Patient selection for transcatheter aortic valve implantation: Patient risk profile and anatomical selection criteria
Sélection des patients avant implantation de valves aortiques percutanées : critères cliniques et anatomiques
 

Nicolo Piazza a, , b , Ruediger Lange b, Giuseppe Martucci a, Patrick W. Serruys c
a Department of Interventional Cardiology, McGill University Health Center, Royal Victoria Hospital, 687, Pine Avenue West, Montreal, Quebec, H3A 1A1, Canada 
b Department of Cardiac Surgery, German Heart Centre, Munich, Germany 
c Department of Interventional Cardiology, Erasmus Medical Center, Rotterdam, The Netherlands 

Corresponding author.
Summary

Patient selection plays a crucial role in the success of transcatheter aortic valve implantation (TAVI). It requires meticulous attention to the smallest of details and needs to be performed in a systematic manner for every patient. In essence, the patient must be assessed from access to implantation site. Becoming over “complacent” and “routine” may lead to failure and impact patient safety. TAVI is indicated for high or prohibitive surgical risk patients with severe aortic stenosis. Some patients, however, are too high risk even for TAVI. In addition to patient risk evaluation, anatomical selection criteria need to be considered. Multimodality imaging, using a combination of angiography, echocardiography and multislice computed tomography, is necessary to determine the anatomical suitability for the procedure.

The full text of this article is available in PDF format.
Résumé

La sélection des patients joue un rôle majeur dans le succès des implantations de valve aortique par voie percutanée. Elle nécessite de porter une attention méticuleuse au moindre détail et doit être réalisée de façon systématique chez chaque patient. Le patient doit être évalué du site d’abord à la valve aortique. Un bilan « complacent » ? (je ne sait pas le traduire) ou trop « routinier » peut conduire à l’échec du geste et entraîner des complications. Les valves aortiques percutanées sont indiquées chez les patients porteurs d’un rétrécissement aortique serré et considérés comme étant à risque chirurgical trop élevé, voire même contre-indiqués. Certains patients peuvent également être à risque trop élevé pour une valve percutanée. À l’évaluation du risque opératoire doit s’ajouter l’évaluation anatomique du patient. Diverses modalités d’imagerie telles que l’angiographie, l’échographie et le scanner, sont nécessaires pour évaluer la faisabilité technique du geste.

The full text of this article is available in PDF format.

Keywords : Aortic stenosis, Patient selection, Transcatheter aortic valve implantation

Mots clés : Sténose aortique, Transcatheter, Sélection

Abbreviations : DEM, MPR, MSCT, SFAR, STS, TAVI, TAV-in-SAV, TEE, TTE, VARC


Background

Patient selection plays a crucial role in the success of transcatheter aortic valve implantation (TAVI). It requires meticulous attention to the smallest of details and needs to be performed in a systematic manner for every patient. In essence, the patient must be assessed from access to implantation site. Becoming over “complacent” and “routine” may lead to failure and impact patient safety.

In Europe, TAVI is indicated for high or prohibitive surgical risk patients with severe aortic stenosis. Some patients, however, are too high risk even for TAVI. In addition to patient risk evaluation, anatomical selection criteria need to be considered. Multimodality imaging, using a combination of angiography, echocardiography and multislice computed tomography (MSCT), is necessary to determine the anatomical suitability for the procedure. In particular, assessment of the peripheral vasculature and aortic valvar complex will allow selection of the access route and prosthesis type and size, respectively.

Patient risk evaluation

A number of cardiac surgical risk algorithms have been developed over the last 20 years. Despite their known limitations, the logistic EuroSCORE I and the STS (Society of Thoracic Surgeons) Predicted Risk of Mortality score have guided enrolment of “high surgical risk” patients into TAVI trials. Because low-to-intermediate surgical risk patients characterized the development of these risk models, their reliability when applied to high or prohibitive surgical risk patients has been rightfully questioned. Furthermore, models may not take into account important comorbidities (e.g. porcelain aorta, chest wall radiation, liver cirrhosis, pulmonary hypertension) and frailty variables that may impact clinical outcomes. The newly updated logistic EuroSCORE II and STS score, however, are expected to incorporate frailty variables.

