Archives of cardiovascular diseases
Volume 104, n° 3
pages 147-154 (mars 2011)
Doi : 10.1016/j.acvd.2010.12.004
Received : 10 September 2010 ;  accepted : 17 December 2010
Spatial distribution of neo-intimal hyperplasia 6 months after zotarolimus-eluting stent implantation, analysed by optical coherence tomography
Distribution spatiale de l’hyperplasie néo-intimale six mois après implantation de stents actifs au zotarolimus analysée par tomographie par cohérence optique
 

Geraud Souteyrand a, b, d, , Sebastien Levesque a, Lemlih Ouchchane b, c, Laurent Sarry b, Romain Eschalier a, Jean-René Lusson a, b, Pascal Motreff a, b
a Department of Cardiology, Clermont-Ferrand University Hospital, Clermont-Ferrand, France 
b ERIM-EA3295, University of Auvergne, Clermont-Ferrand, France 
c Department of Biostatistics, Faculty of Medicine, University of Auvergne, Clermont-Ferrand, France 
d Department of Cardiology, Gabriel Montpied Hospital, CHU Clermont-Ferrand, 63003 Clermont-Ferrand, France 

Corresponding author.
Summary
Background

Optical coherence tomography is a high-resolution imaging technology that allows in vivo assessment of neointimal hyperplasia and strut coverage after coronary stenting.

Aim

Assessment of spatial distribution of healing, 6 months after zotarolimus-eluting stent implantation.

Methods

Forty-two zotarolimus-eluting stents were monitored by optical coherence tomography 6 months after implantation. Mean neointimal strut coverage thickness and percentage of neointimal hyperplasia were measured every millimetre. Non-covered strut ratios were assessed on each slice. In addition, the spatial distribution of neointimal hyperplasia and strut coverage were analysed longitudinally on five stent segments and axially on each slice.

Results

There were no clinical events at 6 months under dual antiplatelet therapy. The optical coherence tomography analysis showed a mean neointimal hyperplasia thickness of 333±147μm and neointimal hyperplasia obstruction of 36.1±12.3%. The percentage of covered struts at 6 months was very high (98.9%). Only 6/745 slices analysed (0.8%) had non-covered strut ratios exceeding 30%. There was no significant heterogeneity in either longitudinal or axial neointimal hyperplasia distribution. No thrombi were observed.

Conclusion

This optical coherence tomography study found relatively constant neointimal hyperplasia thickness, regardless of the zotarolimus-eluting stent length or diameter. This spatially homogeneous neointimal hyperplasia was associated with near-total coverage of all struts, 6 months after implantation.

The full text of this article is available in PDF format.
Résumé
Contexte

L’imagerie par cohérence optique (OCT) est une imagerie de haute résolution permettant l’analyse in vivo de l’hyperplasie néo-intimale et de la couverture des mailles des stents.

Objectif

Analyse quantitative de la couverture néo-intimale de stents au zotarolimus à six mois et de sa distribution spatiale.

Méthode

Quarante-deux stents au zotarolimus ont été contrôlé par OCT six mois après leur implantation. L’épaisseur moyenne de la couverture néo-intimale des mailles et le pourcentage d’hyperplasie néo-intimale ont été mesurés tous les millimètres. Sur chaque coupe, le ratio de mailles non couvertes est mesuré. La distribution spatiale de l’hyperplasie néo-intimale et de la couverture du stent est analysée longitudinalement sur cinq segments du stent et axialement sur chaque coupe.

Résultats

Il n’y a pas eu d’évènements cliniques à six mois sous double anti-agrégation plaquettaire. L’analyse OCT retrouve une épaisseur moyenne d’hyperplasie néo-intimale de 333±147μm et une obstruction néo-intimale de 36,1±12,3 %. Le pourcentage de mailles couvertes à six mois est élevé de 98,9 %. Seulement six coupes sur les 745 analysées (0,8 %) ont un ratio de mailles non couvertes/couvertes de plus de 30 %. La distribution de l’hyperplasie néo-intimale est homogène en axial et en longitudinal. Aucun thrombus n’a été visualisé.

