Archives of cardiovascular diseases
Volume 106, n° 5
pages 295-302 (mai 2013)
Doi : 10.1016/j.acvd.2012.11.005
Received : 3 July 2012 ;  accepted : 14 November 2012
Assessment of valvular surfaces in children with a congenital bicuspid aortic valve: Preliminary three-dimensional echocardiographic study
Estimation des surfaces valvulaires en cas de bicuspidie aortique chez l’enfant : une étude préliminaire d’échocardiographie tridimensionnelle
 

Thomas Cognet a, b, 1, Pierre-Emmanuel Séguéla a, , 1 , Emilie Thomson a, b, Frédéric Bouisset b, Olivier Lairez b, Sébastien Hascoët a, b, Didier Carrié b, Philippe Acar a
a Paediatric Cardiology Unit, Children’s Hospital, Toulouse University Hospital, Toulouse, France 
b Department of Cardiology, Toulouse University Hospital, Toulouse, France 

Corresponding author. Paediatric Cardiology Unit, Children’s Hospital, Toulouse University Hospital, 330, avenue de Grande-Bretagne, TSA 70034, 31059 Toulouse cedex 9, France. Fax: +33 5 34 55 86 63.
Summary
Background

Congenital bicuspid aortic valve (BAV) is the most common congenital heart defect and may be responsible for aortic stenosis early in life. However, its pathogenesis remains unclear. A relationship between the severity of aortic stenosis and valvular surfaces has not been reported in the paediatric population.

Aims

To assess the feasibility of three-dimensional transthoracic echocardiographic planimetry in congenital BAV in children and to evaluate the influence of valvular asymmetry and aortic valve area (AVA) on stenosis severity.

Methods

Seventy consecutive children with BAV were included in this prospective single-centre study. Using the multiplanar review mode, surfaces were measured by planimetry (in systole for AVA and diastole for cusp surfaces). The degree of stenosis was assessed by instantaneous aortic Doppler. Results are expressed as medians and first and third quartiles.

Results

Median age was 5.6 years (2.2–11.5). Feasibility was 97%. Intra- and interobserver concordances were excellent for the measurement of cusp surfaces and AVA. Among the 70 children, 25 had aortic stenosis. The small/large cusp ratio was strongly associated with aortic stenosis (P <0.001). The area under the receiver operating characteristic curve was 0.89 (95% confidence interval 0.82–0.97). The best cut-off value for differentiating stenotic from non-stenotic valve was 0.75, with 84% sensitivity and 83% specificity. When indexed for body surface area, AVA was significantly smaller (P =0.031) in case of stenotic BAV (1.51cm2 [0.99–2.28]) compared with non-stenotic BAV (1.99cm2 [1.57–2.52]).

Conclusions

Three-dimensional echocardiographic planimetry is a feasible and reproducible method for assessing aortic surfaces in congenital BAV. Aortic stenosis seems to strongly depend on the asymmetry of the valve.

The full text of this article is available in PDF format.
Résumé
Contexte

La bicuspidie aortique (BAo) est la malformation cardiaque congénitale la plus fréquente. Elle peut être responsable de sténose aortique (SAo) dès le plus jeune âge. Cependant, sa pathogénie exacte demeure incertaine. La relation entre sévérité de la SAo et importance des surfaces valvulaires n’a jamais été montrée chez l’enfant.

Objectifs

Étude de la faisabilité de l’échocardiographie transthoracique tridimensionnelle (E3D) pour l’étude planimétrique des BAo de l’enfant et étude de l’influence de l’asymétrie valvulaire ainsi que celle de l’orifice valvulaire aortique sur la sévérité de la SAo.

Méthodes

Soixante-dix enfants ayant une BAo ont été inclus consécutivement dans cette étude prospective monocentrique. Grâce au mode de revue multi-plan, les surfaces valvulaires ont été mesurées par planimétrie (en systole pour l’orifice aortique et en diastole pour les feuillets). Le degré de SAo était estimé par le mode Doppler. Les résultats sont présentés sous forme de médiane, premier et troisième quartiles.

Résultats

L’âge médian était de 5,6ans (2,2–11,5). La faisabilité était de 97 % avec une excellente variabilité intra- et interobservateur tant pour la mesure des feuillets valvulaires que pour celle de l’orifice aortique. 25 enfants avaient une SAo. Le ratio petit/grand feuillet était corrélé à l’existence d’une SAo (p <0,001), l’aire sous la courbe ROC étant de 0,89 (IC à 95 % : 0,82–0,97). La meilleure valeur seuil permettant de différencier une valve sténosante d’une non sténosante était de 0,75 (sensibilité 84 % et spécificité 83 %). Indexé à la surface corporelle, l’orifice valvulaire était significativement (p =0,031) plus petit en cas de BAo sténosante (1,51cm2 [0,99–2,28 vs 1,99cm2 [1,57–2,52]] pour les non sténosantes).