In high surgical risk patients, the logistic EuroSCORE I tends to overestimate the observed mortality risk by a factor of 2 to 3; the new logistic EuroSCORE II appears to provide significantly lower mortality estimates than the logistic EuroSCORE I. The STS score has been found to be more reliable than the logistic EuroSCORE I or the Ambler Risk Score for the prediction of operative and long-term mortality in high-risk patients undergoing surgical aortic valve replacement.

Clinical judgment should supersede any surgical risk algorithm. Risk scores should guide but not dictate clinical decision-making.

A need for a transcatheter aortic valve implantation (TAVI) risk score?

A TAVI risk score for TAVI patients, akin to the logistic EuroSCORE or STS score for surgical patients, is currently not available. A TAVI risk score may provide additional information during the preprocedural screening process to better understand the potential outcomes of patients after TAVI.

Recent publications have identified a number of baseline variables to be independently associated with mortality or poor treatment response in patients undergoing TAVI: low body mass, functional status, left ventricular dysfunction, low gradient aortic stenosis, concentration of N-terminal prohormone of brain natriuretic peptide, diabetes, prior stroke, chronic obstructive pulmonary disease, chronic kidney disease, severe tricuspid and mitral regurgitation and baseline anaemia [1–9, 3, 4, 5, 6, 7, 8, 9, 2]. In addition to baseline demographics, anatomical characteristics may also help to predict treatment response after TAVI.

Anatomical patient selection criteria

The most important aspect of anatomical screening involves assessment of the arterial vasculature and aortic valvar complex (left ventricular outflow tract, aortic annulus, sinus of Valsalva, sinutubular junction and ascending aorta) (Figure 1). This information will guide physicians to select the most appropriate access route (i.e. transfemoral, subclavian, apical or direct aortic) and transcatheter valve size. Furthermore, it will alert physicians to potential complications that may arise during the procedure. Information about specific anatomical criteria is provided in the paragraphs that follow.



Figure 1


Figure 1. 

After selecting the appropriate high or prohibitive surgical risk patient, it becomes critical to assess the anatomical eligibility for transcatheter aortic valve implantation (TAVI). This figure demonstrates that every aspect of the vasculature from “skin to heart” needs to be examined with respect to minimal vessel diameter, degree of atherosclerosis, tortuosity and calcification. Furthermore, alternative access sites (e.g. subclavian, transapical direct aortic) should be assessed and kept in mind in case of bail out options.

Zoom

Assessment of the arterial vasculature

The occurrence of vascular complications associated with TAVI is not surprising given the use of large bore catheters and imperfect vascular closure devices. Vascular complications have been reported in 2% to 30% of patients undergoing TAVI; the conflicting reports are due to heterogeneous end-point definitions [10, 11, 12, 13].

Peripheral contrast angiography is practical and associated with relatively lower costs, lower contrast dye injection and lower radiation exposure compared with MSCT [14]. MSCT, however, provides greater appreciation of vessel size, tortuosity and calcific burden [15].

A recent angiographical study identified an outer sheath diameter to femoral artery minimal luminal diameter (SFAR) ratio greater or equal to 1.05 as a predictor of Valve Academic Research Consortium (VARC) major vascular complications as well as 30-day mortality [10]. The SFAR cut-off ratio increased to 1.10 in the absence of calcium and decreased to 1.00 in the presence of calcium. These results need to be confirmed in prospective studies and should also include MSCT.

The femoral artery is typically used as the default vascular access. Peripheral vascular disease is not an absolute contraindication to TAVI; it does, however, increase the risk of complications significantly. Physicians must be skilled or have the necessary resources to treat vascular injuries (percutaneously or surgically). Some physicians attempt to cautiously advance the vascular access sheath and catheter delivery system across the diseased vasculature, implant the valve and repair any complications such as dissections “on the way out” by percutaneous transluminal angioplasty/stent implantation. Alternatively, peripheral vascular interventions (percutaneous transluminal angioplasty or stent implantation) can be performed prior to valve implantation. While this latter approach has been performed successfully in experienced centres, the risk of dislodging the implanted stent during advancement of the vascular access sheath or delivery catheter needs to be considered.

Significant tortuosity alone of the iliofemoral vessels is not necessarily a contraindication to TAVI as long as the vessels are otherwise healthy and compliant — gentle advancement of the stiff guidewire or vascular access sheath will tend to straighten the vessel.

