Accuracy of 4D cardiac magnetic resonance for flow measurement - 06/01/20

Doi : 10.1016/j.acvdsp.2019.09.129 
M. Isorni , J. Guihaire, L. Moisson, A. Sigal, O. Planche, S. Hascoet, P. Brenot
 Hôpital Marie-Lannelongue, Le Plessis Robinson, France 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Résumé

Background

4D Flow CMR has emerged as an innovative imaging modality providing flow assessment at any selected region of interest. The wide range of dynamic properties and velocity of cardiac flows are a commonly admitted technical limit. We aim to assess if this technique is robust enough to validate the conservation of mass principle applied to vessels and valves of the whole heart. Our working hypothesis was that 4D CMR would prove accurate compared to 2D CMR.

Methods

Thirty healthy patients underwent both 2D CMR and 4D CMR using a 1.5-T machine. Ventricular volumes were assessed and cardiac output was calculated and used as a reference value. We mapped using 4D Flow analysis (Arterys), cardiac flows from the vena cava to the aorta. The 4D flows through the cardiac valves, and vessels were independently measured by two readers. Agreement between flows measured using 4D CMR and CMR was assessed by linear regression analysis and Bland-Altman plots.

Results

Mean net flows were 5.5±1.2L/min through venae cavae, 5.4±1.1L/min through the tricuspid valve, 5.5±1.1L/min through the pulmonary valve, 5.4±1.4L/min through the pulmonary artery, 5.3±1.1L/min through the pulmonary veins, 5.5±1.0L/min through the mitral valve, 5.5±1.4L/min through the aortic valve and 5.5±1.4L/min through the aorta compared to 5.6±1.1 with 2D CMR. Correlation coefficients between 4D flows measurements and 2D CMR ranged from 0.60 to 0.86. Agreement was strong between cardiac outputs with a mean bias ranging from 0.01 to 0.33 with no significant bias. Interobserver reproducibility of 4D CMR net flow measurements was good (interclass coefficient ranging from 0.89–0.98; and coefficient of variation of 3.7%–6.3%).

Conclusion

4D CMR flow measurements correlated with cardiac output measured using CMR in healthy volunteers in all valves and vessels, respecting the conservation of mass principle. 4D CMR seems robust enough to obtain precise hemodynamic features of cardiac flow.

Le texte complet de cet article est disponible en PDF.

Plan


© 2019  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 12 - N° 1

P. 60-61 - janvier 2020 Retour au numéro
Article précédent Article précédent
  • The parasternal long axis ultrasound view does not allow the maximum diameter of the aortic root to be measured in atypical BAV
  • P. Masi, O. Milleron, J.F. Paul, F. Arnoult, N. Ould Ouali, E. Lansac, R. Raffoul, P. Ou, G. Jondeau
| Article suivant Article suivant
  • Feasibility and prognostic value of vasodilator stress perfusion Cardiac Magnetic Resonance in elderly patients > 75 years without known CAD
  • T. Pezel, M. Kinnel, P. Garot, T. Unterseeh, S. Champagne, Y. Louvard, M.C. Morice, F. Sanguineti, J. Garot

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.