0305: A new turn in SPECT myocardial perfusion imaging: data-driven cardiac gating is now possible with the new generation CZT SPECT Discovery NM 530c - 12/02/16

Doi : 10.1016/S1878-6480(16)30137-9 
Doumit Daou , ((1)) , Rémy Sabbah ((2)), Carlos Coaguila ((3)), Hatem Boulahdour ((3))
(1) APHP-Hôpital Cochin, Paris, France 
(2) CHU Besançon, Jean Minjoz, Besançon, France 
(3) CH Bigorre, Tarbes, France 

*Corresponding author:

Résumé

Aim

We previously developed a data-driven (DD) respiratory-motion (RM) correction method for conventional SPECT gamma-cameras (REGAT) and adapted it to the new CZT camera (Discovery NM 530c). We recently reported that RM correction with REGAT applied to CZT myocardial perfusion SPECT imaging (MPI) is clinically feasible and impacts substantially myocardial perfusion defects. In this evaluation, we study whether REGAT applied to MPI (Discovery NM 530c) is capable of generating a data-driven (DD) cardiac gating signal allowing the generation of valid global left ventricular (LV) function parameters (EDV: end diastolic volume; ESV: end systolic volume; EF: ejection fraction).

Materials and methods

Were included 7 patients addressed for stress/rest MPI. All patients had prone stress MPI (2 MBq/Kg 99mTc-Tetrofosmin) and rest MPI 3-hours later (6 MBq/Kg). All acquisitions were made on Discovery NM 530c. Each acquisition was processed with REGAT to generate a dynamic SPECT acquisition study. The latter was processed to generate a DD cardiac gating signal and generate a mean DD cardiac GSPECT study (GSPECT-DD). In parallel, a mean ECG cardiac GSPECT study was generated using the ECG trigger signal provided by traditional ECG monitor (GSPECT-M). The 2 generated cardiac GSPECT studies were reconstructed on Xeleris workstation and processed with Emory Cardiac Toolbox (ECT). LV EDV, ESV and EF were compared between cardiac GSPECT-DD and GSPECT-M.

Results

Stress LV EVD, ESV, and EF were 91±24mL, 29±13ml, and 68±10% vs 95±23ml, 29±12ml, and 70±11% with GSPECT-DD vs GSPECTM respectively (P:NS). Rest LV EDV, ESV, and EF were 97±21ml, 32±10ml, and 67±6% vs 101±21ml, 31±10ml, and 69±6% with GSPECT-DD vs GSPECT-M respectively (P:NS).

Conclusion

Data-driven cardiac gating of MPI with Discovery NM 530c processed with REGAT is clinically feasible. It provides LV global systolic function parameters similar to those provided by the traditional clinically used ECG monitor gating.

Le texte complet de cet article est disponible en PDF.

© 2016  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 8 - N° 1

P. 45 - janvier 2016 Retour au numéro
Article précédent Article précédent
  • 0563: Safety, feasibility and interest of transthoracic echocardiography in a deployed French military Ebola virus disease treatment center in Guinea
  • Gilles Rolland Cellarier, Julien Bordes, Raphaël Poyet, Frédéric Pons, Eléonore Capilla, Christophe Jego
| Article suivant Article suivant
  • 0056: Assessment of left ventricular filling pressures: a prospective speckle tracking echocardiography compared to invasive hemodynamic study
  • Julien Voyez, Anthony Foulon, Bertrand Girodet, Mathias Guinot, Karine Bauley, Pierre Nazeyrollas, Damien Metz

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.