Validation des gazométries au cordon ombilical : étude au sein d’une maternité française
Systematic umbilical cord blood analysis at birth: Feasibility and reliability in a French labour ward
 

D. Ernst a, , J. Clerc a, E. Decullier b, c, d, G. Gavanier e, O. Dupuis a
a Service de gynécologie-obstétrique, centre hospitalo-universitaire de Lyon-Sud, hospices civils de Lyon, 165, chemin du Grand-Revoyet, 69310 Pierre-Bénite, France 
b Pôle information médicale évaluation recherche, centre hospitalo-universitaire de Lyon-Sud, hospices civils de Lyon, 69003 Lyon, France 
c EA 4129 santé individu société, RECIF, université de Lyon, 69002 Lyon, France 
d Université Lyon-1, 69008 Lyon, France 
e École polytechnique, département de mécanique, 91128 Palaiseau cedex, France 

Auteur correspondant.
Résumé
Objectif

En salle de naissance, l’évaluation de l’état de santé néonatale est primordiale. La réalisation et l’interprétation des gazométries au cordon de façon systématique permet une telle évaluation. Nous avons voulu établir la faisabilité et la fiabilité des gazométries au sein d’une maternité de type 2B.

Patientes et méthodes

Nous avons étudié rétrospectivement les gazométries au cordon de 1000 accouchements consécutifs, de façon systématique. Premièrement, nous avons établi leur faisabilité, définie par le rapport de gazométries complètes sur le nombre d’accouchements d’enfants nés vivants. Deuxièmement, nous avons établi leur fiabilité, définie par le rapport des gazométries qui remplissaient les critères de qualité initialement définis par Westgate et al., et revus par Kro et al., sur le nombre de gazométries complètes d’enfants nés vivants. Enfin, nous avons cherché des facteurs influençant ces résultats.

Résultats

La faisabilité des prélèvements au cordon systématiquement réalisés était de 91,6 % et la fiabilité de 80,7 %. Concernant le mode d’accouchement, 38,6 % des césariennes en urgence (IC 95 % [30,8–46,3] ; p <0,0001) aboutissaient à des gazométries non valides, alors que seulement 11,3 % des césariennes prophylactiques (IC 95 % [4,3–18,2] ; p <0,0001) donnaient lieu à des gazométries non valides. Faisabilité et fiabilité étaient donc significativement plus faibles lors des césariennes en urgence.

Discussion et conclusion

La réalisation d’une gazométrie systématique s’accompagnait de 8,4 % de gazométries incomplètes, et 19,3 % de gazométries ininterprétables. Une formation rigoureuse serait donc souhaitable pour améliorer la faisabilité et la fiabilité, notamment lors des césariennes en urgence.

The full text of this article is available in PDF format.
Abstract
Objectives

At birth, evaluation of neonatal well-being is crucial. It is though important to perform umbilical cord blood gas analysis, and then to analyze the samples. We wanted to establish the feasibility and reliability of systematic umbilical cord blood sampling in a French labour ward.

Patients and methods

Study of systematic umbilical cord blood gas analysis was realized retrospectively from 1000 consecutive deliveries. We first established the feasibility of the samples. Feasibility was defined by the ratio of complete cord acid-base data on the number of deliveries from alive newborns. Afterwards, we established the reliability on the remaining cord samples. Reliability was the ratio of samples that fulfilled quality criteria defined by Westgate et al. and revised by Kro et al., on the number of complete samples from alive newborns. At last, we looked for factors that would influence these results.

Results

The systematic umbilical cord blood sample feasibility reached 91.6%, and the reliability reached 80.7%. About the delivery mode, 38.6% of emergency caesarians (IC 95% [30.8–46.3]; P <0.0001) led to non-valid samples, when only 11.3% of programmed caesarians (IC 95% [4.3–18.2]; P <0.0001) led to non-valid samples. Umbilical cord blood analysis were significantly less validated during emergency caesarians.

Discussion and conclusion

Realization of systematic cord blood gas analysis was followed by 8.4% of incomplete samples, and by 19.3% that were uninterpretable. Training sessions should be organized to improve the feasibility and reliability, especially during emergency caesarians.

The full text of this article is available in PDF format.

Mots clés : Sang cordon ombilical, Gazométries, Fiabilité, Faisabilité, Critères de qualité, pH au cordon, Validation

Keywords : Umbilical cord blood, Gas analysis, Feasibility, Reliability, Quality criteria, Cord pH, Validation




© 2012  Elsevier Masson SAS. All Rights Reserved.
EM-CONSULTE.COM is registrered at the CNIL, déclaration n° 1286925.
As per the Law relating to information storage and personal integrity, you have the right to oppose (art 26 of that law), access (art 34 of that law) and rectify (art 36 of that law) your personal data. You may thus request that your data, should it be inaccurate, incomplete, unclear, outdated, not be used or stored, be corrected, clarified, updated or deleted.
Personal information regarding our website's visitors, including their identity, is confidential.
The owners of this website hereby guarantee to respect the legal confidentiality conditions, applicable in France, and not to disclose this data to third parties.
Close
Article Outline