Conclusions

La mesure planimétrique des surfaces valvulaires par E3D est une technique faisable et reproductible en cas de BAo chez l’enfant. La SAo semble dépendre étroitement du caractère asymétrique de ces valves.

The full text of this article is available in PDF format.

Keywords : Bicuspid aortic valve, Transthoracic three-dimensional echocardiography, Aortic stenosis, Aortic valve area

Mots clés : Bicuspidie aortique, Échocardiographie transthoracique tridimensionnelle, Orifice valvulaire aortique, Sténose aortique

Abbreviations : 2DE, 2DTTE, 3DE, AVA, BAV, CI, MPR, MRI, RT-3DTTE


Background

Bicuspid aortic valve (BAV) is the most common congenital heart disease, observed in 0.4 to 2.25% of the general population [1, 2]. The clinical spectrum is wide, ranging from asymptomatic valvular disease to severe heart failure. The severity of the valve dysfunction is related to the morphology of the BAV [3]. Moreover, the phenotype of the BAV is predictive of surgical timing [4]. Thus, the challenge is to assess valvular function and determine its probable evolution. Three-dimensional echocardiography (3DE) was reported as an efficient means of assessing valvular morphology in BAV [5]. Indeed, using multiplanar analysis, this technique allows precise measurement of valvular areas [6]. In adults, real-time three-dimensional transthoracic echocardiography (RT-3DTTE) using the biplane mode was consistent with both transoesophageal 3DE and catheterization for measurement of the aortic valve area (AVA) [7]. Although previous studies have demonstrated that the severity of aortic stenosis closely depends on the structural geometry of the aortic valve, the exact pathogenesis remains unclear. The aims of this study were to evaluate the feasibility of RT-3DTTE for the assessment of aortic surfaces in congenital BAV and to assess the relationship between valvular asymmetry, AVA and aortic stenosis.

Methods
Definition of bicuspid aortic valve

The term BAV includes different morphological phenotypes. Indeed, the valve may be composed of either two cusps (purely bicuspid) or three cusps (falsely bicuspid) with fusion (partial or complete) of two of them. The term raphe defines the conjoint or ‘fused” area of the two underdeveloped cusps. For this study, we used the anatomical classification proposed by Sievers and Schmidtke [8], which is based on the presence and position of the raphes. Thus, a BAV type 0 (true BAV) has only two cusps with no raphe and BAV types 1 and 2 have three cusps with one and two raphes, respectively.

Patients

This was a single-centre prospective study performed in our paediatric cardiology unit. From April 2010 to April 2011, we enrolled all children with an echocardiographic diagnosis of BAV. These children were initially referred for the exploration of a cardiac murmur or for follow-up of an already known congenital heart disease. Exclusion criteria were age>18 years, history of endocarditis, concomitant subaortic obstruction, previous surgical aortic commissurotomy and previous percutaneous aortic dilation. Only BAV types 0 and 1 (with 0 raphes and one raphe, respectively) were included because BAV type 2 (three cusps and two raphes) may be considered as a functional unicuspid aortic valve. The study was approved by our local ethics committee. A written consent form was not required according to French law, given that echocardiographic evaluation was part of the regular management of the children and was required for management of their medical conditions. No additional examination was performed for the sole purpose of the study.

