Is determination of the hip rotation center during computer-assisted surgery influenced by the pinless femoral sensor attachment method? An in vitro preliminary study comparing osseous pin fixation - 25/04/16

Doi : 10.1016/j.otsr.2016.01.014 
E. de Thomasson a, , S. Gouzy b, F. Cloerec b, L. Geais b
a Institut Mutualiste Montsouris, 42, boulevard Jourdan, 75014 Paris, France 
b Société Amplitude, 11, cours Jacques-Offenbach, 26000 Valence, France 

Corresponding author. Tel.: +33 1 56 61 64 09; fax: +33 1 56 61 64 38.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Introduction

Computer-assisted surgery improves the positioning of hip prostheses but requires use of transosseous pins requiring a complementary approach exposing the patient to rare but at times serious complications. The use of sensor arrays attached to the skin could advantageously replace pins provided that comparable results are obtained, but their validity has not yet been assessed. We conducted a prospective in vitro study to: measure the possible error of a cutaneous versus transosseous fixation to determine the hip rotation center (HRC) position and determine the inter- and intraobserver reproducibility of the cutaneous versus the transosseous fixation.

Hypothesis

Use of cutaneous sensor arrays while recording the HRC is sufficiently reliable for its calculation algorithm to provide measurement accuracy within 5mm.

Materials and methods

A rigid array attached with either a silicone strap or an adhesive were compared to a transosseous array. Four series of 96 HRC measurements were collected by four operators on two cadavers, half with an array attached with a strap and half with an adhesive. The results were compared to those obtained by a sensor attached with transosseous pins.

Results

On condition that the hip-knee is mobilized in extension, a sensor array attached with an adhesive gives results with comparable accuracy (standard deviation [SD]: 2.89mm [1.9–4.8]) to the results obtained with a transosseous fixation (SD: 1.2mm [0.9–1.6]), with no significant inter- or intraobserver variation (0.97<ICC<0.99). If the knee was flexed, accuracy was within more than 1cm, or twice the predetermined limit. Only the adhesive array gave measurements systematically within the limits established with the data spread (8.7%) close to that obtained with a pin (6%).

Discussion

An adhesive array seems to be a reliable alternative to transosseous pins to evaluate HRC and provides a three-dimensional landmark, indispensable to total hip arthroplasty navigation. These preliminary results confirm our hypothesis but should be validated in vivo.

Level of evidence

Level III, comparative prospective in vitro study.

Le texte complet de cet article est disponible en PDF.

Keywords : THA, Pinless, Navigation, Hip rotation center


Plan


© 2016  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 102 - N° 3

P. 375-378 - mai 2016 Retour au numéro
Article précédent Article précédent
  • Sequential therapy with “vacuum sealing drainage-artificial dermis implantation-thin partial thickness skin grafting” for deep and infected wound surfaces in children
  • X.-g. Yuan, X. Zhang, Y.-x. Fu, X.-f. Tian, Y. Liu, J. Xiao, T.-w. Li, L. Qiu
| Article suivant Article suivant
  • No effect of femoral offset on bone implant micromotion in an experimental model
  • F. Amirouche, G. Solitro, A. Walia

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.