Conclusion

Cette analyse OCT retrouve une épaisseur d’hyperplasie néo-intimale relativement homogène dans le stent, quelle que soit sa longueur ou son diamètre. Cette homogénéité d’hyperplasie néo-intimale est associée à une couverture quasi totale des mailles six mois après l’implantation des stents.

The full text of this article is available in PDF format.

Keywords : Optical coherence tomography, Neointimal hyperplasia, Zotarolimus-eluting stent

Mots clés : Imagerie par cohérence optique, Hyperplasie néo-intimale, Stent actif au zotarolimus

Abbreviations : DES, NIH, OCT, PES, QCA, SES, ZES


Background

Randomised studies have shown that DESs significantly reduce clinical restenosis rates compared with bare-metal stents [1, 2, 3, 4]. This benefit is due to inhibition of intimal neoproliferation [5]. Animal and autopsy studies, however, have shown this effect to be associated with delayed or deficient arterial healing. It is therefore advisable to continue dual antiplatelet therapy beyond the first month post stenting, in order to avoid late intrastent thrombosis [6]. In a series of 8000 DESs, Wenaweser et al. reported a consistent 0.6% annual late thrombosis rate over the first 4 years post stenting [7].

Intrastent thrombosis is a multifactorial phenomenon [8], but one significant factor is failure of stent–strut re-endothelialization [9]. Post-mortem studies of stented subjects [10] have reported deficient re-endothelialization to be more frequently associated with late intrastent thrombosis than with other causes of death [11, 12].

OCT is a high-resolution (around 10microns) imaging technology that is particularly well adapted for the study of the most superficial layers of the vessel wall and for strut-by-strut stent analysis. Several recent studies have focused on OCT analysis of neointimal coverage in bare-metal and first-generation DESs [13, 14, 15, 16].

The present study used OCT to quantify and analyse the spatial distribution of NIH and strut coverage, 6 months after ZES implantation.

Methods
Study population and methods

Between October 2006 and September 2007, 25 patients (21 men, four women) gave consent and were included in this prospective, observational study. These 25 patients were selected from the 220 patients who underwent ZES implantation in our centre during the study period, using the following inclusion criteria: consent and feasibility of OCT. Exclusion criteria were left main coronary artery stenosis, bypass lesion, ostial stenosis, renal insufficiency, acute coronary syndrome, stent overlapping, contraindication for DES and OCT limitations. OCT limitations were tortuous and calcified vessels, proximal lesion or large coronary diameter (>3.5mm) preventing the good saline flush downstream of an occlusion balloon that is necessary for image quality. The 25 patients underwent native coronary stenting, using a total of 42 ZESs. Angiographic analyses were done at baseline and after angioplasty. The angiograms were evaluated by a QCA system (QuantCor, CASS II, Siemens, Erlangen, Germany). An angiographic control with OCT was scheduled at 6 months post stenting. Restenosis was defined as>50% lumen narrowing on QCA.

Optical coherence tomography procedure

Optical coherence tomography at 6 months post stenting immediately followed the control angiography. A 6F catheter was used to inject 30UI/kg of unfractionated heparin intra-arterially. A 0.014-inch guide was introduced into the vessel and positioned distally to the stent. A Helios™ coaxial occlusion balloon catheter (LightLab Imaging, Westford, MA, USA) was then introduced along the guide across the vessel until the balloon marker was at the distal extremity of the stent. The guide was then withdrawn and replaced by a 1.4F optic fibre, connected up to the OCT system (M2 version, LightLab Imaging, Westford, MA, USA) and inserted through the balloon until distal to the stent. The occlusion balloon was then withdrawn until proximal to the DES and inflated to between 0.4 and 0.7atm. Physiological saline was then injected downstream of the occlusion balloon via its coaxial catheter; 30mL were injected during each pullback. OCT gives a clear image once the lumen is sufficiently transparent. Automatic light-source pullback then began (1mm/s, with 15 image/s acquisitions). A 30-mm DICOM-format video recording of the artery was made, including the stented segment. The balloon was then deflated and the saline injection stopped. Several pullbacks were sometimes needed for analysis of the longest stents. Pullback was repeated until perfect visualization of the whole length of all stents was obtained.

