Archives of cardiovascular diseases
Volume 104, n° 1
pages 11-16 (janvier 2011)
Doi : 10.1016/j.acvd.2010.11.005
Received : 28 August 2010 ;  accepted : 12 November 2010
Initial clinical experience with implantation of left ventricular lead guided by Overlay Ref for the treatment of congestive heart failure
Évaluation clinique initiale d’une technique d’implantation d’une sonde de stimulation en position ventriculaire gauche
 

Bu-Chun Zhang, Kai Tang, Ya-Wei Xu
Department of Cardiology, Shanghai Tenth People’s Hospital, Tongji University School of Medicine, 301, Yanchang Road, Shanghai 200072, China 

Corresponding author. Fax: +086-21-66301004.
Summary
Background

Cardiac resynchronization therapy (CRT) improves clinical outcome in selected patients with advanced congestive heart failure. The Overlay Ref technique may facilitate the procedure for implanting left ventricular (LV) pacing leads to deliver CRT.

Aim

To assess the feasibility of deploying a LV pacing lead into a coronary sinus side branch guided by Overlay Ref.

Methods

Data from 88 consecutive patients who met the CRT implantation criteria in our hospital between 28 November 2007 and 30 December 2009 were randomly assigned to two groups. Forty-four patients underwent CRT device implantation using Overlay Ref to guide target vein selection and advance a specifically designed pacing lead into the target vein (Overlay Ref group); 44 patients were conventionally implanted (control group).

Results

LV lead implantation was successful in all patients. Mean CRT total procedure times (skin-to-skin) were: Overlay Ref group, 80.7±18.0min; control group, 98.5±32.2min; p =0.029. Mean placement of LV pacing lead into target vein times were: Overlay Ref group, 16.2±7.7min; control group, 36.4±23.4min; p =0.004. Mean total fluoroscopy times were: Overlay Ref group, 13.6±4.3min; control group, 23.8±15.7min; p =0.007. Mean LV lead fluoroscopy times were: Overlay Ref group, 5.7±2.9min; control group, 14.4±4.6min; p =0.003. No major complications occurred.

Conclusions

Overlay Ref facilitates location of and entry into the coronary sinus, and shortens the duration of LV pacing lead implantation into the target vein.

The full text of this article is available in PDF format.
Résumé
Objectif

Évaluer la faisabilité d’une sonde ventriculaire gauche implantée dans le sinus coronaire, guidée par une technique appropriée.

Justification

La resynchronisation cardiaque (CRT) est une thérapeutique éprouvée de l’insuffisance cardiaque congestive avérée, à condition de respecter les indications. Cependant, la procédure d’implantation des sondes ventriculaires gauches varie en fonction des opérateurs et de leur technique. Il reste à réduire le temps des procédures. Nous proposons une nouvelle technique pour faciliter cette procédure de mise en place de la sonde en position ventriculaire gauche. Cette étude rapporte notre expérience initiale dans l’implantation du traitement de l’insuffisance cardiaque congestive.

Méthode

Les données de 88 patients consécutifs répondant aux critères d’implantation ont été assignées de façon randomisée dans deux groupes, au cours d’une période s’étendant du 28 novembre 2007 au 30 décembre 2009 : 44 patients ont bénéficié d’une implantation selon la nouvelle technique permettant de guider la veine coronaire cible et de déployer ainsi la sonde ventriculaire gauche dans celle-ci (groupe de référence) ; les témoins étaient constitués de 44 patients, traités et implantés de façon conventionnelle (groupe témoin).