The Edwards eSheath features the innovative “dynamic expansion mechanism” (DEM) that allows for transient sheath expansion during valve delivery. Immediately after the Edwards SAPIEN XT prosthesis passes through the sheath, the DEM allows the sheath to return to a low profile diameter. This reduces the time the access vessel is expanded, thereby minimizing the risk of vascular trauma. The unexpanded eSheath has an inner diameter of 5.3mm (16F) and an outer diameter of 6.7mm (20F) for implantation of the 23mm prosthesis; an inner diameter of 5.9mm (18F) and an outer diameter of 7.2mm (21–22F) for implantation of the 26mm prosthesis; and an inner diameter of 6.7mm (20F) and an outer diameter of 8mm (24F) for implantation of the 29mm prosthesis (Table 1). Acceptable minimal iliofemoral diameters for implantation of the 23, 26 and 29mm Edwards SAPIEN XT prostheses are 6.0, 6.5 and 7.0mm, respectively.

In the current version of the Medtronic CoreValve system, a custom introducer sheath is not supplied. Current recommendations are to use a Cook 30cm Check-Flo Performer 18F introducer (Cook, Bloomington, IN, USA). The St. Jude 30cm Ultimum EV 18F introducer (St. Jude, Minnetonka, MN, USA) is less utilized because it has less column strength and is therefore less kink resistant than the Cook sheath. The acceptable minimal iliofemoral diameter for implantation of the 26, 29 and 31mm Medtronic CoreValve prostheses is 6.0mm.

Alternative routes to the transfemoral include the subclavian artery [16, 17], apex of the heart [18, 19] and ascending aorta [20, 21]. In selected cases, non-transfemoral routes may be safer and just as effective. Transapical approaches with the Edwards SAPIEN device, JenaValve and Symetis Acurate system, and the subclavian and direct aortic approaches with the Medtronic CoreValve System are Conformité Européenne (CE) mark approved.

The left subclavian artery (as opposed to the right) is selected in more than 95% of Medtronic CoreValve subclavian cases [16]. In most cases, the approach from the left subclavian is straightforward irrespective of the aortic root angulation. In contrast, if the angulation between the plane of the annulus and a horizontal reference line is more than 30 degrees (suggesting a horizontal aorta or vertical annulus plane), an approach from the right subclavian will be technically challenging [22]. The presence of an internal mammary coronary artery bypass graft is a relative contraindication to the subclavian approach. Having said that, successful cases have been performed in its presence with a minimum vessel diameter requirement of 6.5 to 7.0mm [23]. In order to reduce the risk of retrograde dissection of the internal mammary artery graft, the segment of the subclavian artery proximal to the internal mammary should be free of atherosclerotic disease.

Measurement of the aortic valve annulus

Surgeons commonly define the aortic valve annulus as the semilunar crown-like ring demarcated by the “draping” leaflet attachment line that runs across the aortic root. For the purposes of TAVI, the enigmatic “aortic valve annulus” corresponds to a virtual ring formed by joining the basal attachment points of the leaflets within the left ventricle [24]. This plane represents the inlet from the left ventricular outflow tract into the aortic root. The fact that the aortic valve annulus has a non-circular shape has led to a great deal of debate over the optimal imaging modality for measuring its diameter [25, 26] (Figure 2).



Figure 2


Figure 2. 

Multislice computed tomography axial cuts of the aortic annulus from 12 patients, demonstrating that the aortic annulus is, in fact, non-circular. The difference between the maximum and minimum diameter measurements of the aortic annulus is on average 6.5mm with a standard deviation of approximately 2mm. The non-circularity of the aortic annulus limits applicability of two-dimensional imaging in estimating the annulus diameter for transcatheter valve sizing.

Zoom

The aortic valve annulus diameter can be measured using various imaging modalities. These include transthoracic echocardiography (TTE), transesophageal echocardiography (TEE), MSCT, contrast aortography and magnetic resonance imaging (Figure 3).



Figure 3


Figure 3. 

Various imaging modalities in clinical use for assessing the diameter of the aortic annulus. A. Echocardiography. B. Fluoroscopy. C. Multislice computed tomography. D. Magnetic resonance imaging.