Optical coherence tomography analysis

Optical coherence tomography images were selected from the DICOM pullback recordings every millimetre (every 15 images) over the entire stented segment and were analysed by two independent operators (Figure 1).



Figure 1


Figure 1. 

Stent analysis on optical coherence tomography millimetric images: (a) qualitative strut analysis (A. covered; B. non-covered; C. malapposed; D. facing collateral); (b) quantitative analysis (neointimal hyperplasia area [green]; lumen area [red]; coverage thickness per strut [red arrows]); (c) longitudinal analysis in five segments of equal length (P2: proximal; P1: proximal-medial; M: medial; D1: distal-medial; D2: distal).

Zoom

Each strut was classified into one of four categories:

well apposed to vessel wall with apparent neointimal coverage (A);
well apposed to vessel wall but without neointimal coverage (non-coverage being defined as absence of any visible structure between lumen and vessel on OCT, with strut reflection) (B);
malapposed to vessel wall without neointimal coverage (malapposition being defined as>110μm distance between strut reflection and vessel wall, corresponding to OCT axial resolution+ZES strut thickness) (C);
located at a bifurcation (D).

Struts in class A were assigned their measured coverage thickness in micrometer; otherwise, the assigned value was 0.

Strut information was used to produce cross-sectional criteria:

overall NIH quantification, in terms of mean strut coverage thickness;
cross-sectional heterogeneity of NIH, in terms of the coefficient of variation (i.e. standard deviation divided by mean) of strut coverage thickness;
global coverage deficiency, in terms of the non-covered strut ratio, defined as the proportion of struts in categories B, C and D.

NIH quantification was also assessed in each slice from both lumen area (in millimetre square) and stent area (in millimetre square), computing percentage NIH area, defined as NIH obstruction (%)=(stent area–lumen area)/SA*100.

Longitudinal NIH distribution was assessed using the mean values for each of the aforementioned criteria on each of five predefined longitudinal stent segments of equal length (proximal, proximal-medial, medial, distal-medial and distal).

Statistical analysis

Categorical variables are described as frequencies and percentages. Continuous variables are displayed as means±standard deviations. Longitudinal NIH distribution used a mixed linear model to account for segment nesting within a given stent. Longitudinal effect (i.e. possible longitudinal NIH distribution heterogeneity) was assessed through a global F statistic, completed, as appropriate, by post-hoc multiple comparison of the five predefined stent segments.

The analyses were performed using SAS 9.2 software (SAS Institute, Inc., Cary, NC, USA).

Results
Population characteristics

Population characteristics are given in Table 1. The 25 included patients (mean age, 59.7±10.1 years) were stented at severe coronary stenosis sites (77.5±8.9% diameter stenosis on QCA). Minimum lesion diameter and length were 0.67±0.29mm and 14.45±5.68mm, respectively, with a reference diameter of 2.99±0.33mm.

A total of 42 ZESs (1.68 per patient) were implanted in de novo lesions, with good initial angiographic results. Stent diameter ranged from 2.5 to 4.0mm (mean, 3.01±0.34mm) and length ranged from 8 to 30mm (mean, 18.40±6.30mm). Implantation used a mean balloon pressure of 14.0±2.4 atm, usually without predilatation (81%). Mean acute diameter gain was 2.25±0.34mm.

Clinical and angiographic follow-up

There were no complications secondary to angioplasty during the 6-month follow-up period under dual antiplatelet therapy or during the OCT procedure.

At 6 months, there were no intrastent binary restenoses (angiographic diameter stenosis>50%) requiring further revascularization. Mean restenosis diameter was 20.2±8.7% with a mean late-loss of 0.52±0.30mm on QCA.

