S'abonner

Accuracy of liver ablation zone prediction in a single 2450 MHz 100 Watt generator model microwave ablation system: An in human study - 25/03/20

Doi : 10.1016/j.diii.2019.10.007 
S. Young , M. Rivard, R. Kimyon, T. Sanghvi
 Department of Radiology, University of Minnesota, Division of Interventional Radiology, 420 Delaware St SE, Minneapolis, MN 55455, USA 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Purpose

To compare manufacturer provided predictions and realized ablation dimensions in the liver using one 2450MHz 100 Watt generator model microwave ablation (MWA) system.

Materials and method

Between 1/1/2015 and 2/1/2018, MWAs were performed in 86 patients who underwent a total of 103 MWAs with a single MWA system. There were 64 men and 22 women with a mean age of 63.9±9.9 (SD) years (range: 30–88 years). Demographic, procedural, and outcomes data was recorded. The manufacturer predicted ablation zone sizes in three dimensions (anterior-posterior [AP], transverse [TR], and cranial caudal [CC]) were recorded and then compared to the actual ablation zone sizes at one month follow-up imaging.

Results

MWAs were most commonly performed to treat hepatocellular carcinoma (92/103, 89.3%). Dividing the actual ablation size by the manufacturer prediction in the AP, TR, and CC directions resulted in a mean of 88.3±20.6 (SD) % (range: 33.3–156.4%), 80.2±26.5 (SD) % (range: 29.6–182.9%), and 86.7±25.1 (SD) % (range: 37–186.1%), respectively. The realized AP direction was statistically closer to the manufacturer prediction than the TR (P<0.01). Ablation Watt setting of 100 Watts resulted in more accurate predictions than the 75 or 45 Watt settings in the AP direction (P=0.03).

Conclusions

This 2450MHz 100 Watt generator MWA system manufacturer provided model fairly accurately predicts ablation zone dimensions, but tends to over predict realized dimensions in this mainly hepatocellular carcinoma, and therefore cirrhotic, cohort. The TR is the most inaccurately predicted dimension and manufacturer predictions appear to be best in the 100W setting, important aspects for interventionalists to consider during ablation planning and execution.

Le texte complet de cet article est disponible en PDF.

Keywords : Microwave ablation, Hepatocellular carcinoma, Metastatic liver lesions, Liver neoplasm, Thermal ablation


Plan


© 2019  Société française de radiologie. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 101 - N° 4

P. 225-233 - avril 2020 Retour au numéro
Article précédent Article précédent
  • Detection of secondary causes of spontaneous pneumothorax: Comparison between computed tomography and chest X-ray
  • A.M. Ruppert, D. Sroussi, A. Khallil, M. Giot, J. Assouad, J. Cadranel, V. Gounant
| Article suivant Article suivant
  • MRI combined with PSA density in detecting clinically significant prostate cancer in patients with PSA serum levels of 4?10 ng/mL: Biparametric versus multiparametric MRI
  • C. Han, S. Liu, X.B. Qin, S. Ma, L.N. Zhu, X.Y. Wang

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.