Absence of additional improvement in outcome of patients receiving cardiac resynchronization therapy paced at the most delayed left ventricular region
Stimuler la paroi ventriculaire gauche la plus retardée ne permet pas de bénéfice supplémentaire chez les patients resynchronisés
 

Antoine Deplagne, Stephane Lafitte, Sylvain Reuter, Patricia Reant, Sylvain Ploux, Bilel Mokrani, Raymond Roudaut, Pierre Jais, Michel Haissaguerre, Jacques Clementy, Pierre DosSantos, Pierre Bordachar
Service Pr-Clementy, hôpital Haut-Lévêque, avenue Magellan, 33600 Pessac, France 

Corresponding author. Fax: +11 33 5 57 65 65 09.
Summary
Background

The choice of the optimal left ventricular (LV) pacing site remains an issue in patients requiring cardiac resynchronization therapy (CRT).

Aim

This prospective study compared the outcome of patients paced at the most delayed LV region with that of patients paced at any other LV site.

Methods

Forty-four patients with severe heart failure underwent three-dimensional (3D) echocardiography before implantation and 3 days after implantation of a CRT device, to determine the most delayed LV region during spontaneous rhythm and during right ventricular pacing. The patients were divided subsequently into four groups: group 1 (n =19), LV lead placed at the most delayed echocardiographic site in spontaneous rhythm; group 2 (n =25), LV lead placed at any other site; group 3 (n =21), LV lead placed at the most delayed echocardiographic site during right ventricular pacing; group 4 (n =23), LV lead placed at any other site.

Results

No significant differences were observed between the four groups before implantation. After 6 months of CRT, no significant differences were observed between groups 1 and 2 or between groups 3 and 4 in terms of change in New York Heart Association functional class, Minnesota living with heart failure questionnaire, 6-minute walk test, peak exercise oxygen consumption, 3D ventricular dyssynchrony and 3D LV ejection fraction.

Conclusion

Implantation of the LV lead in the most delayed region of the left ventricle determined by 3D echocardiography did not result in additional improvement in symptoms or LV function.

The full text of this article is available in PDF format.
Résumé
Introduction

Chez les patients bénéficiant d’une resynchronisation biventriculaire, le choix du site optimal de stimulation ventriculaire gauche reste controversé. Cette étude prospective compare le devenir de patients stimulés au niveau de la région ventriculaire gauche (VG) la plus retardée et le devenir de patients stimulés dans une autre région.

Méthode

Quarante-quatre patients insuffisants cardiaques bénéficièrent d’une échocardiographie tridimensionnelle (3D) avant l’implantation et trois jours après pour déterminer la paroi la plus retardée en rythme spontané (RS) et lors d’une stimulation ventriculaire droite (VD). Les patients étaient divisés en quatre groupes : groupe 1 (n =19) : sonde VG placée au niveau de la paroi la plus retardée en RS ; groupe 2 (n =25) : sonde VG placée sur une autre paroi ; groupe 3 (n =21) : sonde VG placée au niveau de la paroi la plus retardée lors d’une stimulation VD ; groupe 4 (n =23) : sonde VG placée sur une autre paroi.

Résultats

Nous n’avons pas retrouvé de différence significative entre les quatre groupes avant l’implantation. Après six mois de stimulation, nous n’avons pas retrouvé de différence significative entre les patients des groupes 1 et 2 et des patients des groupes 3 et 4 en termes de modification de classe New York Heart Association , de questionnaire de qualité de vie, de périmètre de marche, de pic de VO2 , d’asynchronisme 3D et de fraction d’éjection.

Conclusion

L’implantation de la sonde ventriculaire gauche dans la paroi la plus retardée n’a pas permis d’apporter un bénéfice supplémentaire en termes de symptômes et de fonction ventriculaire gauche.

The full text of this article is available in PDF format.

