Comparaison dosimétrique de la radiothérapie conformationnelle, la radiothérapie conformationnelle avec modulation d’intensité, la radiothérapie conformationnelle en conditions stéréotaxiques et la radiothérapie en conditions stéréotaxiques robotisée des tumeurs cérébrales bénignes - 07/07/11

Doi : 10.1016/j.canrad.2011.02.004 
E. Spasic a, , b, c, d , I. Buchheit a, V. Bernier a, A. Noël a, b, c, d
a Département de radiophysique, centre Alexis-Vautrin, avenue de Bourgogne, 54511 Vandœuvre-lès-Nancy cedex, France 
b UMR 7039 CNRS, centre de recherche en automatique de Nancy (Cran), BP 239, 54506 Vandœuvre-lès-Nancy cedex, France 
c Cran UMR 7039, faculté des sciences et techniques, université Henri-Poincaré Nancy 1, BP 239, 54506 Vandœuvre-lès-Nancy cedex, France 
d Cran UMR 7039, institut national polytechnique de Lorraine, BP 239, 54506 Vandœuvre-lès-Nancy cedex, France 

Auteur correspondant.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

pages 7
Iconographies 1
Vidéos 0
Autres 0

Résumé

Objectif de l’étude

Effectuer une comparaison entre la radiothérapie conformationnelle tridimensionnelle, la radiothérapie conformationnelle avec modulation d’intensité (RCMI), la radiothérapie conformationnelle en conditions stéréotaxiques et la radiothérapie en conditions stéréotaxiques robotisée par CyberKnife® afin de déterminer la ou les techniques les plus adaptées à un type de tumeur selon sa localisation, sa forme et sa taille.

Patients et méthodes

Les données anatomiques et dosimétriques (indice de conformation, volume d’isodose) de cinq patients récemment traités par irradiation conformationnelle tridimensionnelle, par irradiation conformationnelle avec modulation d’intensité ou par CyberKnife®ont servi de référence. Ces patients étaient atteints d’un méningiome du sinus caverneux, d’une tumeur de l’hypophyse, d’un craniopharyngiome ou d’un neurinome. Pour chacun des traitements, la dosimétrie de référence réellement délivrée a été comparée aux dosimétries obtenues avec les trois autres techniques. En comparant des traitements basés sur des caractéristiques différentes, différents paramètres ont également été évalués (radiobiologiques, de positionnement…).

Résultats

L’analyse des paramètres dosimétriques montrait que le choix du traitement semblait s’effectuer principalement en fonction de la taille de la tumeur, de sa forme et de sa proximité aux organes à risques et non de sa localisation. Pour les paramètres radiobiologiques, elle n’a pas permis d’observer d’écarts significatifs permettant de choisir une technique plutôt qu’une autre.

Conclusions

En ce qui concerne des tumeurs cérébrales bénignes de taille moyenne ou volumineuse et quelle que soit la forme, la radiothérapie conformationnelle avec modulation d’intensité semblait le traitement le plus adapté. Elle permettait d’obtenir un bon compromis entre la dose reçue par les volumes cibles et celle reçue par les tissus sains pour des tumeurs qui étaient à proximité d’organes à risques. Pour des lésions de petite taille, l’irradiation par CyberKnife® était le traitement de choix.

Le texte complet de cet article est disponible en PDF.

Abstract

Purpose

To compare several techniques in order to determine the best treatment for benign brain tumours.

Methods and patients

A retrospective study was performed for five patients who received 3D-conformal radiotherapy, intensity modulated radiotherapy or CyberKnife®. These patients had a meningioma, a pituitary tumour, a craniopharyngioma or a neurinoma. In each case, these treatment plans were optimised and compared with the three other dosimetries. Radiobiological or positioning parameters were evaluated, as well as dosimetric parameters, in order to compare treatments with different characteristics.

Results

The dosimetric parameters showed that the choice of treatment seemed to be determined mostly by tumour size, shape and proximity with organs at risk (not tumour localisation). Whereas the results showed no significant deviations with regards to the radiobiological parameters. Therefore, with these parameters, it was difficult to give priority to a treatment.

Conclusions

With regards to benign brain tumours of medium or large size, intensity modulated radiotherapy seemed the recommended treatment. It enabled to obtain a good ratio between efficacy and toxicity for tumours that are really close to organs at risk. Concerning small benign brain tumours, the CyberKnife® was probably the best treatment.

Le texte complet de cet article est disponible en PDF.

Mots clés : Comparaison dosimétrique, Tumeurs cérébrales bénignes, CyberKnife®, RCMI

Keywords : Dosimetry comparison, Benign brain tumours, CyberKnife®, IMRT


Plan


© 2011  Publié par Elsevier Masson SAS.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 15 - N° 4

P. 287-293 - juillet 2011 Retour au numéro
Article précédent Article précédent
  • Facteurs prédictifs de réponse à la radiochimiothérapie néoaduvante dans les cancers rectaux localement évolués et corrélation de ces facteurs avec la survie
  • B. Farnault, L. Moureau-Zabotto, C. de Chaisemartin, B. Esterni, B. Lelong, F. Viret, M. Giovannini, G. Monges, J.-R. Delpero, E. Bories, O. Turrini, P. Viens, M. Resbeut
| Article suivant Article suivant
  • Radiosensibilité, radiocurabilité et réparation
  • G. Vogin

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’achat d’article à l’unité est indisponible à l’heure actuelle.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.