Zoom

Transcatheter aortic bioprostheses are typically oversized by 5 to 30% relative to the aortic valve annulus diameter. The intention of oversizing is to create enough interference (or radial force) between the prosthetic valve and aortic valvar complex to ensure adequate anchoring and sealing.

According to MSCT data, the aortic valve annulus is non-circular in the vast majority of patients [25]. The mean difference between the maximum and minimum diameter of the aortic annulus is 6.5mm (95% confidence interval, 5.7–7.2) [26]. This fact explains the potential shortcomings of relying on two-dimensional measurements of the aortic annulus for transcatheter aortic valve sizing (Figure 4).



Figure 4


Figure 4. 

Short-axis superior view of the human aortic valve seen from the ascending aorta. Depending on the orientation of the chord transecting the aortic annulus, and whether or not the chord transects the centre of the aortic valve, different measurements of the aortic annulus can be made by two-dimensional imaging.

Zoom

MSCT multiplanar reconstructions (MPRs) can provide coronal, sagittal and axial images of the aortic root. Methodologies for MPRs of the aortic valve annulus have been previously described [27]. In the axial view, the aortic valve annulus lies at a level just below the basal attachment points of the three leaflets (Figure 5). From this view, the maximum and minimum diameter, perimeter and area of the annulus can be measured (Figure 5, Figure 6). The maximum and minimum diameters typically correspond to coronal and sagittal planes cut across the annulus, respectively. Depending on the orientation of the chord cut across the annulus, two-dimensional echocardiography provides a “one-dimensional view” of the aortic annular dimensions; typically underestimating the annulus diameter with respect to MSCT. Despite this knowledge, we are still left with the conundrum of how to appropriately apply MSCT aortic annular measurements to existing echocardiographic sizing criteria.



Figure 5


Figure 5. 

Serial axial cuts across the aortic valvar complex. A. Basal attachments of the aortic valvar leaflets. B and C. Slice (approximately 0.85mm) below the basal attachments of the aortic valvar leaflets and the level at which the aortic annulus is measured. In (B), the maximum and minimum diameters are obtained whereas in (C), a tracing of the “aortic valve annulus” provides the perimeter and cross-sectional area measurements.

Zoom



Figure 6


Figure 6. 

The Edwards SAPIEN XT valve is currently available in four sizes (20, 23, 26 and 29mm) and can treat aortic annuli diameters ranging from 16 to 27mm.

Zoom

Aortography during preimplantation balloon aortic valvuloplasty (i.e. balloon sizing) is a simple way to confirm the optimal valve size depending on the presence or absence of aortic regurgitation [28].

Having said all of this, two-dimensional echocardiography (TTE or TEE) remains the most commonly used and practical method to assess the aortic valve annulus diameter. Nonetheless, as we acquire more data in the near future, MSCT will likely become the primary imaging modality to measure the dimensions of the aortic valvar complex and guide transcatheter aortic valve size selection.

The Edwards SAPIEN XT THV is currently available in four sizes (20, 23, 26 and 29mm) and can be implanted in native annuli with diameters of 16 to 27mm (Figure 6). The CoreValve is currently available in three sizes (26mm, 29mm and 31mm) and can be implanted in native annuli with diameters ranging from 20 to 29mm (Figure 7).



Figure 7


Figure 7. 

The Medtronic CoreValve is currently available in three sizes (26, 29 and 31mm) and can treat aortic annuli diameters ranging from 20 to 29mm.

Zoom

It’s not just about the aortic valve annular measurements — the aortic valvar complex

In addition to the aortic annulus diameter, other measurements of the aortic valvar complex, such as the height and width of the Sinus of Valsalva, take-off heights of the coronary arteries, left ventricular outflow tract diameter, ascending aorta diameter and calcification burden, may influence the feasibility, safety and effectiveness of the procedure (Table 2).

Coronary artery disease

According to published case series, up to three-quarters of patients undergoing TAVI may have documented coronary artery disease. Expectedly, patients with coronary artery disease have higher surgical risk scores and associated comorbidities than patients without coronary artery disease [29]. Percutaneous coronary intervention in association with TAVI has been reported in 0 to 44% of patients. Although concomitant and staged strategies have been reported successfully, the latter approach appears to be more commonly used [30, 31, 32, 33, 34]. Discrepancies exist amongst studies as to whether coronary artery disease negatively impacts survival post-TAVI [30, 35, 36, 37]. A recent study showed no differences in VARC clinical outcomes between 70 patients who underwent isolated TAVI versus 55 who underwent concomitant TAVI and percutaneous coronary intervention [38].

