Assessment of stiffness and load deflection of orthodontic miniscrews used for palatal anchorage: An in vitro biomechanical study - 11/12/20

Doi : 10.1016/j.ortho.2020.08.005 
Anna Colonna 1, , Margherita Drudi 2, Giorgio Alfredo Spedicato 3, Francesco Mollica 4, Valentina Mazzanti 4, Emanuele Paoletto 5, Giuliano Maino 6, Giuseppe Siciliani 7, Luca Lombardo 7
1 University of Ferrara, Postgraduate school of orthodontics, Ferrara, Italy 
2 Private practice, Ferrara, Italy 
3 Unipol Group, Bologna, Italy 
4 University of Ferrara, Department of Engineering, Ferrara, Italy 
5 Lab Orthomodul, Thiene, Italy 
6 Ferrara University and Insubria University, Postgraduate school of orthodontics, Private practice in Vicenza, Vicenza, Italy 
7 Postgraduate school of orthodontics. University of Ferrara, Ferrara, Italy 

Anna Colonna, University of Ferrara, Postgraduate school of orthodontics, Via Luigi Borsari 46, Ferrara, Italy.University of Ferrara, Postgraduate school of orthodonticsVia Luigi Borsari 46FerraraItaly

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

pages 11
Iconographies 6
Vidéos 0
Autres 0

Summary

Objective

The purpose of this study was to evaluate the biomechanical properties of miniscrews of 5 different lengths, 2 different diameters, and different combinations of insertion used for palatal skeletal anchorage.

Materials and methods

Twenty-four different combinations of a total of 120 miniscrews of two different diameters (2.0mm and 2.3mm) and five different lengths (9mm, 11mm, 13mm and 15mm) were tested at different angles of insertion (90° and 45°) and distances from a synthetic bone block (3mm, 5mm, 7mm). Samples were fixed in an Instron Universal Testing Machine and a load was applied in single cantilever mode to the neck of each miniscrew. The stiffness and maximum load before permanent deformation were recorded. Model-based recursive partitioning testing (CART) was used to evaluate differences between groups.

Results

Significantly higher forces were necessary to deform miniscrews of diameter 2.3mm than those of 2.0mm, those inserted at an angle of 45° with respect to 90°, and at smaller distances between the miniscrew neck and block; in addition, the maximum load and stiffness increased with increasing screw length.

Conclusion

This in vitro experimental study showed strong correlations between deformation load and miniscrew geometry, insertion angle and distance from the synthetic block, results that should be considered when planning miniscrew insertion in order to reduce the risk of unwanted fracture.

Le texte complet de cet article est disponible en PDF.

Keywords : Miniscrews, Anchorage, Palatal anchorage, Load, Temporary anchorage device, In vitro study


Plan


© 2020  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 18 - N° 4

P. 809-819 - décembre 2020 Retour au numéro
Article précédent Article précédent
  • Evaluation of the force degradation and deformation of the open-closed and open springs of NiTi: An in vitro study
  • Thiago Prado, José Guilherme Neves, Lourenço Correr-Sobrinho, Carolina Carmo Menezes, Giovana Cherubini Venezian, Américo Bortolazzo Correr, Ana Rosa Costa
| Article suivant Article suivant
  • How sandblasting on lingual surfaces can be carried out with minimum enamel damage: An in vitro study on human teeth
  • Ted Lundgren, Alva Samuelson, Cajsa Clase, Julia Naoumova

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’achat d’article à l’unité est indisponible à l’heure actuelle.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.