Keywords : Cardiac resynchronization therapy, Left ventricular pacing site, Echocardiography

Mots clés : Resynchronisation biventriculaire, Site de stimulation ventriculaire gauche, Échocardiographie


Abbreviations

3D : three-dimensional
CRT : cardiac resynchronization therapy
LV : left ventricular
LVEDV : left ventricular end-diastolic volume
LVEF : left ventricular ejection fraction
LVESV : left ventricular end-systolic volume

Introduction

CRT has an established role in the management of patients with advanced heart failure and prolonged QRS duration [1, 2, 3]. However, one third of patients do not respond to CRT, despite the application of established electrocardiographic selection criteria [4]. To improve outcome and reduce the proportion of non-responders, three different and complementary approaches have been proposed: optimization of patient selection; optimization of LV lead placement and optimization of the programming of the CRT device [5, 6, 7, 8, 9, 10]. To date, identification of the optimal LV lead position has attracted little attention. Previous studies have suggested that implantation of the LV lead in the area of the most delayed pre-implantation echocardiographically-determined electromechanical activation may be associated with an improvement in the response to CRT and a decrease in the proportion of non-responders [11, 12, 13]. Similarly, placement of the LV electrode in the area with maximal delay during right ventricular pacing may optimize the electromechanical activation. The aim of the present study was to evaluate, by use of 3D echocardiography, the putative favourable impact on the efficacy of CRT, in terms of ventricular dyssynchrony and midterm clinical status, of an echocardiographically-optimized LV lead position targeting the exact region of maximal mechanical delay during spontaneous rhythm and during right ventricular pacing.

Methods
Study population

Forty-four successive patients with refractory heart failure due to severe systolic dysfunction, prolonged QRS duration, in sinus rhythm and scheduled for implantation of a CRT device were enrolled prospectively. Refractory heart failure was defined by the persistence of New York Heart Association functional class III or IV despite optimal pharmacological therapy. Severe systolic dysfunction was defined by a LVEF less than 35%, while prolonged QRS duration was defined by a QRS width greater than 120ms. Patients were excluded from the study if they had a history of sustained atrial arrhythmias or complete atrioventricular block or a poor ultrasonic window that did not allow exploitable 3D acquisitions. All patients provided written, informed consent to the study, which was approved by the institutional clinical research and ethics committee.

Study protocol

All patients underwent echocardiography, a quality-of-life assessment using the Minnesota living with heart failure questionnaire, a 6-minute hall walk test and a maximal bicycle and treadmill cardiopulmonary test with measurement of peak exercise oxygen consumption within 5 days before implantation and after 6 months of simultaneous biventricular pacing.

Biventricular device implantation and programming

All leads were implanted transvenously. The atrial lead was positioned conventionally at the right atrial appendage and the right ventricular lead at the apex of the right ventricle. The LV lead was positioned through the coronary sinus into a posterior (n =9), lateral (n =23) or anterior (n =12) cardiac vein. The physician who implanted the LV lead was blinded to the results of the echocardiographic examination. In clinical practice, a lateral or a posterior vein is usually targeted. In this study, the priority was to achieve a stable position with suitable threshold and absence of diaphragmatic pacing. No intraoperative haemodynamic evaluation was carried out. The pacing leads were connected to a biventricular implantable cardioverter-defibrillator (InSync Sentry 7298, Medtronic [Minnesota, USA] Contak Renewal 4 HE, Guidant [Minneapolis, USA]). For 3 days, the patients were not paced in the ventricles (AAI mode, 40 beats/minute). Three days after implantation, the atrioventricular delay was optimized echocardiographically during biventricular pacing to provide the longest transmitral LV filling time without truncation of the A-wave obtained from pulsed Doppler analysis of the LV filling. The interventricular timing was set to 0 in all patients.

