Pullout strength of a novel hybrid fixation technique (Tape Locking Screw™) in soft-tissue ACL reconstruction: A biomechanical study in human and porcine bone - 29/11/17

Doi : 10.1016/j.otsr.2017.01.006 
M. Ayzenberg , D. Arango, G.E. Gershkovich, P.S. Samuel, M. Saing
 Einstein Medical Center, 5501 Old York Road, Willowcrest Building 4th Floor, Philadelphia, PA 19141, United States 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Introduction

A novel hybrid anterior cruciate ligament (ACL) reconstruction technique known as Tape Locking Screw™ (TLS) is gaining popularity. Utilizing a suspension-type construct in conjunction with an interference screw, this technique has demonstrated successful initial clinical results with the use of quadruple hamstring graft. However, there is currently limited data available on the biomechanical strength of this fixation. This study investigates the pullout strength of the construct in human distal femora as well as in a porcine model. The construct is tested in isolation, without the use of any graft. We hypothesized that the pullout strength of this construct would be similar to or better than current fixation systems available.

Materials and methods

The Tape Locking Screw hybrid fixation system was implanted into twenty-two fresh frozen human distal femora (50–89 years old) randomized to 10×20mm titanium or polyether ether ketone (PEEK) screws by a single sports fellowship trained orthopedic surgeon. Given that the graft is secured to polyethylene terephthalate tape within the construct, the construct was implanted without any graft in order to isolate the device for biomechanical testing. After implantation, a tensile force was applied directly to the loop of tape at a loading rate of 5mm/min using an electromechanical testing system. The failure load was calculated from the resultant load–displacement curve. Specimens were then visually examined for mode of failure. Similar biomechanical tests were performed on sixteen porcine femora.

Results

In the human model, the mean pullout strength was 523±269N with the PEEK screw and 578±245N with the titanium screw. In the porcine femur model, mean strength was 616±177N with PEEK, 584±245N with titanium. There was no statistically significant difference in failure loads between these four groups. Tape slippage at the screw bone interface was the primary mode of failure in all the groups tested.

Discussion

Our results demonstrate that the hybrid technique provides excellent pullout strength in comparison to other soft-tissue ACL fixation methods, with tape slippage being the mode of failure in all specimens tested. This data, in addition to the advantages of the TLS system, support its consideration in the armamentarium of constructs available for soft-tissue ACL reconstruction.

Level of evidence

Laboratory controlled study level 2.

Le texte complet de cet article est disponible en PDF.

Keywords : Anterior cruciate ligament, Fixation, Pull out strength, Hamstring, Tape locking screw, TLS


Plan


© 2017  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 103 - N° 4

P. 591-595 - juin 2017 Retour au numéro
Article précédent Article précédent
  • Tourniquet and closed-suction drains in total knee arthroplasty. No beneficial effects on bleeding management and knee function at a higher cost
  • D. Yin, J. Delisle, A. Banica, A. Senay, P. Ranger, G.Y. Laflamme, J. Jun, J. Fernandes
| Article suivant Article suivant
  • Acute quadriceps tendon rupture repaired by suture anchors: Outcomes at 7 years’ follow-up in 25 cases
  • P. Brossard, G. Le Roux, B. Vasse, Orthopedics, Traumatology Society of Western France (SOO) b

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.