Optical coherence tomography analysis

Optical coherence tomography was feasible in all 25 patients and 69 pullbacks were performed to analyse the 42 stents (1.64 per stent). In all, 745mm of stent were studied, and 10169 struts were visualized. Strut coverage thickness was assessed by two independent operators. Concordance on quantitative measurements was virtually perfect, with an intraclass correlation coefficient of 0.99. The value submitted for analysis was the mean of the two observers’ estimates. Strut coverage was likewise classified in the four exclusive categories by the two independent observers, again with virtually perfect agreement (kappa=0.99). In case of disagreement, the category to be adopted for analysis was determined by a third observer. Results are shown in Table 2.

Neointimal hyperplasia

Mean stent area was 8.31±2.15 m2 and mean lumen area was 5.46±2.17 mm2. Mean NIH area was 2.85±0.82 mm2, with a percentage obstruction of 36.1±12.3%. Considering lumen area and stent area as disks, with diameters calculated as diameter=2*√(area/π), mean stent diameter, lumen diameter and NIH thickness were 3.22±0.41mm, 2.57±0.52mm and 0.65±0.23mm, respectively. Mean coverage thickness measured on each strut (n =10,169) was 333±147μm for 42 ZESs.

Neointimal strut coverage

The incidence of non-covered struts was very low at 6 months (111/10,169; non-covered strut ratio=1.1%). Only 65 struts were observed to be remote from the coronary wall because of malapposition (0.5%) or a collateral branch (0.2%). On most slices, all struts were covered (699/745, 93.8%). No thrombi were observed in any of the 42 stents.

The rate of cross sections with a non-covered strut ratio>0.3 was very low (6/745, 0.8%). On the 745 slices, a converse relation was also found between non-covered strut ratio and strut coverage thickness or NIH percentage obstruction (Figure 2).



Figure 2


Figure 2. 

Relationship between strut coverage thickness (μm), neointimal hyperplasia (NIH) obstruction (%) and non-covered strut ratio, studied on 745 optical coherence tomography slices (699 slices with all struts covered; 28 slices with non-covered strut [NCS] ratio<0.15; 12 slices with NCS ratio between 0.15 and 0.30; six slices with NCS ratio>0.30).

Zoom

Spatial distribution of neointimal hyperplasia

Figure 3 shows mean stent area, lumen area, NIH area, strut coverage thicknesses, NIH obstruction and non-covered strut ratio on the 745 slices and in each segment. There were no significant differences in longitudinal distribution with regard to any of the criteria studied (p >0.05).



Figure 3


Figure 3. 

Spatial distribution of neointimal hyperplasia (NIH) and strut coverage, 6 months after zotarolimus-eluting stent implantation (n =42), from quantitative optical coherence tomography analysis. A. Mean stent area, NIH area (NIHA) and lumen area (LA) (mm2) in all 745 slices; mean areas in five longitudinal stent segments (P2: proximal; P1: proximal-medial; M: medial; D1: distal-medial; D2: distal). B. Mean strut coverage thickness (μm) in all cross sections and in five longitudinal stent segments (bars), and neointimal hyperplasia (NIH) percentage obstruction (points). C. Mean non-covered strut ratio in all cross sections and in five longitudinal stent segments.

Zoom

Mean cross-sectional NIH heterogeneity was very low (0.375±0.107). No longitudinal effect was found on this criterion (p =0.662), suggesting a high level of longitudinal NIH homogeneity.

Discussion

Optical coherence tomography is a particularly effective, high-resolution, imaging technique for analysing the more superficial coronary layers and controlling stenting. The present study confirms the feasibility of the technique in selected lesions and the good reproducibility of its qualitative and quantitative analysis of neointimal stent coverage. All 42 stents initially selected for control were analysed.

