Delayed right ventricular perforation with a pacemaker lead into subcutaneous tissues
Déplacement tardif d’une sonde de stimulation cardiaque ventriculaire droite dans les tissus sous-cutanés
 

Walid Amara , Michel Cymbalista, Jacky Sergent
Service de cardiologie et radiologie, centre hospitalier intercommunal, 10, rue du Général-Leclerc, 93370 le Raincy-Montfermeil, France 

Corresponding author.

Keywords : Ventricular perforation, Pacemaker lead

Mots clés : Perforation ventriculaire, Sonde de stimulation cardiaque


An 82-year-old woman underwent routine pacemaker implantation for paroxysmal atrioventricular block. The operator did not encounter any difficulties at the time of lead placement. At a predischarge check three days later, electrical signals and thresholds were excellent and similar to those at implant. The chest X-ray showed a good lead position in the right ventricular apex. During a routine one-month post-implant visit, the patient pointed out that she felt, two weeks after discharge, a violent left-thoracic pain. The pain started while the patient was lifting a heavy object and ended five days later. Pacing could not to be achieved and the electrograms were small. Fluoroscopy suggested perforation. There was no pericardial effusion on echocardiography. Ventricular perforation was confirmed by computed tomography scan (Figure 1). The tip of the lead (5076 Medtronic Inc., Minneapolis, MN, USA) was found in the subcutaneous tissues. Because the subsequent attempt to remove the lead could induce an acute cardiac tamponade, the procedure was performed in an operating theatre with a suitable emergency backup, that is, immediate availability of echocardiography, equipment for pericardiocentesis and a thoracic surgeon on stand by.



Figure 1


Figure 1. 

Computer tomography scanning is showing right ventricular perforation with a ventricular lead into subcutaneous tissues.

Zoom

The lead was pulled back into the ventricle and repositioned at the same time on the interventricular septum. At the end of the procedure, arterial pressure dropped and echocardiography revealed a moderate pericardial effusion with mild compression of the right ventricle. A subxiphoid incision was made. The surgeon evacuated a mild pericardial haematoma. No ventricular injury was found. A drain was positioned in the pericardial space for 48h. The patient continued to recover well after the hospitalization.

Cardiac perforation by the implanted pacemaker lead occurs at a complication rate of 0.4–2.0%. If the perforation is diagnosed more than one month after implantation, it is described as delayed. The most common manifestations of delayed perforation may be haemopericardium, cardiac tamponade, pectoral muscle or diaphragm stimulation, and pneumothorax. Loss of stimulation properties can be the only sign indicating this complication. Diagnosis is habitually done by fluoroscopy and echocardiography. Computed tomography scanning is a sensitive tool for the diagnosis of cardiac perforation. The major risk factors for cardiac perforation are age, female sex, body mass index below 20, use of anticoagulants or steroids, and use of leads with an extendable fixation lead.

Several options are available to treat this complication. The first is to pull back the lead into the ventricle, performed in the operating theatre, with suitable preparation because the heart often ‘self-seals’ after perforation. The second is to refer the patient to the cardiac surgery department, to withdraw the lead under visual control, usually after sternotomy. An epicardial lead could be implanted during this procedure. Whichever option is chosen, surgical stand-by is necessary, as was the case in our patient. To prevent this complication, we recommend implanting the right ventricular lead in a septal position and advising the patient not to carry heavy objects or undertake vigorous tasks within the month after implantation (Figure 1).



© 2008  Elsevier Masson SAS. All Rights Reserved.
EM-CONSULTE.COM is registrered at the CNIL, déclaration n° 1286925.
As per the Law relating to information storage and personal integrity, you have the right to oppose (art 26 of that law), access (art 34 of that law) and rectify (art 36 of that law) your personal data. You may thus request that your data, should it be inaccurate, incomplete, unclear, outdated, not be used or stored, be corrected, clarified, updated or deleted.
Personal information regarding our website's visitors, including their identity, is confidential.
The owners of this website hereby guarantee to respect the legal confidentiality conditions, applicable in France, and not to disclose this data to third parties.
Close
Article Outline
You can move this window by clicking on the headline
@@#110903@@