Abstract

Background

Coronal alignment is an important factor for the function and longevity of total knee arthroplasty (TKA). Coronal bowing of the lower extremity is common among Asians and it may pose a risk for malalignment of the lower leg and malposition of component.

Hypothesis

We hypothesized that coronal bowing itself has a risk for malalignment of the lower leg and malposition of femoral/tibial components and that navigation TKA is beneficial for patients with coronal bowing. We investigated the incidence of femoral/tibial bowing in patients treated with TKA and compared the radiographic parameters between the navigation group and the conventional group. Additionally, the influence of coronal bowing on these radiographic parameters was investigated.

Materials and methods

We enrolled 35 patients with knee osteoarthritis and 70 bilateral simultaneous TKAs. The patients underwent TKA with the use of a computer tomography-free navigation in one knee and conventional TKA in the contralateral knee. Preoperative coronal bowing were measured, and the subjects were divided into 2 subgroups, i.e. the bowing group and the non-bowing group. Lateral bowing was expressed as plus (+) and medial bowing was expressed as minus (−). Various radiographic parameters, including coronal bowing, lower leg alignment, component position, and outliers were compared between the navigation group and the conventional group.

Results

Femoral bowing varied from −7.4° to 10.9° with an average of 3.0°. Tibial bowing varied from −4.1° to 4.6° with an average of 0.4°. The femoral component was placed more properly in the navigation group. Number of outlier regarding to the coronal femoral component angle to the femoral mechanical axis was 14 cases (37.8%) in the bowing group and 6 cases (18.2%) in the non-bowing group (P=0.04).

Discussion

In conclusion, coronal femoral bowing has an important effect on femoral bone cut in TKA. The navigated TKA was more consistent than conventional TKA in aiding proper alignments of femoral component.

Level of evidence

Level II, comparative prospective study.

Le texte complet de cet article est disponible en PDF.

Keywords : Total knee arthroplasty (TKA), Coronal bowing, Navigation, Alignment


Plan


© 2017  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 103 - N° 2

P. 251-256 - avril 2017 Retour au numéro
Article précédent Article précédent
  • Can three-dimensional patient-specific cutting guides be used to achieve optimal correction for high tibial osteotomy? Pilot study
  • M. Munier, M. Donnez, M. Ollivier, X. Flecher, P. Chabrand, J.-N. Argenson, S. Parratte
| Article suivant Article suivant
  • Periprosthetic fractures of the distal femur after total knee arthroplasty : Plate versus nail fixation
  • S.S. Wallace, D. Bechtold, A. Sassoon

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.