Optimal management of patients with concomitant coronary artery disease undergoing TAVI is currently a subject of debate. It appears that a staged approach (percutaneous coronary intervention followed by TAVI) is prudent in patients with lesions in a dominant proximal right coronary artery, left main, proximal left anterior descending and dominant proximal circumflex artery.

Transcatheter aortic valve implantation (TAVI) in special subgroups
Failing surgical bioprosthetic valves

The operative mortality for elective re-do aortic valve surgery is reported to range from 2 to 7%; but this percentage can increase to more than 30% in high-risk and non-elective patients [39, 40, 41]. Understanding the basic construction and dimensions, radiographic identification and potential failure modes of surgical aortic valve bioprostheses is fundamental for understanding the key principles involved in transcatheter aortic valve-in-surgical aortic valve (TAV-in-SAV) implantation [42]. One important fact is that the labelling size of surgical bioprostheses does not correspond to their internal stent diameter; the internal stent diameter determines the transcatheter aortic valve size. Reference tables relating labelling sizes and internal stent diameters have already been published [42]. Over 100 successful TAV-in-SAV implantations have been reported with the Medtronic CoreValve and Edwards SAPIEN transcatheter heart valve for failing stented and stentless surgical bioprostheses [43]. Transaortic valvular gradients after TAV-in-SAV implantation are around 20mmHg (higher than TAVI for native aortic valve stenosis or surgical aortic valve replacement) and paravalvular leaks are typically trivial to mild.

Bicuspid valves

Although congenital or acquired bicuspid aortic valve stenosis is considered a contraindication to TAVI, several successful case reports have been documented [44, 45, 46, 47, 48, 49, 50, 51]. Anecdotally, stenotic bicuspid aortic annuli are usually larger and more eccentric than stenotic tricuspid aortic valves. As such, three-dimensional imaging is strongly advised for transcatheter aortic valve sizing. Furthermore, assessment of the ascending aorta and its treatment require special attention.

Lower surgical risk patients

TAVI was initially conceived for the treatment of high surgical risk or inoperable patients. A recent observational report noted a paradigm shift toward the selection of lower surgical risk patients for TAVI. Furthermore, significantly better clinical outcomes were observed in the lower (mean STS score 4%) than higher (mean STS score 7%) surgical risk patients undergoing TAVI, at 30-day and 6-month follow-up [52]. With increasing volumes and continued encouraging results, it appears that lower surgical risk patients are receiving TAVI. This is further corroborated by the ongoing preparation of the SURgical aortic valve replacement versus Transcatheter Aortic Valve Implantation (SURTAVI) trial and Placement of AoRTic traNscathetER (PARTNER II) valve trial, which are expected to randomize intermediate surgical risk patients with an STS score of 4 to 8% to TAVI or surgical aortic valve replacement.

Disclosure of interest

Nicolo Piazza is a consultant and proctor for Medtronic CoreValve and Ruediger Lange is a consultant for Medtronic CoreValve.

