Planification de la radiothérapie du cancer de la prostate par l’imagerie par résonance magnétique - 15/06/19

Radiotherapy treatment planning of prostate cancer using magnetic resonance imaging

Doi : 10.1016/j.canrad.2018.09.005 
L. Vanquin a, , C. Boydev a, b, J. Korhonen c, E. Rault a, F. Crop a, T. Lacornerie a, A. Wagner a, J. Laffarguette a, D. Pasquier d, e, N. Reynaert a, f
a Unité de physique médicale, centre Oscar-Lambret, 3, rue Combemale, 59020 Lille, France 
b Unité de physique médicale, institut Curie, 35, rue Dailly, 92210 Saint-Cloud, France 
c Department of radiation oncology, Comprehensive Cancer Center, Helsinki university central hospital, 00029 Helsinki, Finlande 
d Département universitaire de radiothérapie, centre Oscar-Lambret, 3, rue Combemale, 59020 Lille, France 
e CRIStAL UMR CNRS 9189, université Lille 1, M3, avenue Carl-Gauss, 59650 Villeneuve-d’Ascq, France 
f Institut Jules-Bordet, 121, boulevard de Waterloo, 1000 Bruxelles, Belgique 

Auteur correspondant. Physique médicale, centre Oscar-Lambret, 3, rue Frédéric-Combemale, 59020 Lille cedex, France.Physique médicale, centre Oscar-Lambret3, rue Frédéric-CombemaleLille cedex59020France

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

pages 9
Iconographies 5
Vidéos 0
Autres 0

Résumé

Objectif de l’étude

L’utilisation de l’imagerie par résonance magnétique (IRM) seule pour la planification dosimétrique de la radiothérapie externe soulève le problème de l’estimation de la densité électronique à partir des IRM. L’objectif de ce travail était de développer une méthode de création d’images tomodensitométriques (TDM) synthétiques (ou pseudo-TDM) à partir des IRM puis d’évaluer sa fiabilité quant à son utilisation pour la planification dosimétrique des traitements du cancer de la prostate par irradiation externe.

Matériel et méthodes

Les pseudo-TDM du pelvis ont été créés par déformations élastiques d’atlas sur dix patients qui devaient être traités pour un cancer de la prostate. Pour chaque patient, la planification dosimétrique était effectuée sur la TDM puis reprojetée sur la pseudo-TDM. Les différences dosimétriques entre pseudo-TDM et TDM ont ensuite été évaluées.

Résultats

Pour la dose moyenne au volume cible prévisionnel, la différence relative de dose entre pseudo-TDM et TDM allait de 0,1 % à 1,7 % ; elle était en moyenne de 0,9 %. Pour le rectum, les têtes fémorales droite et gauche et la vessie, elle était en moyenne de respectivement 1,1 %, 0,8 %, 0,8 % et 1,0 %.

Conclusion

Les écarts dosimétriques entre les TDM et les pseudo-TDM étaient faibles pour les organes à risques et pour le volume cible, laissant envisager une possible utilisation d’une méthode de génération de pseudo-TDM de ce type pour la planification en radiothérapie à partir de l’IRM.

Le texte complet de cet article est disponible en PDF.

Abstract

Purpose

Magnetic resonance imaging (MRI) plays an increasing role in radiotherapy dose planning. Indeed, MRI offers superior soft tissue contrast compared to computerized tomography (CT) and therefore could provide a better delineation of target volumes and organs at risk than CT for radiotherapy. Furthermore, an MRI-only radiotherapy workflow would suppress registration errors inherent to the registration of MRI with CT. However, the estimation of the electronic density of tissues using MRI images is still a challenging issue. The purpose of this work was to design and evaluate a pseudo-CT generation method for prostate cancer treatments.

Materials and methods

A pseudo-CT was generated for ten prostate cancer patients using an elastic deformation based method. For each patient, dose delivered to the patient was calculated using both the planning CT and the pseudo-CT. Dose differences between CT and pseudo-CT were investigated.

Results

Mean dose relative difference in the planning target volume is 0.9% on average and ranges from 0.1% to 1.7%. In organs at risks, this value is 1.8%, 0.8%, 0.8% and 1% on average in the rectum, the right and left femoral heads, and the bladder respectively.

Conclusion

The dose calculated using the pseudo-CT is very close to the dose calculated using the CT for both organs at risk and PTV. These results confirm that pseudo-CT images generated using the proposed method could be used to calculate radiotherapy treatment doses on MRI images.

Le texte complet de cet article est disponible en PDF.

Mots clés : Recalage élastique, Radiothérapie, IRM, Pseudo-CT (Pseudo-Computerized Tomography)

Keywords : Elastic registration, Radiotherapy, MRI, Pseudo-CT (Pseudo-Computerized Tomography)


Plan


© 2019  Société française de radiothérapie oncologique (SFRO). Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 23 - N° 4

P. 281-289 - juillet 2019 Retour au numéro
Article précédent Article précédent
  • Toxicité de la chimiothérapie d’induction par docétaxel, platine, 5-fluorouracile (TPF) pour les cancers des voies aérodigestives supérieures localement évolués en routine clinique : importance du statut nutritionnel
  • M. Bernadach, M. Lapeyre, A.F. Dillies, J. Miroir, J. Moreau, F. Kwiatkowski, N. Pham-Dang, N. Saroul, X. Durando, J. Biau
| Article suivant Article suivant
  • Méningiomes de la base du crâne : efficacité clinique et radiologique basée sur une analyse volumétrique quantitative
  • Y. Brahimi, D. Antoni, R. Srour, P. Wagner, F. Proust, A. Thiery, A. Labani, G. Noël

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’achat d’article à l’unité est indisponible à l’heure actuelle.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.