Previous angiographic findings showed greater late-loss after implantation of a ZES compared with a first-generation DES (SES or PES). In 11 randomised trials, Pocock et al. reported a late-loss of less than 0.21mm with SESs, between 0.30 and 0.49mm with PESs, between 0.60 and 0.61mm with ZESs, and between 0.80 and 1.06mm with bare-metal stents [17].

Comparative OCT studies have confirmed greater NIH after ZES implantation. Kim et al. [18] observed a mean NIH thickness of 251.2 vs 85.5μm (p <0.001) and a percentage NIH area of 27.9 vs 11.2% (p <0.001) at 9 months with ZESs and SESs, respectively. In a previous study [19], we found a significant difference between ZESs and PESs: mean strut coverage thickness 333 vs 154μm (p <0.001) and percentage NIH area 38 vs 17% (p <0.001), respectively.

Guagliumi et al. reported a similar difference in the ODESSA trial [20], in which 77 patients (189 stents) were randomised and monitored by OCT 6 months after long-lesion stenting requiring overlapping stents: mean percentage NIH was 19.3% with SESs, 31.5% with PESs, 45.2% with ZESs and 57.8% with bare-metal stents.

The present mean 0.333mm strut coverage thickness (to be multiplied by two to obtain the OCT diameter late-loss value) agrees with the late-loss values reported in the ENDEAVOR II and IV studies (0.61 and 0.67mm, respectively [21]). It was noteworthy that coverage thickness was independent of stent diameter or length (Figure 4), making percentage obstruction greater with small-diameter stents (Figure 4D); with a constant strut coverage thickness of 0.333mm, obstruction with a 2.5mm diameter stent theoretically approximates 50%.



Figure 4


Figure 4. 

Relationship between mean strut coverage thickness, neointimal hyperplasia (NIH) obstruction measured by optical coherence tomography and stent length (A, B) and stent diameter (C, D), 6 months after zotarolimus-eluting stent implantation (n =42).

Zoom

ZES strut coverage was virtually complete by 6 months. In other series [18, 19, 20], as in the present, no thrombi were ever reported and ZES strut coverage was almost total on OCT. First-generation DES coverage was less at 6 or 9 months, with 8 to 12% of SES struts remaining non-covered [14, 18, 20], and was still incomplete at 24 months (5% non-covered SES struts) [22].

The originality of the present study lay in its analysis of the spatial distribution of NIH at six months post stenting. NIH proved homogeneous in terms of both coverage thickness and non-covered strut ratio, longitudinally and axially. There were no significant differences between the five longitudinal stent segments, while NIH variability was very low on all slices. These data suggest that, in the present series, the coronary response to zotarolimus was homogeneous.

Non-uniform NIH was observed in previous intravascular ultrasound studies of SESs, attributed by the authors to possible non-uniform stent distribution, stent under-expansion [23, 24] or a heterogeneous local response to the drug [25].

In the present series, the homogeneous reaction may be explained by the non-occurrence of stent under-expansion (confirmed by stent area measurement), lesion selection for OCT feasibility, which excluded more complex plaques, and the absence of overlapping. The specific impact of zotarolimus or of the stent platform will need to be investigated in larger-scale, comparative studies.

Anatomopathology studies reported delayed endothelialization after DES implantation. Post-mortem studies on first-generation DESs found NIH to be sometimes heterogeneous, with a converse relationship between non-covered strut ratio and NIH thickness [11]. Finn et al. reported that a non-covered strut ratio of>0.30 in a given cross section increased the intrastent thrombosis risk ninefold. The most recent OCT studies have adopted this prognostic index [16, 18]. We also found in the present series a relationship between NIH and non-covered strut ratio (Figure 2).

Deficient strut coverage on OCT can be considered to be a thrombosis risk marker. There is, however, a discrepancy between the rate of non-covered active stents at several months post implantation and the much lower rate of late thrombosis. Such “vulnerable” stents may be at increased risk of thrombosis in case of termination of or resistance to antiplatelet therapy. If a correlation could be demonstrated between OCT findings and large-scale clinical study results, OCT might be able to guide the duration of dual antiplatelet therapy, comparing theoretical late thrombosis risk for the various kinds of stent.

