S'abonner

Joint Position Paper of the Working Group of Pacing and Electrophysiology of the French Society of Cardiology (SFC) and the Société française d’imagerie cardiaque et vasculaire diagnostique et interventionnelle (SFICV) on magnetic resonance imaging in patients with cardiac electronic implantable devices - 22/02/20

Doi : 10.1016/j.diii.2020.02.003 
J.-N. Dacher a, , 1 , E. Gandjbakhch b, 1, J. Taieb c, M. Chauvin d, F. Anselme e, A. Bartoli f, L. Boyer g, L. Cassagnes g, H. Cochet h, B. Dubourg a, L. Fauchier i, D. Gras j, D. Klug k, G. Laurent l, J. Mansourati m, E. Marijon n, P. Maury o, O. Piot p, F. Pontana q, F. Sacher r, N. Sadoul s, S. Boveda t, A. Jacquier f

Working Group of Pacing, Electrophysiology of the French Society of Cardiology, Société française d’imagerie cardiaque et vasculaire diagnostique et interventionnelle (SFICV)

a Normandie UNIV, UNIROUEN, Inserm U1096, CHU Rouen, Department of Radiology, Cardiac Imaging Unit, 76000 Rouen, France 
b Sorbonne Universités, AP–HP, Heart Institute, La Pitié-Salpêtrière University Hospital, 75013 Paris, France 
c Hospital of Aix-en-Provence, Department of Cardiology, 13100 Aix-en-Provence, France 
d Université de Strasbourg, CHU Strasbourg, Department of Cardiology, 67000 Strasbourg, France 
e Normandie UNIV, UNIROUEN, CHU Rouen, Department of Cardiology, 76000 Rouen, France 
f Université Aix-Marseille, Centre Hospitalo-Universitaire Timone, AP–HM, Department of Radiology, CNRS, CRMBM, CEMEREM, 13005 Marseille, France 
g Université Clermont Auvergne, CHU Clermont-Ferrand, Department of Radiology, 63000 Clermont-Ferrand, France 
h Université de Bordeaux-Inserm, IHU LIRYC, CHU de Bordeaux, Department of Cardiovascular Imaging, Hôpital Cardiologique du Haut-Lévêque, 33600 Pessac, France 
i Université de Tours, CHU de Tours, Department of Cardiology, 37000 Tours, France 
j Nouvelles Cliniques Nantaises, Department of Cardiology, 44200 Nantes, France 
k Université de Lille, CHRU de Lille, Department of Cardiology, 59000 Lille, France 
l Université de Dijon, CHU de Dijon, Department of Cardiology, 21000 Dijon, France 
m Université de Bretagne Occidentale, CHU de Brest, Department of Cardiology, 29200 Brest, France 
n Université de Paris, AP–HP, Department of Cardiology, Georges-Pompidou European University Hospital, 75015 Paris, France 
o Université de Toulouse, Inserm U1048, Department of Cardiology, Hospital Rangueil, 31059 Toulouse, France 
p Centre Cardiologique du Nord, Department of Cardiology, 93200 Saint-Denis, France 
q Université de Lille, Inserm U1011, Department of Cardiovascular Radiology, Institut Cœur-Poumon, 59000 Lille, France 
r Université de Bordeaux-Inserm, IHU LIRYC, CHU de Bordeaux, Department of Cardiology, Hôpital Cardiologique du Haut-Lévêque, CHU de Bordeaux, 33600 Pessac, France 
s Université de Nancy Lorraine, CHU de Nancy, Department of Cardiology, 54511 Vandœuvre-lès-Nancy, France 
t Clinique Pasteur, Department of Cardiology, 31076 Toulouse, France 

Corresponding author. Department of Radiology, CHU Rouen, Hôpital Charles-Nicolle, 1, boulevard Gambetta, 76000 Rouen, France.Department of Radiology, CHU Rouen, Hôpital Charles-Nicolle1, boulevard GambettaRouen76000France
Sous presse. Épreuves corrigées par l'auteur. Disponible en ligne depuis le Saturday 22 February 2020
Cet article a été publié dans un numéro de la revue, cliquez ici pour y accéder

Highlights

This position paper gives the common position of the Working Group of Pacing and Electrophysiology of the French Society of Cardiology (SFC) and the Société française d’imagerie cardiaque et vasculaire diagnostique et interventionnelle (SFICV) on the technical conditions of MRI in patients with MR-conditional and -nonconditional CEID that could serve as a basis for institutional MRI protocols in patients with a CEID.
MRI-conditional materials are not “MRI-safe” hence require specific device programing and patient monitoring.
For all patients carrying a MRI-conditional or nonconditional cardiac electronic implantable device (CEID), MRI should be integrated into a standardized workflow defined in an institutional protocol involving both radiologists and device specialists.

Le texte complet de cet article est disponible en PDF.

Abstract

Magnetic resonance imaging (MRI) has become the reference imaging for the management of a large number of diseases. The number of MR examinations increases every year, simultaneously with the number of patients receiving a cardiac electronic implantable device (CEID). A CEID was considered an absolute contraindication for MRI for years. The progressive replacement of conventional pacemakers and defibrillators by MR-conditional CEIDs and recent data on the safety of MRI in patients with “MR-nonconditional” CEIDs have progressively increased the demand for MRI in patients with a CEID. However, some risks are associated with MRI in CEID carriers, even with “MR-conditional” devices because these devices are not “MR-safe”. A specific programing of the device in “MR-mode” and monitoring patients during MRI remain mandatory for all patients with a CEID. A standardized patient workflow based on an institutional protocol should be established in each institution performing such examinations. This joint position paper of the Working Group of Pacing and Electrophysiology of the French Society of Cardiology and the Société française d’imagerie cardiaque et vasculaire diagnostique et interventionnelle (SFICV) describes the effect and risks associated with MRI in CEID carriers. We propose recommendations for patient workflow and monitoring and CEID programming in MR-conditional, “MR-conditional nonguaranteed” and MR-nonconditional devices.

Le texte complet de cet article est disponible en PDF.

Keywords : Magnetic resonance imaging (MRI), Safety, Cardiac pacing, Defibrillators, Pacemaker


Plan


© 2020  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.