The Merete BioBall system in hip revision surgery: A systematic review - 23/11/18

Doi : 10.1016/j.otsr.2018.06.016 
Carlos D. Novoa a, , Mustafa Citak b, Akos Zahar b, Roxana E. López a, Thorsten Gehrke b, José L. Rodrigo a
a Orthopedics Surgery and Traumatology Department, Dr Peset University Hospital, avenue Gaspar Aguilar 90, 47017 Valencia, Spain 
b Department of Orthopaedic Surgery, Helios ENDO-Klinik Hamburg, Holstenstrasse 2, 22767 Hamburg, Germany 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Background

Modular hip arthroplasty is widely recognized in hip surgery; nevertheless, despite advances in prosthetic design, the incidence of hip revision surgery is increasing. In these procedures, a modular adapter that engages the femoral stem and the modular femoral head could be useful in order to protect the neck-head junction and restore preoperative biomechanics and soft tissue tension when the femoral or acetabular component should be retained. The Merete™ BioBall™ system (Merete Medical, Berlin, Germany) could fill this need. However the literature regarding this device is sparse, therefore we performed a systematic review to analyze: the indications for its use and clinical results and the causes of second revision and survival in hip revision surgery, mainly focus in complications of its use.

Materials and methods

Databases including PubMed and Google Scholar were searched for English-language articles published between 01/01/1999 and 01/04/2017, using search terms related to the system Merete™ BioBall™. Studies that reported the clinical use of the Merete™ BioBall™ system in hip revision surgery were selected. Data related to the indications, characteristics of the implant, clinical results, causes of second revision and survival were gathered.

Results

A total of 14 studies with a level IV evidence were analyzed. Information related to the implant was provided in 194 cases. Isolated acetabular revision was the main indication of implant use (139/194 [71.6%]). The most common combination used was the standard implant (69/110 specified [59.4%]), 3XL length (28/107 specified [26.2%]) with 32mm metal head (58/187 specified [31%]). After a follow-up ranging from 2 to 97 months, there were 16 complications, thirteen (81,2%) were revised, including one neck fracture of the stem (6.25%), two ceramic head fractures (12,5%) and six recurrent dislocations (37,5%). Implant survival was 92.8% at 52.5 months of follow-up in the longest published series. Clinical and functional results were provided in a heterogeneous way.

Conclusions

There is insufficient evidence to recommend for or contraindicate the use of the Merete™ BioBall™ system. The Isolated acetabular revision was the main indication of implant use. When dislocation is the indication of its use the rate of second revision is high. Some important complications possibly related with the implant design have been reported but as isolated cases. The neck adapter failure or corrosion phenomena have not been reported. Further prospective and controlled trials are needed to assess the use of this type of implant and its survival.

Level of evidence

Level III systematic review of case series.

Le texte complet de cet article est disponible en PDF.

Keywords : Total Hip Arthroplasty, Hip revision arthroplasty, Head-neck adapter, Merete Bioball, Systematic review


Plan


© 2018  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 104 - N° 8

P. 1171-1178 - décembre 2018 Retour au numéro
Article précédent Article précédent
  • Total hip arthroplasty after failed transtrochanteric rotational osteotomy for osteonecrosis of the femoral head: A systematic review and meta-analysis
  • Tomonori Shigemura, Yohei Yamamoto, Yasuaki Murata, Takashi Sato, Ryuto Tsuchiya, Norichika Mizuki, Yasunori Toki, Yuichi Wada
| Article suivant Article suivant
  • Are dual mobility cups associated with increased metal ions in the blood? Clinical study of nickel and chromium levels with 29 months’ follow-up
  • Laura Marie-Hardy, Padhraig O’Laughlin, Michel Bonnin, Tarik Ait Si Selmi

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.