References

Buellesfeld L., Wenaweser P., Gerckens U., and al. Transcatheter aortic valve implantation: predictors of procedural success – the Siegburg-Bern experience Eur Heart J 2010 ;  31 : 984-991 [cross-ref]
D’Onofrio A., Gasparetto V., Napodano M., and al. Impact of preoperative mitral valve regurgitation on outcomes after transcatheter aortic valve implantation Eur J Cardiothorac Surg 2011 Dec 29 ; [Epub ahead of print, PMID: 22219481, PubMed - as supplied by publisher].
Gotzmann M., Pljakic A., Bojara W., and al. Transcatheter aortic valve implantation in patients with severe symptomatic aortic valve stenosis-predictors of mortality and poor treatment response Am Heart J 2011 ;  162 : 238-2450 [inter-ref]
Moat N.E., Ludman P., de Belder M.A., and al. Long-term outcomes after transcatheter aortic valve implantation in high-risk patients with severe aortic stenosis: the UK TAVI (United Kingdom Transcatheter Aortic Valve Implantation) Registry J Am Coll Cardiol 2011 ;  58 : 2130-2138 [cross-ref]
Sinning J.M., Ghanem A., Steinhauser H., and al. Renal function as predictor of mortality in patients after percutaneous transcatheter aortic valve implantation JACC Cardiovasc Interv 2010 ;  3 : 1141-1149 [cross-ref]
Spargias K., Polymeros S., Dimopoulos A., and al. The predictive value and evolution of N-terminal pro-B-type natriuretic peptide levels following transcutaneous aortic valve implantation J Interv Cardiol 2011 ;  24 : 462-469 [cross-ref]
Tamburino C., Capodanno D., Ramondo A., and al. Incidence and predictors of early and late mortality after transcatheter aortic valve implantation in 663 patients with severe aortic stenosis Circulation 2011 ;  123 : 299-308 [cross-ref]
Van Mieghem N.M., Nuis R.J., Tzikas A., and al. Prevalence and prognostic implications of baseline anaemia in patients undergoing transcatheter aortic valve implantation EuroIntervention 2011 ;  7 : 184-191 [cross-ref]
Wenaweser P., Pilgrim T., Roth N., and al. Clinical outcome and predictors for adverse events after transcatheter aortic valve implantation with the use of different devices and access routes Am Heart J 2011 ;  161 : 1114-1124 [inter-ref]
Hayashida K., Lefevre T., Chevalier B., and al. Transfemoral aortic valve implantation new criteria to predict vascular complications JACC Cardiovasc Interv 2011 ;  4 : 851-858 [cross-ref]
Kahlert P., Al-Rashid F., Weber M., and al. Vascular access site complications after percutaneous transfemoral aortic valve implantation Herz 2009 ;  34 : 398-408 [cross-ref]
Tchetche D., Dumonteil N., Sauguet A., and al. Thirty-day outcome and vascular complications after transarterial aortic valve implantation using both Edwards Sapien and Medtronic CoreValve bioprostheses in a mixed population EuroIntervention 2010 ;  5 : 659-665 [cross-ref]
Van Mieghem N.M., Nuis R.J., Piazza N., and al. Vascular complications with transcatheter aortic valve implantation using the 18 Fr Medtronic CoreValve System: the Rotterdam experience EuroIntervention 2010 ;  5 : 673-679 [cross-ref]
Eltchaninoff H., Kerkeni M., Zajarias A., and al. Aorto-iliac angiography as a screening tool in selecting patients for transfemoral aortic valve implantation with the Edwards SAPIEN bioprosthesis EuroIntervention 2009 ;  5 : 438-442 [cross-ref]
Delgado V., Ewe S.H., Ng A.C., and al. Multimodality imaging in transcatheter aortic valve implantation: key steps to assess procedural feasibility EuroIntervention 2010 ;  6 : 643-652 [cross-ref]
Petronio A.S., De Carlo M., Bedogni F., and al. Safety and efficacy of the subclavian approach for transcatheter aortic valve implantation with the CoreValve revalving system Circ Cardiovasc Interv 2010 ;  3 : 359-366 [cross-ref]
Sharp A.S., Michev I., Colombo A. First trans-axillary implantation of Edwards Sapien valve to treat an incompetent aortic bioprosthesis Catheter Cardiovasc Interv 2010 ;  75 : 507-510
D’Onofrio A., Rubino P., Fusari M., and al. Clinical and hemodynamic outcomes of “all-comers” undergoing transapical aortic valve implantation: results from the Italian Registry of Trans-Apical Aortic Valve Implantation (I-TA) J Thorac Cardiovasc Surg 2011 ;  142 : 768-775