Echocardiographic evaluations

Real-time 3D echocardiography was performed 3 days before implantation, 3 days after implantation and after 6 months of simultaneous biventricular pacing, using a 3D probe connected to a 3D digital ultrasound system (Vivid 7, GE Vingmed, Horton, Norway). During the echocardiography 3 days after implantation, echocardiographic recordings were made during atrio-sensed right ventricular pacing and atrio-sensed simultaneous biventricular pacing. All echocardiographic recordings were made by the same physician to minimize variability between examinations. All images were recorded digitally and analysed off-line. The off-line analysis was performed by an observer blinded to the fluoroscopic lead positions (Figure 1). A full-volume loop of the left ventricle was acquired using an apical position of the probe during a short breath hold. The LVEF, LVESV and LVEDV were determined off-line with the aid of semiautomatic contour tracing software (4D LV-Analysis, TomTec, Unterschleissheim, Germany) as described previously [14]. A 3D parameter of dyssynchrony was obtained from the time-course of shortening in 16 LV segments and the resulting segmental volume/time curves. The 3D dyssynchrony index was defined as the standard deviation of the 16 segmental shortening durations to reach minimum segmental volume [15]. To allow comparisons between patients with significantly different heart rates, this variable was expressed as a percentage of the cardiac cycle duration.



Figure 1


Figure 1. 

After implantation of the cardiac resynchronization therapy device, fluoroscopic orthogonal views (right anterior oblique [RAO] at 30° and left anterior oblique [LAO] at 60°) were acquired. LAO at 60°: the resized 16-segment scheme was projected. RAO: divided into basal, medial and apical sections.

Zoom

Based on the resulting segmental volume/time curve in each patient, the segment with the latest minimum of systolic volume as an indicator for latest systolic contraction was identified before implantation and defined as the segment with maximum mechanical delay. Similarly, 3 days after implantation, the segment with the latest minimum of systolic volume as an indicator for latest systolic contraction was identified during right ventricular pacing.

Concordance between most delayed echocardiographic left ventricular segment (spontaneous rhythm and right ventricular pacing) and left ventricular pacing site

After CRT implantation, biplane fluoroscopy was performed in left anterior and right anterior oblique orthogonal views. These images were analysed by two independent physicians using a 16-segment scheme, identical to the scheme used in the 4D LV-analysis (TomTec, Unterschleissheim, Germany) and projected onto the left anterior oblique view to determine the anatomical location of the LV lead within the circumference of that level. The right anterior oblique view was used subsequently to define the basal, medial or apical level of the LV lead.

A correlation between the fluoroscopic LV pacing site and the most delayed echocardiographically-determined LV site was assessed blindly, as described previously [11], by two observers with full concordance. The physicians performing echocardiographic analysis and classification of LV lead position as optimal or non-optimal were blinded with regard to possible difficulties during LV lead placement and anatomical limitations.

Statistical analysis

Interobserver and intraobserver reproducibility of 3D echocardiographic measurements were assessed with linear regression analysis and the Bland-Altman method in 32 patients (mean age 53±16 years; 28 men) who were not included in the present study. These patients were selected to demonstrate different levels of LV dysfunction ranging from normal heart to severe cardiomyopathy. Continuous variables are presented as means±standard deviations. Categorical data are presented as frequencies and percentages. Sequential data measurements were analysed by repeated measures analysis of variance followed by Scheffé’s procedure for multiple comparisons. Statistical significance was established at p <0.05.

Results

The patients’ baseline characteristics are presented in Table 1. The entire study protocol was completed in all 44 patients. No patient died during the 6-month follow-up period. Interobserver variability mean average error and 95% confidence interval values obtained from the Bland and Altman analysis of 3D LVEF, 3D LVESV and 3D LVEDV were −0.2 and 3.3%, −1.2 and 13.7mL, and −0.2 and 16.1mL, respectively. Intraobserver variability mean average error and 95% confidence interval values obtained from the Bland and Altman analysis of 3D LVEF, 3D LVESV and 3D LVEDV were 0.1 and 2.8%, −0.1 and 8.4mL, and 0.3 and 9.6mL, respectively (Figure 2, Figure 3).



Figure 2


Figure 2. 

Interobserver reproducibility in terms of three-dimensional left ventricular ejection fraction (LVEF), left ventricular end-diastolic volume (LVEDV) and left ventricular end-systolic volume (LVESV).

Zoom



Figure 3


Figure 3. 

Intraobserver reproducibility in terms of three-dimensional left ventricular ejection fraction (LVEF), left ventricular end-diastolic volume (LVEDV) and left ventricular end-systolic volume (LVESV).

