Advertising



Journal Français d'Ophtalmologie
Volume 41, n° 1
pages 14-20 (janvier 2018)
Doi : 10.1016/j.jfo.2017.07.003
Received : 1 April 2017 ;  accepted : 6 July 2017
Articles originaux

Agreement between Pentacam and optical coherence tomography in the assessment of iridocorneal angle width in a large healthy population
Accord entre Pentacam et tomographie par cohérence optique pour l’évaluation de l’ouverture de l’angle irido-cornéen dans une large population normale
 

Figure 1




Figure 1 : 

Measurements of the trabecular-iris angle by Pentacam (left) and optical coherence tomography (right).


Figure 2




Figure 2 : 

Bland–Altman plots showing the agreement between the trabecular-iris angle (TIA) measurements by Pentacam and by optical coherence tomography (OCT) in the temporal (left) and nasal (right) quadrants. The 95% confidence interval is also illustrated. This figure shows, in both quadrants, that the Pentacam tends to overestimate the measurements of narrow angles and underestimate those of open angles in comparison with OCT measurements.

EM-CONSULTE.COM is registrered at the CNIL, déclaration n° 1286925.
As per the Law relating to information storage and personal integrity, you have the right to oppose (art 26 of that law), access (art 34 of that law) and rectify (art 36 of that law) your personal data. You may thus request that your data, should it be inaccurate, incomplete, unclear, outdated, not be used or stored, be corrected, clarified, updated or deleted.
Personal information regarding our website's visitors, including their identity, is confidential.
The owners of this website hereby guarantee to respect the legal confidentiality conditions, applicable in France, and not to disclose this data to third parties.
Close
Article Outline