Evaluation of a new baseplate in reverse total shoulder arthroplasty – comparison of biomechanical testing of stability with roentgenological follow up criteria - 01/04/15

Doi : 10.1016/j.otsr.2014.11.015 
U. Irlenbusch a, , G. Kohut b
a Sports Clinic Erfurt, Am Urbicher Kreuz 7, 99099 Erfurt, Germany 
b Hòpital cantonal Fribourg, Fribourg, Switzerland 

Corresponding author. Tel.: +49 361 60221209; fax: +49 361 6022 111.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Background and purpose

To minimize notching problem associated with reversed prostheses, inferior positioning of base plate is recommended. This reduces the risk of notching, but does not eliminate it completely. Both polyethylene/PE-induced osteolysis and implant-to-bone or implant-to-implant contact may still occur, contributing to the risk of screw-breakage and resulting long-term failure. Therefore, the stability and integration of a newly developed base plate without inferior screw and inversion of bearing materials was investigated.

Patients and methods

Biomechanical assessment of primary stability of the two types of glenoid baseplate (1- and 2-pegged) was carried out according to ASTM F-2028-02 (American Society for Testing and Materials). Patients with a follow-up period of at least 2 years were clinically (n=78) and for most of them radiologically (n=61) examined. The X-rays were evaluated for loosening and scapular notching.

Results

The mean values of micromotions after 100,000 cycles showed no relevant differences between the 2-peg and the 1-peg base plates (47μm for the 2-peg design and 43μm for the 1-peg design), i.e. both were below the borderline for secure Osseointegration of 150μm. Radiologically, no signs of loosening or radiolucent lines/RLL were found for both base plates. The mean incidence of inferior scapular notching was 23.6% (42mm glenoid sphere: 15.8%). Only grade 1 and grade 2 notching was observed. Additionally as result of absence of PE-induced osteolysis shape, size, borderline and location of notching differed from those observed with conventional reverse total shoulder arthroplasty bearing materials.

Conclusion

In combination with modified inferior operating technique, the newly designed implant has the potential to reduce the incidence of scapular notching and to avoid both PE-induced osteolysis and metal-screw contact. The new design did not compromise stability of the base plate in any way during the investigation period, as demonstrated both by the data from the biomechanical investigation and also by the radiological follow-up.

Level of evidence

Level III, case-control study.

Le texte complet de cet article est disponible en PDF.

Keywords : Reverse prosthesis, Inversed shoulder prostheses, Micromotion, Scapular notching


Plan


© 2015  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 101 - N° 2

P. 185-190 - avril 2015 Retour au numéro
Article précédent Article précédent
  • Percutaneous bunionette correction: Results of a 49-case retrospective study at a mean 34 months’ follow-up
  • O. Laffenêtre, B. Millet-Barbé, V. Darcel, J. Lucas y Hernandez, D. Chauveaux
| Article suivant Article suivant
  • Can reverse shoulder arthroplasty in post-traumatic revision surgery restore the ability to perform activities of daily living?
  • M.W. Maier, F. Zeifang, M. Caspers, T. Dreher, M.C. Klotz, O. Rettig, S.I. Wolf, P. Kasten

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.