Kempfert J., Rastan A., Holzhey D., and al. Transapical aortic valve implantation: analysis of risk factors and learning experience in 299 patients Circulation 2011 ;  124 : S124-S129
Bruschi G., De Marco F., Fratto P., and al. Direct aortic access through right minithoracotomy for implantation of self-expanding aortic bioprosthetic valves J Thorac Cardiovasc Surg 2010 ;  140 : 715-717 [cross-ref]
Etienne P.Y., Papadatos S., El Khoury E., and al. Transaortic transcatheter aortic valve implantation with the Edwards SAPIEN valve: feasibility, technical considerations, and clinical advantages Ann Thorac Surg 2011 ;  92 : 746-748 [cross-ref]
Piazza N., de Jaegere P., Manoharan G., and al. Patient selection for the CoreValve ReValving System Transcatheter aortic valve implantation: tips and tricks to avoid failure New York: Informa Healthcare (2011).  82
De Marco F., Bruschi G., Klugmann S. Transcatheter aortic valve implantation by left subclavian access in the presence of a patent LIMA to LAD graft Catheter Cardiovasc Interv 2011 ;  77 : 430-434 [cross-ref]
Piazza N., de Jaegere P., Schultz C., and al. Anatomy of the aortic valvar complex and its implications for transcatheter implantation of the aortic valve Circ Cardiovasc Interv 2008 ;  1 : 74-81 [cross-ref]
Hamdan A., Guetta V., Konen E., and al. Deformation dynamics and mechanical properties of the aortic annulus by 4-dimensional computed tomography: insights into the functional anatomy of the aortic valve complex and implications for transcatheter aortic valve therapy J Am Coll Cardiol 2012 ;  59 : 119-127 [cross-ref]
Schultz C.J., Moelker A., Piazza N., and al. Three-dimensional evaluation of the aortic annulus using multislice computer tomography: are manufacturer’s guidelines for sizing for percutaneous aortic valve replacement helpful? Eur Heart J 2010 ;  31 : 849-856
Schultz C., Moelker A., Tzikas A., and al. The use of MSCT for the evaluation of the aortic root before transcutaneous aortic valve implantation: the Rotterdam approach EuroIntervention 2010 ;  6 : 505-511 [cross-ref]
Babaliaros V.C., Junagadhwalla Z., Lerakis S., and al. Use of balloon aortic valvuloplasty to size the aortic annulus before implantation of a balloon-expandable transcatheter heart valve JACC Cardiovasc Interv 2010 ;  3 : 114-118 [cross-ref]
Wenaweser P., Pilgrim T., Guerios E., and al. Impact of coronary artery disease and percutaneous coronary intervention on outcomes in patients with severe aortic stenosis undergoing transcatheter aortic valve implantation EuroIntervention 2011 ;  7 : 541-548 [cross-ref]
Conradi L., Seiffert M., Franzen O., and al. First experience with transcatheter aortic valve implantation and concomitant percutaneous coronary intervention Clin Res Cardiol 2011 ;  100 : 311-316 [cross-ref]
Fontes-Carvalho R., Caeiro D., Santos L., and al. Combined percutaneous treatment of severe aortic stenosis and left main disease in a high surgical risk patient Rev Port Cardiol 2011 ;  30 : 229-234
Gerckens U., Latsios G., Mueller R., and al. Left main PCI after trans-subclavian CoreValve implantation. Successful outcome of a combined procedure for management of a rare complication Clin Res Cardiol 2009 ;  98 : 687-690 [cross-ref]
Mostafa A.E., Geist V., Abdel-Wahab M. Ad-hoc percutaneous coronary intervention and transcatheter aortic valve implantation as a combined transfemoral procedure J Invasive Cardiol 2011 ;  23 : E102-E105
Piazza N., Serruys P.W., de Jaegere P. Feasibility of complex coronary intervention in combination with percutaneous aortic valve implantation in patients with aortic stenosis using percutaneous left ventricular assist device (TandemHeart) Catheter Cardiovasc Interv 2009 ;  73 : 161-166 [cross-ref]
Dewey T.M., Brown D.L., Herbert M.A., and al. Effect of concomitant coronary artery disease on procedural and late outcomes of transcatheter aortic valve implantation Ann Thorac Surg 2010 ;  89 : 758-767[discussion 67].  [cross-ref]
