P3: Hair and whole blood matrix effect in identification of a model analyte and determination of its limit of detection - 28/06/14

Doi : 10.1016/S2352-0078(14)70064-2 
W. Lechowicz
 Institute of forensic research in Cracow, Poland 

Résumé

Introduction

Any kind of biological material matrix may represent a serious problem during real samples analyses, especially using LCMS method. It causes both signal intensity loss and/or signal to noise ratio (S/N) decrease. Unfortunately, this phenomena touches selected ions sometime in a different way. It means that even deuterated standards in some cases cannot protect from precision disturbance or even substance misidentification.

Aim

The study aim was to elaborate a control method of a hair or whole blood matrix influence on a limit of detection (LOD) in identification and quantitation aspects.

Methods

Waters Alliance 2695 Separations Module with Quattro Micro Micromass were applied in the research. Analyses were performed using LiChroCART 55x4 Purospher RP-18 column from Merck as well as by direct injection without separation in flow injection analysis (FIA). The mobile phase consisted of 0.1 % (v/v) formic acid in acetonitrile and water. Flow rate was set to 0.5ml/ min. Hair samples free of drugs were used in testing. Hair were powdered in ball-mill and 50mg weighed samples were macerated in 1ml of methanol during 72 hours in 37°C. Results were compared to extract of 0,5ml whole blood sample (also free of drugs) obtained by ethyl acetate at pH 9. Short segments of mass range were set (25 amu) to improve sensitivity in scan mode. SIR and MRM modes were applied in LOD testing. Morphine was engaged as a model drug for matrix influence investigation.

Results

Ions map for m/z 50 to m/z 450 was created both for blood and hair matrix. That allowed to compare possible interferences from those matrices. Signal intensity and S/N were used for calculations. Plots of S/N versus analyte concentration appeared to better show LOD than signals intensity. Some ions (not the weakest) or MRM transitions disappeared from chromatograms of morphine causing lack of all parameters for appropriate identification although analyte was present in higher than LOD concentration.

Conclusions

Standard infusion during real sample analysis a few seconds after analyte retention time helped to solve problem of analyte presence in a sample. Matrix effect very often cannot be avoided but always should be controlled.

Le texte complet de cet article est disponible en PDF.

© 2014  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 26 - N° 2S

P. S32 - juin 2014 Retour au numéro
Article précédent Article précédent
  • P2 The detection of propofol glucuronide in human hair using LC-MS/MS: Case study and survey of high-risk health professionals
  • J. Jones, M. Jones, C. Plate, D. Lewis
| Article suivant Article suivant
  • P4: Multi-residue analysis of organic pollutants in hair and urine for matrices comparison
  • E. Hardy, R. Duca, G. Salquebre, B.M.R. Appenzeller

Bienvenue sur EM-consulte, la référence des professionnels de santé.
L’accès au texte intégral de cet article nécessite un abonnement.

Déjà abonné à cette revue ?

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.