Zoom

In 19 patients (Group 1), the LV lead position determined by fluoroscopy was found to be concordant with the site of the most delayed mechanical activation determined in spontaneous rhythm before implantation. The LV lead was positioned in the following segments: nine lateral (three basal, five medial and one apical), six anterior (two basal and four medial) and four posterior (one basal and three medial). In 25 patients (Group 2), the LV lead position did not match with the most delayed site. The LV lead was positioned in the following segments: 14 lateral (four basal, nine medial and one apical), six anterior (four basal and two medial) and five posterior (four medial and one apical).

In 21 patients (Group 3), the LV lead position determined by fluoroscopy was found to be concordant with the site of the most delayed mechanical activation determined during right ventricular pacing after implantation. The LV lead was positioned in the following segments: 11 lateral (five basal, five medial and one apical), four anterior (three basal and one medial) and six posterior (one basal and five medial). In 23 patients (Group 4), the LV lead position did not match with the most delayed site during right ventricular pacing. The LV lead was positioned in the following segments: 12 lateral (two basal, nine medial and one apical), eight anterior (three basal and five medial) and three posterior (two medial and one apical).

There was no difference between patients in Groups 1 and 2 or between patients in Groups 3 and 4 in terms of demographic, clinical and echocardiographic characteristics at baseline, or in the distribution of the location of the most delayed segment (Table 2, Table 3).

Three days after implantation, a significant increase was observed compared with baseline in real-time 3D echocardiographically-determined LVEF (31.8±7% vs 25.8±7%, r <0.05) and a decrease in the dyssynchrony index (6.7±1.8% vs 9.9±2.9%; p <0.05) in the overall population, without significant difference between the four groups.

After 6 months of simultaneous biventricular pacing, a significant improvement was observed in clinical status in terms of New York Heart Association functional class (2.3±0.3 vs 3.2±0.3, p <0.05), Minnesota living with heart failure questionnaire (30±15 vs 54±18, p <0.05), 6-minute walk test (415±83m vs 296±69m, p <0.05) and peak oxygen consumption (15.2±3.8mL/kg/min vs 13.4±3.5mL/kg/min, p <0.05) in the overall population, compared with baseline. These beneficial effects of CRT were supported by a decrease in LVESV (99±48mL vs 130±69mL, p <0.05) and LVEDV (144±60mL vs 176±80mL, p <0.05), an increase in LVEF (33.3±6.4% vs 27.5±7%, p <0.05) and a decrease in the dyssynchrony index (4.9±1.2% vs 9.9±2.9%, p <0.05).

After 6 months of simultaneous biventricular pacing, no significant differences were observed in any of the studied variables between patients in Group 1 and Group 2 or between patients in Group 3 and Group 4.

Discussion
Impact of the left ventricular pacing site

In patients requiring CRT, the choice of optimal LV pacing site remains an issue. Systematic implantation of the LV lead at the lateral wall is often recommended, simplifying the choice when a suitable lateral vein can be cannulated. However, the existence of an identical, unique, optimal pacing site, i.e., the lateral wall, for all implanted patients, despite major interpatient differences in electromechanical activation, seems very unlikely. Tailoring of the LV lead position may provide additional haemodynamic improvement and optimization of the pacing site for each patient may be recommended. Therefore, any technique or strategy that might optimize LV lead position deserves interest.

Echocardiography to select the optimal pacing site

In an experimental study that included a very limited number of patients, tailoring of the LV lead position to the area of maximal electrical delay resulted in the best acute resynchronization effect [16]. The human results in terms of clinical impact, reverse remodelling and reduction of mortality associated with this strategy are still controversial [17, 18].

