneous wave-free ratio, and fractional flow reserve, to assess the functional significance of LAD stenosis of intermediate severity - 12/02/16

Doi : 10.1016/S1878-6480(16)30030-1 
Patrick Meimoun , Dorothée Ardourel, Jerome Clerc, Frederic Elmkies, Anne Luycx-Bore, Sonia Martis, Hamdane Zemir, Jacques Boulanger
 CH Compiègne, Compiègne, France 

*Corresponding author:

Résumé

Introduction

Assessment of the functional significance of left anterior descending coronary artery (LAD) stenosis of intermediate severity (IS) is challenging. A direct comparison of fractional flow reserve (FFR), instantaneous wave-free ratio (IFR), and non-invasive coronary flow reserve (CFR) has never been performed. Our objective was to test the usefulness of noninvasive CFR by comparison to invasive FFR and IFR in patients with LAD stenosis of angiographic IS and stable coronary artery disease.

Methods

58 stable consecutive patients (mean age, 68±10 years; with angiographic proximal or mid LAD stenosis of IS (40-70% diameter stenosis on quantitative coronary angiography), no previous anterior myocardial infarction, were prospectively studied. They underwent iFR which was calculated as a trans-lesion pressure ratio during a specific period of baseline diastole, FFR with intracoronary bolus adenosine (150μg), and CFR using intravenous adenosine (140μg/kg/min over 2min) in the distal part of the LAD, the same day. CFR was defined as hyperemic peak diastolic LAD flow velocity divided by baseline flow velocity, and FFR was defined as distal pressure divided by mean aortic pressure during maximal hyperemia.

Results

The mean values of iFR, FFR, and CFR were 0.88±0.07, 0.81±0.08, and 2.4±0.6 respectively. A significant correlation was found between CFR and FFR (r=0.72, curvilinear relationship), FFR and IFR (r=0.63, linear relationship), and between CFR and IFR (r=0.44) (all, p<0.01). Using a ROC curve analysis, the best cut-off to detect a significant lesion based on FFR assessment (FFR<0.8, n=16) was iFR≤0.86 with a sensitivity (Se) of 75%, specificity (Sp) of 81%, AUC 0.8±0.05; and CFR≤2 with a Se of 82%, Sp of 85%, AUC 0.9±0.03, (all, p<0.001). Based on these cut-offs, discordant results between CFR and FFR were observed in 9 cases (accuracy 84%), between CFR and iFR in 14 cases (accuracy 76%), and between iFR and FFR in 11 cases (accuracy 81%).

Conclusion

In stable patients with LAD stenosis of IS, non-invasive CFR is a useful tool to detect a significant lesion based on FFR. Furthermore, CFR is better correlated to FFR than to iFR, and its accuracy seems as good as iFR in this setting.

Le texte complet de cet article est disponible en PDF.

© 2016  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 8 - N° 1

P. 10 - janvier 2016 Retour au numéro
Article précédent Article précédent
  • 0041: Contrast-induced microvascular dilatation: implications for fractional flow reserve measurements
  • Julien Adjedj, Gabor Toth, Mariano Pellicano, Guiseppe Di Gioia, Angela Ferrara, Vincent Flore, Nils Johnson, Olivier Muller, Emanuele Barbato, Bernard de Bruyne
| Article suivant Article suivant
  • 0459: Super-Minimalist Immediate Mechanical Intervention (Super-MIMI) study
  • Loic Belle, Hend Madiot, Hélène Bouvaist, Nicolas Delarche, Fréderic Bouisset, Pierre-Yves Petiteau, Ziad Boueri, Mohamed Abdellaoui, Chloé Durier, Olivier Dubreuil

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.