Abstract

Background

Despite excellent long-term outcomes, posterior stabilisation by a third condyle continues to receive unwarranted criticism regarding patellar complications and instability.

Hypothesis

Complication rates with a tri-condylar posterior-stabilised implant are similar to those with other posterior-stabilised prostheses and have diminished over time due to improvements in prosthesis design.

Material and methods

Post-operative complications and revision rates were assessed retrospectively in a prospective cohort of 4189 consecutive patients who had primary total knee arthroplasty (TKA) using a tri-condylar posterior-stabilised implant (Wright-Tornier) and were then followed-up for at least 24 months. The analysis included 2844 knees. The prosthesis generations were HLS1®, n=20; HLS2®, n=220; HLS Evolution®, n=636; HLS Noetos®, n=1373; and HLS KneeTec®, n=595. Complications were compared across generations by applying Fisher's exact test, and survival was compared using the Kaplan-Meier method.

Results

At last follow-up, there had been 341 (12%) post-operative complications in 306 (10.8%) knees, including 168 (5.9%) related to the implant, 41 (1.4%) infections, and 132 (4.6%) secondary complications unrelated to the implant. Re-operation was required for 200 complications (7%), including 87 (3.1%) consisting in revision of the prosthesis. Implant-related complications were stiffness (n=67, 2.4%), patellar fracture (n=34, 1.2%), patellar clunk syndrome (n=25, 0.9%), patellar loosening (n=3, 0.1%), tibial/femoral loosening (n=15, 0.5%), polyethylene wear (n=3, 0.1%), and implant rupture (n=1, 0.04%). Significant differences across generations were found for stiffness (P<0.0001), patellar fracture (P=0.03), clunk syndrome (P=0.03), and polyethylene wear (P=0.004), whose frequencies declined from one generation to the next. Overall 10-year survival was 92% with no significant difference across generations (P=0.1).

Discussion

Outcomes of tri-condylar posterior-stabilised TKA are similar to those obtained using other posterior-stabilised implants. Neither patellar complications nor instability are more common, and improvements in implant design have contributed to correct early flaws.

Level of evidence

IV, historical cohort, retrospective assessment of prospectively collected data.

Le texte complet de cet article est disponible en PDF.

Keywords : Total knee arthroplasty, Complications, Posterior stabilisation, Third midline condyle, Implant design


Plan


© 2016  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 102 - N° 8

P. 1061-1068 - décembre 2016 Retour au numéro
Article précédent Article précédent
  • Validation of a French patient-reported outcome measure for patello-femoral disorders: The Lille Patello-Femoral Score
  • S. Putman, F. Rémy, G. Pasquier, F. Gougeon, H. Migaud, A. Duhamel
| Article suivant Article suivant
  • Tibiofemoral syndesmosis injury treated by temporary screw fixation and ligament repair
  • S. Steinmetz, B. Puliero, D. Brinkert, N. Meyer, P. Adam, F. Bonnomet, M. Ehlinger

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.