Gautier M., Pepin M., Himbert D., and al. Impact of coronary artery disease on indications for transcatheter aortic valve implantation and on procedural outcomes EuroIntervention 2011 ;  7 : 549-555 [cross-ref]
Masson J.B., Lee M., Boone R.H., and al. Impact of coronary artery disease on outcomes after transcatheter aortic valve implantation Catheter Cardiovasc Interv 2010 ;  76 : 165-173 [cross-ref]
Abdel-Wahab M., Mostafa A.E., Geist V., and al. Comparison of outcomes in patients having isolated transcatheter aortic valve implantation versus combined with preprocedural percutaneous coronary intervention Am J Cardiol 2012 ;  109 : 581-586 [inter-ref]
Christiansen S., Schmid M., Autschbach R. Perioperative risk of redo aortic valve replacement Ann Thorac Cardiovasc Surg 2009 ;  15 : 105-110
Jamieson W.R., Burr L.H., Miyagishima R.T., and al. Reoperation for bioprosthetic aortic structural failure – risk assessment Eur J Cardiothorac Surg 2003 ;  24 : 873-878 [cross-ref]
Vogt P.R., Brunner-LaRocca H., Sidler P., and al. Reoperative surgery for degenerated aortic bioprostheses: predictors for emergency surgery and reoperative mortality Eur J Cardiothorac Surg 2000 ;  17 : 134-139 [cross-ref]
Piazza N., Bleiziffer S., Brockmann G., and al. Transcatheter aortic valve implantation for failing surgical aortic bioprosthetic valve: from concept to clinical application and evaluation (part 1) JACC Cardiovasc Interv 2011 ;  4 : 721-732 [cross-ref]
Piazza N., Bleiziffer S., Brockmann G., and al. Transcatheter aortic valve implantation for failing surgical aortic bioprosthetic valve: from concept to clinical application and evaluation (part 2) JACC Cardiovasc Interv 2011 ;  4 : 733-742 [cross-ref]
Chiam P.T., Chao V.T., Tan S.Y., and al. Percutaneous transcatheter heart valve implantation in a bicuspid aortic valve JACC Cardiovasc Interv 2010 ;  3 : 559-561 [cross-ref]
Delgado V., Tops L.F., Schuijf J.D., and al. Successful deployment of a transcatheter aortic valve in bicuspid aortic stenosis: role of imaging with multislice computed tomography Circ Cardiovasc Imaging 2009 ;  2 : e12-e13
Ferrari E., Locca D., Sulzer C., and al. Successful transapical aortic valve implantation in a congenital bicuspid aortic valve Ann Thorac Surg 2010 ;  90 : 630-632 [cross-ref]
Jilaihawi H., Asgar A., Bonan R. Good outcome and valve function despite Medtronic corevalve underexpansion Catheter Cardiovasc Interv 2010 ;  76 : 1022-1025 [cross-ref]
Kochman J., Huczek Z., Koltowski L., and al. Transcatheter implantation of an aortic valve prosthesis in a female patient with severe bicuspid aortic stenosis Eur Heart J 2012 ;  33 : 112
Raja Y., Holloway B., Doshi S.N. Symmetrical expansion of an Edwards Sapien valve in a congenitally bicuspid aortic valve Heart 2011 ;  97 : 1113
Unsworth B., Malik I., Mikhail G.W. Recognising bicuspid aortic stenosis in patients referred for transcatheter aortic valve implantation: routine screening with three-dimensional transoesophageal echocardiography Heart 2010 ;  96 : 645
Wijesinghe N., Ye J., Rodes-Cabau J., and al. Transcatheter aortic valve implantation in patients with bicuspid aortic valve stenosis JACC Cardiovasc Interv 2010 ;  3 : 1122-1125 [cross-ref]
Lange R., Bleiziffer S., Mazzitelli D., and al. Improvements in transcatheter aortic valve implantation outcomes in lower surgical risk patients a glimpse into the future J Am Coll Cardiol 2012 ;  59 : 280-287 [cross-ref]



© 2012  Elsevier Masson SAS. All Rights Reserved.
EM-CONSULTE.COM is registrered at the CNIL, déclaration n° 1286925.
As per the Law relating to information storage and personal integrity, you have the right to oppose (art 26 of that law), access (art 34 of that law) and rectify (art 36 of that law) your personal data. You may thus request that your data, should it be inaccurate, incomplete, unclear, outdated, not be used or stored, be corrected, clarified, updated or deleted.
Personal information regarding our website's visitors, including their identity, is confidential.
The owners of this website hereby guarantee to respect the legal confidentiality conditions, applicable in France, and not to disclose this data to third parties.
Close
Article Outline