Echocardiographic acquisitions and off-line analysis

All patients underwent standardized and complete two-dimensional transthoracic echocardiography (2DTTE), performed using high-quality commercially available ultrasound systems (iE33, Philips Medical Systems, Andover, MA, USA). When a BAV was diagnosed, the examination was completed by RT-3DTTE. X3-1 and X7-2 matrix probes (Philips Medical Systems, Andover, MA, USA) were used, depending on the age of the patient. The best short-axis cross-sectional view was selected to see the most circular annulus and the commissural areas. RT-3DTTE data were acquired–as much as possible considering the extreme youth of some children – during end expiratory. When necessary, 3D zoom was used to improve image resolution. Q-Lab software (Philips Medical Systems, Andover, MA, USA) was used for off-line analysis. In the multiplanar review (MPR) mode, three orthogonal cutting planes can be moved independently of each other through the data set. The MPR mode allows better morphological analysis because the valve can be seen in the best 2D cutting plans at any stage of the cardiac cycle. 3D images were acquired by triggering to the electrocardiogram R-wave and quality was improved using gain and compression controls. Two 3D volumes were acquired for each patient. Cusp surfaces and AVA were determined by planimetry. The degree of asymmetry was assessed by manually surfacing the leaflets (Figure 1). Aortic cusp surfaces were measured in diastole. For BAV type 1, the sum of the two fused cusps was considered as a single functional cusp. The ratio between the surfaces of the two functional cusps was calculated to express the degree of asymmetry (small/large cusp ratio). AVA was measured at the inner leaflet edges at the time of maximal opening in systole. All areas were corrected for body surface area. Mean and maximal instantaneous aortic Doppler gradients were measured to evaluate aortic stenosis. Aortic Doppler values were obtained from an apical five-chamber view. The valve was defined as stenotic for a maximal instantaneous aortic Doppler gradient>20mmHg. The ascending aorta, measured in parasternal long-axis view, was defined as dilated if the Z-score of the diameter was>2.



Figure 1


Figure 1. 

Measurement of aortic cusp surfaces by planimetry using the multiplanar review mode. A. Schematic representation of the different types of bicuspid aortic valve (BAV) according to the classification described by Sievers and Schmidtke. BAV type 0 has two cusps with no raphe and BAV type 1 has three cusps with one raphe. Raphes are represented by black lines. In type 0, ap and lat refer to the spatial position of the cusps. In type 1, the position of the raphe is indicated by L-R (between the left and right coronary sinuses), R-N (between the right and non-coronary sinuses) and N-L (between the non-coronary and left coronary sinuses). The small/large cusp ratio corresponds to the surface of the white cusp/surface of the blue cusp ratio. Adapted from Sievers and Schmidtke [8]. B. Real-time transthoracic three-dimensional echocardiography of a pure BAV (type 0 ap). This valve is perfectly symmetrical as shown by measurement of the cusp surfaces. C. Multiplanar review mode from three-dimensional acquisition with the three orthogonal cutting plans. Cropping allows perfect visualization of the valve in a short-axis plane. Ap: anterior-posterior; lat: lateral; L: left coronary sinus; N: non-coronary sinus; R: right coronary sinus.

Zoom

Operators

RT-3DTTE acquisitions were performed by a single cardiologist. Two independent operators performed the off-line analysis, randomly and on separate days, to assess interobserver variability. To evaluate intraobserver variability, the same observer performed measurements twice, on different days, in a random order and blinded to the prior results.

Statistical analysis

Statistical analysis was performed on STATA statistical software, version 11.1 (Stata Corporation, College Station, TX, USA). Continuous variables are summarized as medians and interquartile ranges. Categorical variables are presented as proportions. In univariate analysis, qualitative variables were compared with the Chi2 test (or Fisher’s exact test; when necessary). The Mann–Whitney U test was used to compare ranges of continuous non-normally distributed variables according to qualitative variables. Reproducibility of the measurements was tested on 16 randomly selected patients by calculating intraclass correlation coefficients. Receiver operating characteristic curves were used to determine the best threshold values for sensitivity and specificity. Linear regression analyses expressing Doppler gradients according to AVA or small/large cusp ratio are presented. A P value<0.05 was considered statistically significant.

Results
Population characteristics

Among the 70 patients included in the study, 44 (62.9%) were male. Median age was 5.6 years (2.2–11.5) and median weight was 16.5kg (11–32) (Table 1). The most common associated cardiac lesion was coarctation of the aorta, encountered in 21 (30%) patients (Table 2). Aortic stenosis was noted in 25 patients (35.7%). The median value of the mean instantaneous aortic Doppler gradient was 16mmHg (10–30) for stenotic BAV and 2mmHg (2–3) for non-stenotic valves. The median value of the maximal instantaneous aortic Doppler gradient was 30mmHg (24–50) in case of aortic stenosis and 4mmHg (4–5) in case of non-stenotic BAV. Dilation of the ascending aorta occurred in 22 patients (31.4%) and was more frequent in case of aortic stenosis (48.0% vs. 22.2%; P =0.03). Four patients were excluded from the study because they had BAV type 2; three of them had aortic stenosis (one patient underwent a surgical commissurotomy). According to standard cross-sectional echocardiography, all patients had normal left ventricular function.

Feasibility

Cusp surface and AVA measurements were feasible for 97% of patients. RT-3DTTE analysis was not possible for only two patients because of poor image resolution attributed to poor echogenicity; these two patients were aged 14.9 and 15.7 years. Tachycardia was not a limiting factor in our population.