In our present study, echocardiographic optimization of the LV lead position based on a detailed analysis of the myocardial contraction sequence before implantation was not associated with additional short-term or long-term improvement. Our results contradict some results published previously, despite having a similar study design [11]. This discordance highlights the disparity in results obtained with echocardiography in the context of CRT. Our study (and others) combined complex techniques, such as tissue Doppler imaging, tissue synchronization imaging or real-time 3D echocardiography, to define the segment with most delayed activation. In a recent, prospective, multicentre trial, the ability of several conventional two-dimensional echocardiographic variables and techniques to predict response to CRT was evaluated [19]. Poor interobserver reproducibility with the use of these variables and techniques may explain why studies with similar design produce conflicting results. However, the absence of an additional beneficial effect with LV implantation in the most delayed region in our study might have a more pathophysiological explanation. In a high proportion of patients, biventricular pacing is performed with complete capture and absence of fusion between intrinsic and ectopic rhythms. Therefore, the electromechanical activation after CRT depends on the location of the pacing leads and on the electromechanical properties of the LV myocardium, independent of the LV activation pattern before implantation. As ventricular pacing induces areas of early activation around the pacing lead and areas of late activation at a distance from the lead, it is likely that stimulation in the most delayed region, determined in spontaneous rhythm, will convert this zone into an early activated site, whereas previous regions of early activation will be delayed. We believe that there is no reason for these modifications in electromechanical activation to be associated with a maximal reduction in ventricular dyssynchrony or with optimal acute haemodynamic or long-term clinical improvements. Furthermore, in patients with ischaemic cardiomyopathy, the most delayed segments are often composed of viable but ischaemic tissue, while pacing the ischaemic area has been demonstrated to be ineffective, if not deleterious [20]. Preprocedural echocardiography may not therefore be the adequate tool for predicting the optimal pacing sites.

In contrast, periprocedural echocardiographic comparisons between the different pacing sites may be technically difficult but more appropriate for determining the LV pacing site, associated with maximal reduction in ventricular dyssynchrony and maximal haemodynamic improvement. Echocardiography may be performed during the procedure to determine the optimal LV pacing site based on a detailed analysis of the myocardial contraction sequence once the correct ventricular lead has been implanted. Our present study assessed the putative favourable impact of targeting the most delayed LV region during right ventricular pacing. Unfortunately, we did not demonstrate an increased benefit with an echocardiographically-optimized LV lead position targeting the exact region of maximal mechanical delay during right ventricular pacing. Therefore, just like preprocedural echocardiography, this strategy may not be helpful in improving the response to the therapy.

Limitations

The study population size was rather small and the statistical analysis may have been underpowered. Moreover, the duration of follow-up was rather short. Therefore, this study should probably be considered as a pilot study. A multicentre, randomized study is warranted to confirm the clinical impact of pre-implantation echocardiographically guided selection of the LV pacing site.

Conclusions

This prospective study compared the outcome of patients paced at the most delayed LV region with that of patients paced at any other LV site. Implantation of the LV lead in the most delayed region of the left ventricle determined by 3D echocardiography did not result in additional improvement in symptoms or LV function.

Conflict of interest

None.

