S'abonner

What factors predict ceramic liner malseating after ceramic-on-ceramic total hip arthroplasty? - 21/04/19

Doi : 10.1016/j.otsr.2019.01.009 
Yoshitoshi Higuchi , Taisuke Seki, Yasuhiko Takegami, Yusuke Osawa, Taiki Kusano, Naoki Ishiguro
 Department of Orthopaedic Surgery, Nagoya University Graduate School of Medicine, 65 Tsurumai-cho, Showa-ku, Nagoya, Japan 466-8550 

Corresponding author.

Bienvenue sur EM-consulte, la référence des professionnels de santé.
Article gratuit.

Connectez-vous pour en bénéficier!

Abstract

Background

There is a lack of evidence about the risk factors associated with the malseating of the acetabular liner after ceramic-on-ceramic (CoC) total hip arthroplasty (THA). Therefore, we performed a complementary retrospective case-control study to determine the factors predicting the malseating of the acetabular liner after CoC THA and to evaluate the relationship between malseating and (1) osteoarthritis, (2) particularly in terms of the new radiographic parameter “bone sclerotic length” of the acetabular bone.

Hypothesis

Osteoarthritis, particularly bone sclerotic length, was an independent risk factor for malseating of ceramic liners.

Patients and methods

In total, 219 CoC THAs (174 women and 45 men) were evaluated to determine the risk factors influencing the malseating of the acetabular ceramic liner. An average patient age at the time of surgery was 55.9±9.5 years (range, 23 to 75 years). Data on patient background and preoperative radiographs, such as Tönnis grades; Crowe classification; and indices of acetabular osteoarthritis change; including bone cyst, osteophyte and bone sclerosis, were assessed. The bone sclerotic length in patients with osteoarthritis was measured as the slant distance between the bilateral edges of the sclerosis lesion of the acetabulum on the anteroposterior view.

Results

Preoperative less hip flexion (hazard ratio [HR]: 0.98; 95% CI: 0.97–0.99), osteoarthritis (HR: 3.15; 95% CI: 1.02–9.70) and the bone sclerotic length (HR: 1.83; 95% CI: 1.35–2.48) were independent risk factors determining the malseating of ceramic liners. Receiver operating characteristic curve analysis showed that a bone sclerotic length of 24.6mm was defined as the cut-off point for the malseating of the ceramic liner.

Discussion

Age, preoperative less flextion, osteoarthritis, and the bone sclerotic were independent risk factors determining malseating of ceramic liners. The acetabular shell can also deform upon insertion of the cup with sclerotic bone of the acetabulum and prevent correct seating of liners. Therefore, these factors must be taken into consideration when seating the ceramic liner.

Level of evidence

Case control study III, case control retrospective design.

Le texte complet de cet article est disponible en PDF.

Keywords : Ceramic on ceramic bearings, Total hip arthroplasty, Liner malseating, Risk factor, Ceramic liner with a metal-back titanium sleeve


Plan


© 2019  Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 105 - N° 3

P. 453-459 - mai 2019 Retour au numéro
Article précédent Article précédent
  • Does the direct anterior approach allow earlier recovery of walking following total hip arthroplasty? A randomized prospective trial using accelerometry
  • Gautier Bon, Elias Ben Kacem, Pierre Marie Lepretre, Thierry Weissland, Patrice Mertl, Massinissa Dehl, Antoine Gabrion
| Article suivant Article suivant
  • Dual mobility device reduces the risk of prosthetic hip instability for patients with degenerated spine: A case-control study
  • Louis Dagneaux, Stanislas Marouby, Cedric Maillot, François Canovas, Charles Rivière

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.