S'abonner

Iterative denoising accelerated 3D SPACE FLAIR sequence for brain MR imaging at 3T - 04/01/22

Doi : 10.1016/j.diii.2021.09.004 
Michael Eliezer a, b, , Alexis Vaussy c, Solenn Toupin c, Rémy Barbe a, Stephan Kannengiesser d, Alto Stemmer d, Emmanuel Houdart a, b
a Department of Neuroradiology, Lariboisiere University Hospital, 75010 Paris, France 
b Université de Paris, Faculté de Médecine, 75010 Paris, France 
c Siemens Healthineers France, 93210 Saint-Denis, France 
d Siemens Healthineers, 91052 Erlangen, Germany 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Highlights

A 37% scanning time reduction is obtained with 3D FLAIR while preserving image quality for brain MRI using iterative denoising (ID).
Higher degrees of interobserver agreement for image quality ,are obtained with accelerated FLAIR with ID compared to accelerated FLAIR without ID.
The use of ID reconstruction results in greater SNR and CNR with accelerated 3D FLAIR.

Le texte complet de cet article est disponible en PDF.

Abstract

Purpose

The purpose of this study was to prospectively evaluate image quality of three-dimensional fluid attenuated inversion recovery (3D-FLAIR) sequence acquired with a high acceleration factor and reconstructed with iterative denoising (ID) for brain magnetic resonance imaging (MRI) at 3-T.

Material and methods

Patients with brain tumor who underwent brain MRI were consecutively included. Two 3D-FLAIR sequences were successively performed for each patient. A first conventional FLAIR acquisition (conv-FLAIR) was performed with an acceleration factor of 6. The second acquisition was performed with an increased acceleration factor of 9. Two series one without ID (acc-FLAIR) and one with ID (acc-FLAIR-ID) were reconstructed. Two neuroradiologists independently assessed image quality, deep brain nuclei visualization and white matter/gray matter (WM/GM) differentiation on a 4-point scale.

Results

Thirty patients with brain tumor were consecutively included in this study. There were 16 women and 14 men with a mean age of 54 ± 17 (SD) years (range: 22–78 years). Scanning time of Acc-FLAIR-ID and Acc-FLAIR (4 min 40 sec) was 37% shorter than that of conv-FLAIR (2 min 50 sec) (P < 0.01). Improved image quality score was significantly different for both conv-FLAIR and acc-FLAIR-ID compared to acc-FLAIR (P < 0.01 for both). WM/GM differentiation score of conv-FLAIR was not significantly different compared to acc-FLAIR-ID (P = 0.10). Improved WM/GM differentiation score was different for both sequences compared to acc-FLAIR (P = 0.017 and P < 0.001). Deep brain nuclei visualization score was not different between conv-FLAIR and acc-FLAIR-ID (P = 0.71). However, the improved deep brain nuclei visualization score was significantly different for both sequences compared to acc-FLAIR (P < 0.001 for both).

Conclusion

Scanning time of 3D-FLAIR sequence using a high acceleration factor reconstructed with ID algorithm can be reduced by 37% while preserving image quality for brain MRI.

Le texte complet de cet article est disponible en PDF.

Keywords : Magnetic resonance imaging (MRI), 3D FLAIR, Iterative denoising, Brain MRI

Abbreviations : 3D, FLAIR, acc-FLAIR, acc-FLAIR-ID, CAIPIRINHA, CI, CNR, conv-FLAIR, CS, GRAPPA, ID, MPRAGE, MRI, PI, SD, SENSE, SNR, SPACE, STROBE, VIBE, WM/GM, TSE


Plan


© 2021  Société française de radiologie. Publié par Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 103 - N° 1

P. 13-20 - janvier 2022 Retour au numéro
Article précédent Article précédent
  • Hemorrhoid embolization: A review of current evidences
  • Reza Talaie, Pooya Torkian, Arash Dooghaie Moghadam, Farouk Tradi, Vincent Vidal, Marc Sapoval, Jafar Golzarian
| Article suivant Article suivant
  • Comparison of two deep learning image reconstruction algorithms in chest CT images: A task-based image quality assessment on phantom data
  • Joël Greffier, Julien Frandon, Salim Si-Mohamed, Djamel Dabli, Aymeric Hamard, Asmaa Belaouni, Philippe Akessoul, Francis Besse, Boris Guiu, Jean-Paul Beregi

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.