References

Abraham W.T., Fisher W.G., Smith A.L., and al. Cardiac resynchronization in chronic heart failure N Engl J Med 2002 ;  346 : 1845-1853 [cross-ref]
Bristow M.R., Saxon L.A., Boehmer J., and al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure N Engl J Med 2004 ;  350 : 2140-2150 [cross-ref]
Cleland J.G., Daubert J.C., Erdmann E., and al. The effect of cardiac resynchronization on morbidity and mortality in heart failure N Engl J Med 2005 ;  352 : 1539-1549 [cross-ref]
Reuter S., Garrigue S., Barold S.S., and al. Comparison of characteristics in responders versus nonresponders with biventricular pacing for drug-resistant congestive heart failure Am J Cardiol 2002 ;  89 : 346-350 [cross-ref]
Bax J.J., Marwick T.H., Molhoek S.G., and al. Left ventricular dyssynchrony predicts benefit of cardiac resynchronization therapy in patients with end-stage heart failure before pacemaker implantation Am J Cardiol 2003 ;  92 : 1238-1240 [cross-ref]
Bordachar P., Lafitte S., Reuter S., and al. Echocardiographic parameters of ventricular dyssynchrony validation in patients with heart failure using sequential biventricular pacing J Am Coll Cardiol 2004 ;  44 : 2157-2165 [cross-ref]
Breithardt O.A., Stellbrink C., Kramer A.P., and al. Echocardiographic quantification of left ventricular asynchrony predicts an acute hemodynamic benefit of cardiac resynchronization therapy J Am Coll Cardiol 2002 ;  40 : 536-545 [cross-ref]
Kindermann M., Frohlig G., Doerr T., and al. Optimizing the AV delay in DDD pacemaker patients with high degree AV block: mitral valve Doppler versus impedance cardiography Pacing Clin Electrophysiol 1997 ;  20 : 2453-2462 [cross-ref]
Yu C.M., Fung J.W., Zhang Q., and al. Tissue Doppler imaging is superior to strain rate imaging and postsystolic shortening on the prediction of reverse remodeling in both ischemic and nonischemic heart failure after cardiac resynchronization therapy Circulation 2004 ;  110 : 66-73 [cross-ref]
Yu C.M., Fung W.H., Lin H., and al. Predictors of left ventricular reverse remodeling after cardiac resynchronization therapy for heart failure secondary to idiopathic dilated or ischemic cardiomyopathy Am J Cardiol 2003 ;  91 : 684-688 [cross-ref]
Becker M., Hoffmann R., Schmitz F., and al. Relation of optimal lead positioning as defined by three-dimensional echocardiography to long-term benefit of cardiac resynchronization Am J Cardiol 2007 ;  100 : 1671-1676 [cross-ref]
Becker M., Kramann R., Franke A., and al. Impact of left ventricular lead position in cardiac resynchronization therapy on left ventricular remodelling. A circumferential strain analysis based on 2D echocardiography Eur Heart J 2007 ;  28 : 1211-1220 [cross-ref]
Murphy R.T., Sigurdsson G., Mulamalla S., and al. Tissue synchronization imaging and optimal left ventricular pacing site in cardiac resynchronization therapy Am J Cardiol 2006 ;  97 : 1615-1621 [cross-ref]
Sugeng L., Mor-Avi V., Weinert L., and al. Quantitative assessment of left ventricular size and function: side-by-side comparison of real-time three-dimensional echocardiography and computed tomography with magnetic resonance reference Circulation 2006 ;  114 : 654-661 [cross-ref]
Kapetanakis S., Kearney M.T., Siva A., and al. Real-time three-dimensional echocardiography: a novel technique to quantify global left ventricular mechanical dyssynchrony Circulation 2005 ;  112 : 992-1000 [cross-ref]
Auricchio A., Fantoni C., Regoli F., and al. Characterization of left ventricular activation in patients with heart failure and left bundle-branch block Circulation 2004 ;  109 : 1133-1139 [cross-ref]
Fung J.W., Lam Y.Y., Zhang Q., and al. Effect of left ventricular lead concordance to the delayed contraction segment on echocardiographic and clinical outcomes after cardiac resynchronization therapy J Cardiovasc Electrophysiol 2009 ;  20 : 530-535 [cross-ref]
Ypenburg C., van Bommel R.J., Delgado V., and al. Optimal left ventricular lead position predicts reverse remodeling and survival after cardiac resynchronization therapy J Am Coll Cardiol 2008 ;  52 : 1402-1409 [cross-ref]
Chung E.S., Leon A.R., Tavazzi L., and al. Results of the Predictors of Response to CRT (PROSPECT) trial Circulation 2008 ;  117 : 2608-2616 [cross-ref]
Bordachar P., Labrousse L., Thambo J.B., and al. Haemodynamic impact of the left ventricular pacing site during graded ischaemia in an open-chest pig model Europace 2008 ;  10 : 242-248 [cross-ref]



© 2009  Elsevier Masson SAS. All Rights Reserved.
EM-CONSULTE.COM is registrered at the CNIL, déclaration n° 1286925.
As per the Law relating to information storage and personal integrity, you have the right to oppose (art 26 of that law), access (art 34 of that law) and rectify (art 36 of that law) your personal data. You may thus request that your data, should it be inaccurate, incomplete, unclear, outdated, not be used or stored, be corrected, clarified, updated or deleted.
Personal information regarding our website's visitors, including their identity, is confidential.
The owners of this website hereby guarantee to respect the legal confidentiality conditions, applicable in France, and not to disclose this data to third parties.
Close
Article Outline