Reproducibility

Intra- and interobserver concordances were excellent for the measurement of the cusp areas and the orifice area. Indeed, intraclass correlation coefficients estimated for the assessment of intraobserver variability were 0.992 (95% confidence interval [CI] 0.979–0.997) for the small cusp surface, 0.998 (95% CI 0.993–0.999) for the large cusp surface and 0.998 (95% CI 0.995–0.999) for the AVA. Correlation coefficients expressing the interobserver variability were also excellent: 0.964 (95% CI 0.900–0.987) for the small cusp surface, 0.961 (95% CI 0.892–0.986) for the large cusp surface and 0.983 (95% CI 0.953–0.994) for the AVA.

Valvular asymmetry and aortic stenosis

According to the functional status of the valve (stenotic/non-stenotic), there was no difference between cusp areas for the large cusp (1.27cm2 vs. 1.31cm2; P =0.93) or for the small cusp (0.72cm2 vs. 1.13cm2; P =0.15). However, the small/large cusp ratio was strongly associated with aortic stenosis (P <0.001). Indeed, the median ratio was 0.59 [0.52–0.68] for aortic stenosis and 0.86 [0.79–0.91] for non-stenotic valves (Table 3). Showing the ability of this ratio to predict aortic stenosis, the area under the receiver operating characteristic curve (Figure 2) was excellent at 0.89 (95% CI 0.82–0.97) and the best cut-off value to distinguish stenotic valves from non-stenotic valves was 0.75, with 84% sensitivity and 83% specificity. When considering only valvular asymmetry, 20 of 25 patients (80%) with stenotic BAV had a small/large cusp ratio<0.75 (P <0.001). For a small/large cusp ratio<0.75, the odds ratio, calculated using a logistic regression, was 21.7 (95% CI 6.1–77.2; P <0.001). Finally, the degree of asymmetry correlated well with the maximal instantaneous aortic Doppler gradient (r =–0.6379, P <0.001) and the mean instantaneous aortic Doppler gradient (r =–0.6102, P <0.001) (Figure 3).



Figure 2


Figure 2. 

Receiver operating characteristic curve analysing the ability of valvular asymmetry to discriminate stenotic from non-stenotic bicuspid aortic valves. The valvular asymmetry is expressed by the small/large cusp ratio. Area under the curve is 0.89 with a 95% confidence interval of 0.82 to 0.97. The best cut-off value to distinguish the stenotic from the non-stenotic valve is 0.75, with 84% sensitivity and 83% specificity.

Zoom



Figure 3


Figure 3. 

Linear regression analysis showing the relationship between valvular asymmetry and aortic stenosis. A. The degree of stenosis is represented by the maximal instantaneous aortic Doppler gradient. Linear regression analysis shows a good relationship (r =–0.6379, P <0.001) with valvular asymmetry expressed by the small/large cusp ratio. B. The degree of aortic stenosis is represented by the mean instantaneous aortic Doppler gradient and linear regression analysis shows a good relationship (r =–0.6102, P <0.001) with valvular asymmetry. Red lines represent regression lines.

Zoom

Aortic valve area and aortic stenosis

Indexed AVA was significantly smaller in patients with a stenotic BAV than in patients with a non-stenotic BAV (1.51cm2 [0.99–2.28] vs. 1.99cm2 [1.57–2.52]; P =0.031). When not indexed, AVA was still significantly smaller in patients with a stenotic BAV than in patients with a non-stenotic BAV (1cm2 [0.57–1.82] vs. 1.82cm2 [0.82–2.66]; P =0.048). However, both AVA and indexed AVA were weakly correlated with the maximal instantaneous aortic Doppler gradient (respectively, r =–0.2772, P =0.011 and r =–0.2995, P =0.006) or with the mean instantaneous aortic Doppler gradient (respectively, r =–0.2697, P =0.012 and r =–0.2953, P =0.007).

Aortic valve area and valvular asymmetry

The association between indexed AVA and valvular asymmetry was not significant (r =0.0296, P =0.16). BAV with a small/large cusp ratio<0.75 had a non-significantly smaller indexed AVA than BAV with a small/large cusp ratio>0.75 (1.43cm2 [1.06–2.48] vs. 1.93cm2 [1.72–2.49]; P =0.09).

Asymmetry and aortic dilation

Children with aortic dilation (n =22) had a significantly lower small/large cusp ratio than children without aortic dilation (0.68 [0.50–0.84] vs. 0.81 [0.65–0.9]; P =0.027).