Résultats

La mise en place de la sonde ventriculaire gauche a été obtenue avec succès dans tous les cas. La durée moyenne de la procédure d’implantation (peau à peau) dans le groupe évalué a été de 80,7±18 comparée à 98,5±32,2minutes dans le groupe témoin (p =0,029). Le délai moyen d’obtention de la stimulation ventriculaire gauche a été respectivement de 16,2±7,7 et 36,4±23,4minutes (p =0,004). Le temps moyen d’irradiation per procédure était respectivement de 13,6±4,6 et 23,8±15,7minutes (p =0,007). Le déploiement de la sonde en position ventriculaire gauche a abouti à une irradiation per procédurale de 5,7±2,9 et 14,4±4,6minutes (p =0,003). Il n’y a pas eu de complication majeure recensée.

Conclusion

Cette nouvelle technique d’implantation d’une sonde en position ventriculaire gauche pour resynchronisation cardiaque a permis un repérage plus aisé du sinus coronaire, et une réduction tant du délai de cathétérisme de celle-ci que de la durée d’irradiation en salle de cathétérisme.

The full text of this article is available in PDF format.

Keywords : Left ventricular pacing, Overlay Ref technique, Computed tomography, Implantation, Congestive heart failure

Mots clés : Stimulation ventriculaire gauche, Scanner, Implantation, Insuffisance cardiaque congestive

Abbreviations : CRT, CS, DSA, LV


Background

Multiple randomized clinical studies have suggested that patients in heart failure with left bundle branch block may benefit from cardiac resynchronization therapy (CRT) [1, 2, 3]. One of the greatest challenges to this therapy is the technical difficulty encountered with implantation of the left ventricular (LV) lead. Implantation of LV pacing leads to deliver CRT typically involves locating and entering the coronary sinus (CS), performing coronary venous angiography, selecting a target vein and advancing a specifically designed pacing lead into the target vein. To overcome this problem, a new technique named Overlay Ref has been introduced into clinical electrophysiology; this technique can overlap the reference images on the real-time fluorescence monitor, to facilitate implantation of LV pacing leads. This article reports our initial experiences with implantation of LV pacing leads guided by Overlay Ref.

Methods
Patient selection

Between 28 November 2007 and 30 December 2009, biventricular pacemaker implantation was performed in 88 consecutive candidates for CRT, who were randomized in a 1:1 ratio based on a permuted blocked randomization list, to provide an approximate balance between treatment groups. The randomization list was stratified by site. Randomization was done by opening sealed allocation envelopes (indicating ‘Overlay Ref’ or ‘conventional’ treatment).

Forty-four randomized patients underwent LV lead placement using the Overlay Ref technique to control the tip of a 0.014-inch guidewire (Medtronic Inc., Minneapolis, MN, USA). The control group was made up of 44 randomized patients scheduled for CRT during the same time period. The Overlay Ref technique was not used in the control group. Standard 0.014-inch guidewires were used in the control group, as well as CS venography. The right ventricular and right atrial leads were initially placed in a standard fashion. The CS leads were commercially available over-the wire leads (Medtronic Inc.).

The institutional review board of our institution approved the study protocol. Each patient signed a written consent form before treatment.

Inclusion criteria were as follows: symptomatic heart failure with idiopathic dilated or ischaemic cardiomyopathy; New York Heart Association functional class III or IV; LV end-diastolic diameter greater or equal to 60mm; QRS width greater or equal to 120ms (measured on surface electrocardiogram leads); and obvious mitral regurgitation by echocardiography.

Device

Impulse generators: Insync 8040, Insync III 8042 (Medtronic Inc.). LV leads: 2188, 2187, 4189, 4191, 4193 (Medtronic Inc.); Attain® OTW 6215 (Medtronic Inc.).

Implantation procedure: Overlay Ref group

After informed consent was obtained, implantation was performed for the patient under local anaesthesia and conscious sedation. After puncturing the left subclavian vein, two steerable hydrophilic guides were inserted. The short guide was used to introduce the atrial and right ventricular leads, and the long guide to introduce the LV lead delivery system.

