Ocular Response Analyzer : étude de fiabilité et de corrélation sur des yeux normaux

Introduction : Étudier la fiabilité des paramètres de l’ORA (CH, CRF, IOPcc et IOPg) et leur corrélation aux données biométriques de l’Orbscan® (Orbtek, Bausch et Lomb, Salt lake City, Utah, États-Unis).

Matériel et méthodes : L’Ocular Response Analyzer (ORA de Reichert, Depew, New York, États-Unis) donne la rigidité de la cornée (CH), son coefficient viscoélastique (CRF) ainsi qu’une pression intra-oculaire corrigé (IOPcc). Les paramètres de l’ORA ont été étudiés sur une population de 100 yeux normaux, puis corrélés aux données biométriques de la cornée mesurées à l’aide de l’Orbscan®.

Résultats : Une bonne corrélation et une faible différence des mesures entre les yeux droit et gauche ont confirmé la fiabilité des mesures (rS = 0,84, p < 0,001 et 8,52 %). La répartition des paramètres de l’ORA est considérée comme gaussienne, et avec une moyenne de 10,25 ± 1,6 mmHg (6,5 à 14,4) pour l’hystérésis et de 10,25 ± 1,85 mmHg (4,9 à 14,2) pour le coefficient de rigidité. De plus, ils sont très corrélés à la pachymétrie. La PIO corrigée n’est pas corrélée à la pachymétrie, mais n’est pas significativement différente de la PIO de Goldmann. Il existe une faible corrélation entre les facteurs biomécaniques (CH, CRF) et le diamètre cornéen et la puissance du cylindre.

Conclusion : Les valeurs normales et la répartition des facteurs biomécaniques sont équivalentes à ce qu’on retrouve dans la littérature et sont corrélées aux données biométriques de la cornée (pachymétrie, diamètre cornéen). La nouvelle mesure de PIO corrigée est indépendante de la pachymétrie, mais n’est pas significativement différente de la PIO tonométrique à air donc son intérêt reste à démontrer.

Ocular Response Analyzer: feasibility study and correlation with normal eyes

Introduction: To evaluate the accuracy of ocular response analyzer (ORA) parameters (corneal hysteresis (CH), corneal resistance factor (CRF), and ocular tension (IOPcc and IOPg)) and the correlation with corneal biometry measured with the Orbscan® topographer.

Material and methods: The Ocular Response Analyzer is a new instrument that measures ocular rigidity/elasticity (CH and CRF) and intraocular pressure (IOPcc), which is assumed to be independent of corneal pachymetry. We compared the results of the ORA in 100 eyes with normal slit lamp examination, except ametropia, with the results from corneal measurements obtained with the Orbscan®.

Results: A high correlation and a low difference between the right and left eyes show the accuracy of measurements (rS=0.84, p<0.001 and 8.52%). The mean CH and CRF were 10.25±1.6 mmHg (range, 6.5-14.4) and 10.25±1.85 mmHg (range, 4.9-14.2), respectively with a Gaussian distribution in normal eyes. The relationship between CH and CRF was significant with corneal pachymetry but not with IOPcc (corneal corrected). CH and CRF were related, although to a lesser extent, with corneal diameter and astigmatism power, but not to keratometry, sex, age, or spherical equivalent.

Conclusion: The mean and the distribution of biomechanical factors are similar to the values found in the literature but the relationship to pachymetry seems to be stronger (pachymetry and corneal diameter) than what has been reported in previous publications. The new IOP corrected for the cornea is independent of pachymetry but is not significantly different from Goldmann IOP or IOP measured with a standard noncontact tonometer.


Mots clés : Ocular Response Analyzer , biomécanique de la cornée , Orbscan® , pression intra-oculaire

Keywords: Ocular Response Analyzer , Orbscan® , corneal biomechanics , intraocular pressure


Plan



© 2007 Elsevier Masson SAS. Tous droits réservés.
Ajouter à ma bibliothèque Retirer de ma bibliothèque Imprimer
Export

    Export citations

  • Fichier

  • Contenu

Vol 30 - N° 10

P. 978-984 - décembre 2007 Retour au numéro
Article suivant Article suivant
  • Schwannomes orbitaires isolés : aspects cliniques, radiologiques et chirurgicaux
  • H. Kiratli, K. Erkan, F. Söylemezoglu

Bienvenue sur EM-consulte, la référence des professionnels de santé.

Mon compte


Plateformes Elsevier Masson

Déclaration CNIL

EM-CONSULTE.COM est déclaré à la CNIL, déclaration n° 1286925.

En application de la loi nº78-17 du 6 janvier 1978 relative à l'informatique, aux fichiers et aux libertés, vous disposez des droits d'opposition (art.26 de la loi), d'accès (art.34 à 38 de la loi), et de rectification (art.36 de la loi) des données vous concernant. Ainsi, vous pouvez exiger que soient rectifiées, complétées, clarifiées, mises à jour ou effacées les informations vous concernant qui sont inexactes, incomplètes, équivoques, périmées ou dont la collecte ou l'utilisation ou la conservation est interdite.
Les informations personnelles concernant les visiteurs de notre site, y compris leur identité, sont confidentielles.
Le responsable du site s'engage sur l'honneur à respecter les conditions légales de confidentialité applicables en France et à ne pas divulguer ces informations à des tiers.


Tout le contenu de ce site: Copyright © 2024 Elsevier, ses concédants de licence et ses contributeurs. Tout les droits sont réservés, y compris ceux relatifs à l'exploration de textes et de données, a la formation en IA et aux technologies similaires. Pour tout contenu en libre accès, les conditions de licence Creative Commons s'appliquent.