Évaluation de l’exposition au champ magnétique statique des manipulateurs en électroradiologie médicale d’un centre de recherche travaillant à proximité d’IRM 3T et 7T à l’aide d’un dispositif portable - 25/05/23

Assessment of the occupational exposure to the static magnetic field of radiographers from a research facility working in the vicinity of 3T and 7T MRI scanners with a portable magnetic monitoring device

Doi : 10.1016/j.admp.2022.101702 
M. Gimbert a, , M. Doyen b, c, N. Weber b, A. Delmas b, d, A. Vignaud e, I. Fabre f, C. Ginisty g, Y. Lecomte g, V. Berland g, S. Becuwe Desmidt h, S. Roger g, J. Felblinger b, d
a CHRU de Nancy, université de Lorraine, Service de santé au travail, 54000 Nancy, France 
b IADI, INSERM-U1254, université de Lorraine, CHRU de Nancy, 54000 Nancy, France 
c CHRU de Nancy, Nancyclotep Imaging Platform, 54000 Nancy, France 
d CIC-IT 1433, 54000 Nancy, France 
e CEA, NeuroSpin, université de Paris-Saclay, CNRS BAOBAB, 91190 Gif-Sur-Yvette, France 
f CEA, Service de santé au travail, 91190 Gif-Sur-Yvette, France 
g CEA, NeuroSpin, Université de Paris-Saclay, Unité INSERM UNIACT, 91190 Gif-Sur-Yvette, France 
h CEA, NeuroSpin, Université de Paris-Saclay, Unité INSERM UNICOG, 91190 Gif-Sur-Yvette, France 

Auteur correspondant. Service de santé au travail, CHRU de Nancy, rue du Morvan, 54000 Nancy, FranceService de santé au travail, CIC-IT 1433, IADI, Unité INSERM U1254, Université de Lorraine, CHRU de Nancyrue du MorvanNancy54000France

Résumé

Introduction

Un appareil d’Imagerie par Résonance Magnétique (IRM) émet de façon permanente un Champ Magnétique Statique (CMS), B0. La Directive Européenne 2013/35/UE, retranscrite dans le droit français avec le décret du 3 aout 2016 restreint l’exposition des salariés au CMS à 2T. Elle permet cependant aux salariés formés qui travaillent à proximité d’une IRM de dépasser ce seuil. Un salarié qui se déplace dans un CMS génère un gradient de champ magnétique, dB/dt. L’International Commission on Non-Ionizing Radiation Protection (ICNIRP) recommande de ne pas atteindre une exposition à dB/dt>2,7T.s−1 ou ΔB/3sec>2T afin d’éviter la survenue de symptômes transitoires tels que des vertiges.

Méthode

Entre 2016 et 2020, 5 manipulateurs en électroradiologie médicale (MERM) travaillant à proximité d’IRM 3T et 7T au sein du département de neuro-imagerie d’un centre de recherche ont porté un dispositif portable de mesure de l’exposition au CMS développé à Nancy. Les enregistrements ont été analysés afin de quantifier le nombre d’évènements correspondant à une exposition à B0>2T, ainsi que le nombre de dB/dt>2,7T.s−1 et de ΔB/3sec>2T. Le temps moyen passé par chaque MERM parmi les 5 catégories d’exposition suivantes a également été évalué : [0,1T–0,5T[ ; [0,5T–1T[ ; [1T-1,5T[ ; [1,5T–2T[ ;>2T.

Résultats

Au total, 966 journées de port de l’exposimètre ont été évaluées correspondant à 4839,8heures d’enregistrement dont environ 231heures où l’exposition était supérieure à 0,1T. 111 évènements où l’exposition à B0 était supérieure à 2T ont été identifiés. Un MERM a totalisé 80 évènements et a obtenu le B0 le plus élevé de 2,67T. Seul un évènement était précédé d’un dB/dt>2,7T.s−1. 14 évènements étaient suivis d’un ΔB/3sec>2T.

Conclusion

Ce dispositif pourrait aider les salariés qui travaillent à proximité d’IRM à contrôler leurs mouvements pour respecter les limites d’exposition recommandées par la Directive Européenne et ne pas expérimenter de symptômes au cours de leur travail.

Le texte complet de cet article est disponible en PDF.

Summary

Introduction

Magnetic Resonance Imaging (MRI) scanners permanently produce a Static Magnetic Field (SMF), B0. The European Directive 2013/35/EU, transposed into French law with the decree of August 3rd, 2016 restricts workers’ exposure regarding the SMF to 2T. It does however allow trained employees working in the vicinity of MRI scanners to exceed this threshold. A worker moving in the SMF generates a motion-induced time-varying magnetic field, dB/dt. The International Commission on Non-Ionizing Radiation Protection (ICNIRP) recommends not to reach an exposition to dB/dt>2.7T.s−1 or ΔB/3sec>2T to avoid transient symptoms such as vertigo.

Method

Between 2016 and 2020, 5 radiographers working in the vicinity of a 3T and a 7T MRI scanners from the neuro-imaging department in an MRI research facility wore a portable magnetic monitoring device developed in Nancy. All recorded data were analyzed in order to quantify the number of events corresponding to an exposure to B0>2T, and the number of dB/dt>2.7T.s−1 and ΔB/3sec>2T. The average time spent by each radiographers for the 5 following exposure categories was also assessed: [0.1T–0.5T[ ; [0.5T–1T[ ; [1T–1.5T[ ; [1.5T–2T[ ; >2T.

Results

In total, 966 shifts carrying the device were analyzed, corresponding to 4839.8hours of recording, including approximately 231hours with an exposure greater than 0.1T. 111 events with an exposure to B0>2T were identified. A radiographer totalized 80 events and was exposed to the higher B0 of 2,67T. Only one event was preceded by a dB/dt>2.7T.s−1. 14 events were followed by a ΔB/3sec>2T.

Conclusion

This device might help MRI workers control their movements in order to comply with the exposure threshold recommended by the European Directive and not experience symptoms during their work.

Le texte complet de cet article est disponible en PDF.

Mots clés : Imagerie par Résonance Magnétique (IRM), Exposition professionnelle, Champs Électromagnétiques, Évaluation des risques, France

Keywords : Magnetic Resonance Imaging (MRI), Occupational exposure, Electromagnetic fields, Risk assessment, France


Plan


© 2022  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 84 - N° 3

Article 101702- juin 2023 Retour au numéro
Article précédent Article précédent
  • Editorial board
| Article suivant Article suivant
  • Santé mentale à l’hôpital : étude transversale des représentations des professionnels d’un CHU français
  • L. Nohales, P. Leblanc, L. Rodriguez-Borlado-Salazar, N. Guibert, M. Ottavy, F. Subtil, P. Michel, J.B. Fassier

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.