Radiation dose and magnification in pelvic X-ray: EOS™ imaging system versus plain radiographs - 23/11/17

Doi : 10.1016/j.otsr.2017.07.018 
P. Chiron, L. Demoulin, K. Wytrykowski, E. Cavaignac, N. Reina, J. Murgier
 Département d’orthopédie traumatologie, hôpital Pierre-Paul-Riquet, CHU de Toulouse, place du Docteur-Baylac, TSA 40031, 31059 Toulouse cedex 9, France 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Background

In plain pelvic X-ray, magnification makes measurement unreliable. The EOS™ (EOS Imaging, Paris France) imaging system is reputed to reproduce patient anatomy exactly, with a lower radiation dose. This, however, has not been assessed according to patient weight, although both magnification and irradiation are known to vary with weight. We therefore conducted a prospective comparative study, to compare: (1) image magnification and (2) radiation dose between the EOS imaging system and plain X-ray.

Hypothesis

The EOS imaging system reproduces patient anatomy exactly, regardless of weight, unlike plain X-ray.

Material and method

A single-center comparative study of plain pelvic X-ray and 2D EOS radiography was performed in 183 patients: 186 arthroplasties; 104 male, 81 female; mean age 61.3±13.7years (range, 24–87years). Magnification and radiation dose (dose-area product [DAP]) were compared between the two systems in 186 hips in patients with a mean body-mass index (BMI) of 27.1±5.3kg/m2 (range, 17.6–42.3kg/m2), including 7 with morbid obesity.

Results

Mean magnification was zero using the EOS system, regardless of patient weight, compared to 1.15±0.05 (range, 1–1.32) on plain X-ray (P<10−5). In patients with BMI<25, mean magnification on plain X-ray was 1.15±0.05 (range, 1–1.25) and, in patients with morbid obesity, 1.22±0.06 (range, 1.18–1.32). The mean radiation dose was 8.19±2.63dGy/cm2 (range, 1.77–14.24) with the EOS system, versus 19.38±12.37dGy/cm2 (range, 4.77–81.75) with plain X-ray (P<10−4). For BMI >40, mean radiation dose was 9.36±2.57dGy/cm2 (range, 7.4–14.2) with the EOS system, versus 44.76±22.21 (range, 25.2–81.7) with plain X-ray. Radiation dose increased by 0.20dGy with each extra BMI point for the EOS system, versus 0.74dGy for plain X-ray.

Conclusion

Magnification did not vary with patient weight using the EOS system, unlike plain X-ray, and radiation dose was 2.5-fold lower.

Level of evidence

3, prospective case-control study.

Le texte complet de cet article est disponible en PDF.

Keywords : EOS imaging system, Radiation dose, Magnification, Planning, Pelvic X-ray


Plan


© 2017  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 103 - N° 8

P. 1155-1159 - décembre 2017 Retour au numéro
Article précédent Article précédent
  • Outcomes of cup revision for ilio-psoas impingement after total hip arthroplasty: Retrospective study of 46 patients
  • C. Batailler, N. Bonin, A. Nogier, S. Martres, E. Ollier, O. May, S. Lustig, the French Arthroscopy Society (SFA) h
| Article suivant Article suivant
  • Prospective assessment of patients’ knowledge and informational needs and of surgeon-to-patient information transfer before and after knee or hip arthroplasty
  • L. Billon, B. Décaudin, G. Pasquier, A. Lons, V. Deken-Delannoy, A.-F. Germe, P. Odou, H. Migaud

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.