Location of the coronary sinus ostium

One of the key steps in the implantation procedure for CRT is access to the CS. We used an electrophysiological guided approach to locate the CS ostium. The standard diagnostic CS electrophysiology catheter was used to cannulate the CS. When the catheter was put into the CS, the delivery sheath was then slipped over the catheter inside the CS as deeply as possible and the catheter was exchanged for a balloon catheter to perform an occlusive CS venogram; it then remained in the ostium to secure access (Figure 1) [4].



Figure 1


Figure 1. 

(a) coronary sinus angiography (b) ideal left ventricular side branch was obtained (in anterior posterior view).

Zoom

Left ventricular lead implantation

The LV lead was selected on the basis of the anatomy of the CS branches. For large veins with few tortuosities, a large bore lead (2187) was used. For medium-sized or small veins with significant tortuosities, over-the-wire leads (mostly 4193) were selected. The lead was placed preferentially into the lateral or posterolateral vein using percutaneous transluminal coronary angioplasty (Figure 2) [5, 6]. Sensing and pacing thresholds were determined, along with the pacing threshold for phrenic nerve stimulation. After lead measurement, the sheath was removed and the right ventricular and right atrial leads were implanted. The right ventricular lead was generally placed at a site anatomically remote from the LV lead, with good electrical separation and, preferably, early in the QRS complex.



Figure 2


Figure 2. 

Typical lead placement in the anterior vein in (a) left anterior oblique view and (b) posterior-anterior view. In the same patient, the lead was positioned in the lateral vein by the over-the-wire technique, depicted in (c) left anterior oblique view and (d) posterior-anterior view. Photo reference from Sack et al [5].

Zoom

The digital subtraction angiography (DSA) system (AXIOM Artis dFC equipped with VB31E software package including Overlay Ref technique; Siemens, Erlangen, Germany) was used to perform selective CS angiography in both the right anterior oblique 30° view and the left anterior oblique 45° view. Meanwhile, the images were recorded in order to direct lead placement. The main branches of the CS are the great vein and the lateral, posterolateral, posterior and middle veins (Figure 3, Figure 4). The snapshot that showed the clearest contour of CS was drawn from the venography movie and transferred to the reference monitor; this served as the reference image. When the ‘Automap’ button was pressed, the DSA machine automatically restored the same zoom stage, matrix size and patient position as the reference image. The ‘Overlay’ button was then pressed, and the (inverted) reference image was faded into the live fluoroscopic image, so that the position of the LV lead could be located directly over the inverted and faded reference image. Hence, Overlay Ref was able to render the position of the LV lead during the implantation procedure. CS landmarks were chosen from the CS ostia region (Figure 3, Figure 4).



Figure 3


Figure 3. 

A percutaneous transluminal coronary angioplasty wire was placed into the target vein, guided by the Overlay Ref technique, in left anterior oblique view. (a) Reference image. (b) Fluoroscopic image.

Zoom



Figure 4


Figure 4. 

An over-the-wire left ventricular lead was placed into the target vein, guided by the Overlay Ref technique, in left anterior oblique view. (a) Reference image. (b) Fluoroscopic image.

Zoom

Implantation procedure: control group

The 44 control patients underwent routine CRT device implantation during the same period of time, as previously described [7]. Selective CS angiography at both the right anterior oblique 30° view and the left anterior oblique 45° view were performed. No reference images were recorded and LV pacing leads were guided by the use of conventional fluoroscopy.

Statistical analysis

Statistical analyses were performed using SPSS13.0 for Windows (Apache Software Foundation, SPSS Inc., Chicago, IL, USA). Continuous variables were expressed as means±standard deviations and compared using Student’s t -test. Absolute numbers and percentages were compared using the chi-square test or Fisher’s exact test. A p value less than 0.05 was considered statistically significant.

