Titanium-titanium modular neck for primary THA. Result of a prospective series of 170 cemented THA with a minimum follow-up of 5 years - 17/02/15

Doi : 10.1016/j.otsr.2014.12.013 
M. Ollivier, S. Parratte , A. Galland, A. Lunebourg, X. Flecher, J.-N. Argenson
 Institut du mouvement et de l’appareil locomoteur, UMR CNRS 787/AMU, hôpital Sainte-Marguerite, centre hospitalo-universitaire Sud, boulevard de Sainte-Marguerite, 13009 Marseille, France 

Corresponding author. Service de chirurgie orthopédique, institut du mouvement et de l’appareil locomoteur, UMR CNRS 787/AMU, hôpital Sainte-Marguerite, centre hospitalo-universitaire Sud, boulevard de Sainte-Marguerite, 13009 Marseille, France. Tel.: +33 4 91 74 50 11; fax: +33 4 91 74 56 25.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Sous presse. Épreuves corrigées par l'auteur. Disponible en ligne depuis le Tuesday 17 February 2015
Cet article a été publié dans un numéro de la revue, cliquez ici pour y accéder

Abstract

Background

Although they have been in use since the end of the 1980s, modular titanium neck components are associated with a risk of wear or fracture, and their safety has recently become a subject of debate and has never been evaluated in a consecutive series of patients. The goal of this study was to evaluate: revision-free survival of these implants after a minimum follow-up of 5 years; clinical and radiographic results; and the potential complications associated with the use of modular titanium neck components.

Hypothesis

The use of titanium modular neck on cemented titanium THA is safe at a minimum follow-up of 5 years.

Patients and methods

Between January 2006 and December 2008, we prospectively followed 170 patients (170 hips) who underwent primary anatomical THA with a modular cemented titanium stem design implant. The indications were unilateral THA for primary (n=160) or secondary (n=10) hip osteoarthritis (aseptic osteonecrosis of the femoral head or hip dysplasia). Mean age of patients was 75.4±5.8 years old (52–85), and mean BMI was 26.1±4.5kg/m2 (16.6–42.1). Patients were operated on by a modified Watson-Jones anterolateral approach based on preoperative 2D planning. All patients underwent annual clinical and radiological follow-up by an independent observer.

Results

At a mean follow-up of 71±8 months (60–84), 5 patients died and 7 were lost to follow-up. There was no revision of THA after a maximum follow-up of 84 months. The Harris score improved significantly from 50.4±11.3 (0–76) preoperatively to 84.5±15.2 (14–100) at the final follow-up. There was no difference in postoperative femoral offset or the position of the center of rotation compared to the opposite side. On the other hand, the neck-shaft angle (NSA) and limb length were corrected (2±5° [−11 to +14°] and 2.16±3.6mm [−7.4 to +12.7mm]) respectively. Fifteen patients (9%) had limb length discrepancies of more than 5mm and 4 patients (2%) of more than 10mm. There were no complications due to the modular implant design.

Discussion

Our study suggests that the use of cemented titanium implants with a modular titanium stem is safe at a follow-up of 5 years. The modular design does not prevent limb length discrepancies but restores femoral offset.

Level of evidence

IV: prospective, non-comparative study.

Le texte complet de cet article est disponible en PDF.

Keywords : Total hip replacement, Hip anatomy, Modular neck, Cemented stem


Plan


© 2015  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.