Study limitations

There are some limitations to this study. The series comprised only 25 patients and analysis concerned only 42 stents. Lesions were selected for OCT (M2 version, LightLab Imaging) feasibility. The most complex stenoses (bifurcations, overlapping, in tortuous or calcified coronary arteries) were excluded. Moreover, despite its good performance, OCT cannot be assimilated to in vivo histology. Certain aspects of coverage may, in fact, be due to fibrin or microthrombus. Conversely, endothelialization may be underestimated on OCT, when thin and translucent.

Conclusion

OCT provides precise measurement of NIH volume and stent–strut coverage. The present study found constant, homogeneous NIH, 6 months after ZES implantation. The mean strut coverage thickness of 333μm did not lead to significant angiographic stenosis or recurrent angina, but could be harmful for the smallest stent diameters. On the other hand, ZES strut coverage at 6 months was almost total and homogeneous. The converse relationship between mean strut coverage thickness and non-covered strut ratio reported in anatomopathology was confirmed on OCT. The relevance of these findings to the issue of intrastent thrombosis, especially on termination of antiplatelet biotherapy, should be investigated in larger-scale clinical studies.

Disclosure of interest

The authors declare that they have no conflicts of interest concerning this article.

References

Fajadet J., Wijns W., Laarman G.J., and al. Randomized, double-blind, multicenter study of the Endeavor zotarolimus-eluting phosphorylcholine-encapsulated stent for treatment of native coronary artery lesions: clinical and angiographic results of the ENDEAVOR II trial Circulation 2006 ;  114 : 798-806 [cross-ref]
Moses J.W., Leon M.B., Popma J.J., and al. Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery N Engl J Med 2003 ;  349 : 1315-1323 [cross-ref]
Stone G.W., Ellis S.G., Cox D.A., and al. One-year clinical results with the slow-release, polymer-based, paclitaxel-eluting TAXUS stent: the TAXUS-IV trial Circulation 2004 ;  109 : 1942-1947 [cross-ref]
Stone G.W., Ellis S.G., Cox D.A., and al. A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease N Engl J Med 2004 ;  350 : 221-231 [cross-ref]
Babapulle M.N., Joseph L., Belisle P., and al. A hierarchical Bayesian meta-analysis of randomised clinical trials of drug-eluting stents Lancet 2004 ;  364 : 583-591 [cross-ref]
Iakovou I., Schmidt T., Bonizzoni E., and al. Incidence, predictors, and outcome of thrombosis after successful implantation of drug-eluting stents JAMA 2005 ;  293 : 2126-2130 [cross-ref]
Wenaweser P., Daemen J., Zwahlen M., and al. Incidence and correlates of drug-eluting stent thrombosis in routine clinical practice. 4-year results from a large 2-institutional cohort study J Am Coll Cardiol 2008 ;  52 : 1134-1140 [cross-ref]
Holmes D.R., Kereiakes D.J., Laskey W.K., and al. Thrombosis and drug-eluting stents: an objective appraisal J Am Coll Cardiol 2007 ;  50 : 109-118 [cross-ref]
Awata M., Kotani J., Uematsu M., and al. Serial angioscopic evidence of incomplete neointimal coverage after sirolimus-eluting stent implantation: comparison with bare-metal stents Circulation 2007 ;  116 : 910-916 [cross-ref]
Finn A.V., Nakazawa G., Joner M., and al. Vascular responses to drug-eluting stents: importance of delayed healing Arterioscler Thromb Vasc Biol 2007 ;  27 : 1500-1510 [cross-ref]
Finn A.V., Joner M., Nakazawa G., and al. Pathological correlates of late drug-eluting stent thrombosis: strut coverage as a marker of endothelialization Circulation 2007 ;  115 : 2435-2441 [cross-ref]
Joner M., Finn A.V., Farb A., and al. Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk J Am Coll Cardiol 2006 ;  48 : 193-202 [cross-ref]
