Influence of iterative reconstruction and dose levels on metallic artifact reduction: A phantom study within four CT systems - 26/04/19

Doi : 10.1016/j.diii.2018.12.007 
J. Greffier , A. Larbi, J. Frandon, P.A. Daviau, J.P. Beregi, F. Pereira
 Department of Radiology, CHU Nîmes, Université Montpellier, Medical Imaging Group Nîmes, EA 2415, 30029 Nîmes cedex, France 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Purpose

To compare metallic artifact reduction (MAR) algorithms proposed by four vendors according to the delivered dose and iterative level using a phantom study.

Methods

Four CT systems (Revolution GSI®, Ingenuity Elite®, Somatom Edge®, and Aquilion Prime®) equipped with MAR algorithms (Smart MAR®, O-MAR®, iMAR®, and SEMAR®) were compared. Acquisitions were performed with CIRS model 062M Phantom containing a titanium rod core insert using 120kV and two dose levels (3 and 7mGy). Images were reconstructed with and without MAR algorithms using standard “soft tissue” kernel for filtered back projection (FBP) and intermediary iterative level. Artifact propagation was assessed by counting the number of pixels containing a HU outside a defined threshold interval (>100HU and<−80 HU). Artifact correction was evaluated by computing the differences between images with and without MAR.

Results

Accuracy of NCT values increased significantly using MAR algorithms, IR, and high dose levels (P<0.001). Image noise reduced −31±15 (SD) % (range: −50%;−14%) with Smart-MAR®, −28±3 (SD) % (range −31%; −25%) with O-MAR®, −32±7 (SD) % (range: −40%; −24%) with iMARCN®, −52±8 (SD) % (range: −60%; −42%) with iMARTH® and −29±6 (SD) % (range: −37%; −23%) with SEMAR®. The number of pixels outside the threshold interval was also reduced using MAR algorithms. Each MAR algorithm corrected in distinct patterns, with satisfactory artifact correction for all MAR algorithms.

Conclusion

This study demonstrates that artifact correction using MAR algorithms differs according to the main manufacturers, although corrections are satisfactory for all systems. Corrections also improved by using IR and increasing the dose level.

Le texte complet de cet article est disponible en PDF.

Keywords : Metallic artifact reduction (MAR), Computed tomography (CT), Image quality, Iterative reconstruction

Abbreviations : CT, FBP, iMAR®, IR, MAR, O-MAR®, ROI, SEMAR®, Smart-MAR®


Plan


© 2019  Société française de radiologie. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 100 - N° 5

P. 269-277 - mai 2019 Retour au numéro
Article précédent Article précédent
  • Sensorineural hearing loss in patients with vestibular schwannoma correlates with the presence of utricular hydrops as diagnosed on heavily T2-weighted MRI
  • M. Eliezer, G. Poillon, C. Maquet, A. Gillibert, J. Horion, J.-P. Marie, J.-P. Guichard, N. Magne, A. Attyé
| Article suivant Article suivant
  • Temporal evaluation of the microwave ablation zone and comparison of CT and gross sizes during the first month post-ablation in swine lung
  • H. Kodama, E. Ueshima, K. Howk, S.W. Lee, J.P. Erinjeri, S.B. Solomon, G. Srimathveeravalli

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.