Results
Left ventricular lead placement

CS access and LV lead placement were successfully performed in all Overlay Ref and control group patients. There were two coronary venous anatomical difficulties in the Overlay Ref group: one patient had an atypically located CS ostium; another patient had undergone previous valve surgery that distorted the atrial anatomy. In 53 patients, the target CS side branch was located in the posterolateral wall of the left ventricle; in 32 patients the ideal CS side branch was considered to be in the lateral LV wall; in three patients the anterior LV wall was used due to the lack of acceptable pacing thresholds in other locations.

Comparison between Overlay Ref and control groups

Patient clinical characteristics were similar in the two groups (Table 1). Implantation procedural data are shown in Table 2. Using the Overlay Ref technique, placement of the LV pacing lead into the target vein took 16.2±7.7min. Total CRT procedure duration (skin-to-skin) was 80.7±18.0min, total fluoroscopy time was 13.6±4.3min and LV lead fluoroscopy time was 5.7±2.9min. Compared with the control group, Overlay Ref was associated with significantly shorter times for LV lead placement.

There were no implantation procedure-related deaths, CS dissections or LV lead dislodgements.

Discussion

CRT, by simultaneous pacing of the right and left ventricles, can improve symptoms and reduce mortality. With the publication of the CARE-HF trial, CRT gained further acceptance as an effective intervention for patients with advanced heart failure and LV systolic dysfunction [1]. The major problem with this therapy has been the implantation of LV pacing leads. Placing the LV lead at the target site involves cannulating the CS, performing coronary venous angiography and placing the LV lead to achieve adequate resynchronization.

The procedure is complicated. Firstly, some patients with advanced heart failure symptoms and depressed cardiac function cannot withstand this time-consuming procedure because they need to lie still during the implantation. Secondly, previous studies have suggested that for the majority of patients, the optimal LV pacing site is considered to be a lateral or posterior lateral CS tributary [8, 9]. However, as far as the tributaries of the CS are concerned, there are all kinds of variation. Moreover, the varied anatomy of the coronary venous system may be further complicated by distortion of the anatomy of heart failure patients [10]. As a result, it is difficult to advance the LV lead to the desired position inside the vein. Thirdly, the available LV leads are still not ideal. For example, the model 2187 Attain LV 7F (Medtronic, Inc.) is a stylet-driven lead with a preshaped stylet used to catheterize the CS without the use of a left heart delivery system; it is not only hard but also has a large curvature, and is therefore not appropriate for the thin branch. Finally, CRT is usually used for patients with severe heart failure. In such cases, the heart is significantly enlarged, which makes CS identification or cannulation more difficult.

This study demonstrates the feasibility of the Overlay Ref technique for guiding implantation of LV pacing leads for the treatment of heart failure. Our previous study showed that the Overlay Ref technique can facilitate the catheter ablation of paroxysmal atrial fibrillation and can help to standardize the landmarks selection procedure [11]. Using the Overlay Ref technique, the total duration of the procedure in the present study was shortened to 80.7±18.0minutes and we rarely saw intermittent contrast injection through the guide catheter. This suggests that the Overlay Ref technique might improve the accuracy of LV pacing lead implantation and shorten the duration of the total procedure.

The coronary venous anatomy may limit placement of the LV lead and the ability to pace a defined region of the LV for CRT in some cases. The use of the Overlay Ref technique might help to overcome some of the technical difficulties, such as cannulation of small veins, or track the lead to a more distal portion of the coronary vein. In addition, there was no learning curve associated with this technique.

The accuracy of Overlay Ref depends on the image quality of the venography of the CS. Only when the reference image outlines the contours of the CS clearly can Overlay Ref locate the catheter exactly. It should also be noted that respiratory motion and cardiac beat motion can change the catheter position, thus affecting the precision of Overlay Ref.

One reason for the precision of Overlay Ref could be that it is a flat, two-dimensional technique, rather than a three-dimensional mapping system. In fact, when Overlay Ref is started, the DSA machine automatically restores to the same zoom stage, matrix size and patient position as the reference image, ensuring precise location of the catheter guided by the flat fluorescence.