Chen B.X., Ma F.Y., Luo W., and al. Neointimal coverage of bare-metal and sirolimus-eluting stents evaluated with optical coherence tomography Heart 2008 ;  94 : 566-570 [cross-ref]
Matsumoto D., Shite J., Shinke T., and al. Neointimal coverage of sirolimus-eluting stents at 6-month follow-up: evaluated by optical coherence tomography Eur Heart J 2007 ;  28 : 961-967 [cross-ref]
Takano M., Inami S., Jang I.K., and al. Evaluation by optical coherence tomography of neointimal coverage of sirolimus-eluting stent three months after implantation Am J Cardiol 2007 ;  99 : 1033-1038 [inter-ref]
Takano M., Yamamoto M., Inami S., and al. Long-term follow-up evaluation after sirolimus-eluting stent implantation by optical coherence tomography: do uncovered struts persist? J Am Coll Cardiol 2008 ;  51 : 968-969 [cross-ref]
Pocock S.J., Lansky A.J., Mehran R., and al. Angiographic surrogate end points in drug-eluting stent trials: a systematic evaluation based on individual patient data from 11 randomized, controlled trials J Am Coll Cardiol 2008 ;  51 : 23-32 [cross-ref]
Kim J.S., Jang I.K., Kim T.H., and al. Optical coherence tomography evaluation of zotarolimus-eluting stents at 9-month follow-up: comparison with sirolimus-eluting stents Heart 2009 ;  95 : 1907-1912 [cross-ref]
Motreff P., Souteyrand G., Levesque S., and al. Comparative analysis of neointimal coverage with paclitaxel and zotarolimus drug-eluting stents, using optical coherence tomography 6 months after implantation Arch Cardiovasc Dis 2009 ;  102 : 617-624 [inter-ref]
Guagliumi G, Sirbu V, Suzuki N, et al. Prospective, randomized, controlled study using optical coherence tomography to evaluate strut coverage of Sirolimus-, Paclitaxel- and Zotarolimus-Eluting coronary stents in long lesions requiring overlapping (ODESSA-OCT for DES safety). Am J Cardiol 200;102. (abstract).
Kornovski R. The long-term results of the “Endeavor” stent Catheter Cardiovasc Interv 2009 ;  74 : 996-999 [cross-ref]
Takano M., Yamamoto M., Xie Y., and al. Serial long-term evaluation of neointimal stent coverage and thrombus after sirolimus-eluting stent implantation by use of coronary angioscopy Heart 2007 ;  93 : 1533-1536
Sano K., Mintz G.S., Carlier S.G., and al. Volumetric intravascular ultrasound assessment of neointimal hyperplasia and non-uniform stent strut distribution in sirolimus-eluting stent restenosis Am J Cardiol 2006 ;  98 : 1559-1562 [inter-ref]
Takebayashi H., Mintz G.S., Carlier S.G., and al. Non-uniform strut distribution correlates with more neointimal hyperplasia after sirolimus-eluting stent implantation Circulation 2004 ;  110 : 3430-3434 [cross-ref]
Kaneda H., Ako J., Terashima M., and al. Distribution pattern of neointimal hyperplasia following sirolimus-eluting stent implantation assessed by 3-dimensional intravascular ultrasound Int J Cardiol 2009 ;  135 : 243-245 [cross-ref]



© 2011  Elsevier Masson SAS. All Rights Reserved.
EM-CONSULTE.COM is registrered at the CNIL, déclaration n° 1286925.
As per the Law relating to information storage and personal integrity, you have the right to oppose (art 26 of that law), access (art 34 of that law) and rectify (art 36 of that law) your personal data. You may thus request that your data, should it be inaccurate, incomplete, unclear, outdated, not be used or stored, be corrected, clarified, updated or deleted.
Personal information regarding our website's visitors, including their identity, is confidential.
The owners of this website hereby guarantee to respect the legal confidentiality conditions, applicable in France, and not to disclose this data to third parties.
Close
Article Outline