The limitation of this technique is that the Overlay Ref image cannot be stored and reviewed after implantation procedure due to technical reasons.

Study limitation

A limitation of this study was the small number of patients included, Although the study showed that the Overlay Ref procedure is safe and feasible, a larger, randomized trial is warranted.

Conclusion

The Overlay Ref technique is a safe, feasible and efficient tool for assisting LV lead placement during CRT, with the advantage of timesavings.

Conflicts of interest statement

None.

References

Cleland J.G., Daubert J.C., Erdmann E., and al. The effect of cardiac resynchronization on morbidity and mortality in heart failure N Engl J Med 2005 ;  352 : 1539-1549 [cross-ref]
Hunt S.A., Abraham W.T., Chin M.H., and al. ACC/AHA 2005 Guideline Update for the Diagnosis and Management of Chronic Heart Failure in the Adult: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Update the 2001 Guidelines for the Evaluation and Management of Heart Failure): developed in collaboration with the American College of Chest Physicians and the International Society for Heart and Lung Transplantation: endorsed by the Heart Rhythm Society Circulation 2005 ;  112 : e154-e235
Moss A.J., Hall W.J., Cannom D.S., and al. Cardiac-resynchronization therapy for the prevention of heart-failure events N Engl J Med 2009 ;  361 : 1329-1338 [cross-ref]
Makaryus A.N., Boutis L., Goldner B., and al. Coronary venous angioplasty and stenting for biventricular pacemaker left ventricular lead implantation J Invasive Cardiol 2007 ;  19 : E128-E130
Sack S., Heinzel F., Dagres N., and al. Stimulation of the left ventricle through the coronary sinus with a newly developed ‘over the wire’ lead system – early experiences with lead handling and positioning Europace 2001 ;  3 : 317-323 [cross-ref]
Yu C.M., Miu R. A new technique for the transvenous implantation of the over-the-wire left ventricular pacing lead in a patient with heart failure J Interv Card Electrophysiol 2002 ;  7 : 189-191 [cross-ref]
Daubert J.C., Ritter P., Le Breton H., and al. Permanent left ventricular pacing with transvenous leads inserted into the coronary veins Pacing Clin Electrophysiol 1998 ;  21 : 239-245 [cross-ref]
Albertsen A.E., Nielsen J.C., Pedersen A.K., and al. Left ventricular lead performance in cardiac resynchronization therapy: impact of lead localization and complications Pacing Clin Electrophysiol 2005 ;  28 : 483-488 [cross-ref]
Butter C., Auricchio A., Stellbrink C., and al. Effect of resynchronization therapy stimulation site on the systolic function of heart failure patients Circulation 2001 ;  104 : 3026-3029 [cross-ref]
Singh J.P., Houser S., Heist E.K., and al. The coronary venous anatomy: a segmental approach to aid cardiac resynchronization therapy J Am Coll Cardiol 2005 ;  46 : 68-74 [cross-ref]
Tang K., Zhao D.D., Zhang J.Y., and al. Initial experience with circumferential pulmonary vein isolation guided by Overlay Ref and CartoMerge in the treatment of paroxysmal atrial fibrillation Chin Med J (Engl) 2010 ;  123 : 1269-1272



© 2010  Elsevier Masson SAS. All Rights Reserved.
EM-CONSULTE.COM is registrered at the CNIL, déclaration n° 1286925.
As per the Law relating to information storage and personal integrity, you have the right to oppose (art 26 of that law), access (art 34 of that law) and rectify (art 36 of that law) your personal data. You may thus request that your data, should it be inaccurate, incomplete, unclear, outdated, not be used or stored, be corrected, clarified, updated or deleted.
Personal information regarding our website's visitors, including their identity, is confidential.
The owners of this website hereby guarantee to respect the legal confidentiality conditions, applicable in France, and not to disclose this data to third parties.
Close
Article Outline