Discussion
Feasibility of aortic planimetry using real-time three-dimensional transthoracic echocardiography

This study shows that RT-3DTTE is a feasible method for analysing cusp surfaces and AVA in children with a BAV. Proving the good reproducibility of the method, both intra- and interobserver variability were low. Our data are consistent with those from previous adult studies. When compared with catheterization and two-dimensional echocardiography (2DE), 3DE was proved to be accurate and reproducible for the quantification of AVA [9, 10]. Moreover, using the MPR mode, 3DE has been reported to be more precise than 2DE because it overcomes the physiological deformation of aortic annulus during the cardiac cycle [11]. An excellent correlation was also reported between 3DE planimetry of AVA and 2DE planimetry of AVA, which require the use of the Doppler continuity equation [7, 12, 13].

Role of aortic valve area and valvular asymmetry in aortic stenosis

In a recent paediatric study, Bharucha et al. showed that the AVA value obtained by 3DE was not well correlated with 2DE Doppler gradients [14]. In agreement with these data, we found a poor relationship between the maximal and mean Doppler gradients and the AVA calculated by RT-3DTTE. Nevertheless, when considering aortic stenosis for a maximal instantaneous gradient>20mmHg, AVA was significantly smaller for the stenotic valves. Above all, our results suggest that, in children, aortic stenosis strongly depends on the valvular asymmetry. As proved by linear regression analysis, the degree of stenosis is well correlated with the severity of asymmetry and a small/large cusp ratio<0.75 seems to be predictive of an increased risk of stenotic valve. Because aortic stenosis strongly influences prognosis, assessment of valvular symmetry would to be of interest in these patients [15]. Furthermore, even if the prognosis depends on the severity of the valvular lesion at the time of diagnosis, a BAV may have the potential for evolution towards stenosis during life [16]. As 3DE is valuable for clearly depicting the valvular morphology of a BAV [17, 18, 19], this new technique seems to provide good quantification of valvular surfaces and clinicians must keep in mind that a very asymmetric valve might become more stenotic.

Aortic stenosis:pathophysiology

As proved by a catheter-based study, the severity of the stenosis is influenced, in congenital BAV, by the structural geometry of the leaflets [20]. Furthermore, four-dimensional flow magnetic resonance imaging (MRI) showed abnormal helical systolic flow in the ascending aorta of patients with a BAV [21, 22]. This local flow pattern was also reported in isolated ascending aortic aneurysms [23], suggesting that modification of the aortic flow could be responsible for lesions of the ascending aorta. Because a disturbed flow may result in local changes in aortic wall stress and may cause aortic remodelling, asymmetry of the valve might explain the role of the BAV in aortic dilation. By showing a lower small/large cusp ratio in the population of BAV with aortic dilation, our results tend to support this hypothesis.

Study limitations

Several limitations must be outlined. First, even if this 3DE study is one of the largest ever conducted in children with a BAV, we did not show any relationship between the type of BAV and valvular asymmetry. Because certain types of BAV are rare, a study including more patients is necessary to identify whether one particular phenotype is more asymmetric than others. Secondly, no gold standard was used to compare the 3DE planimetric values. Indeed, it would have been ethically difficult to systematically perform cardiac MRI in these children, who are, for the most part, asymptomatic. Because cardiac MRI, which was proved to be efficient for the assessment of valvular morphology in a BAV population [24], sometimes requires general anaesthesia, especially in young children, it did not appear ethical to perform this invasive examination for the sole purpose of the study. Third, we deliberately excluded patients with a BAV type 2 because measuring the small/large cusp ratio was impossible as these valves may be considered as monocuspid (no commissure individualized). However, in our experience, this type of BAV appears to be frequently stenotic: three of the four excluded patients with a BAV type 2 had aortic stenosis. Finally, planimetric measurements of valvular surfaces using 3DE is time consuming because of the off-line analysis.

Conclusions

RT-3DTTE is a feasible and reproducible method of assessing valvular surfaces in the population of children with a BAV. Our study shows that aortic stenosis seems to depend strongly on the asymmetry of the valve. Furthermore, a small AVA assessed by RT-3DTTE is also associated with aortic stenosis. When percutaneous balloon valvuloplasty is considered, this technique may be useful for accurately choosing the balloon size, according to the aortic annulus and the AVA. However, further prospective studies are needed to evaluate this non-invasive imaging technique in daily clinical practice.

Disclosure of interest

The authors declare that they have no conflicts of interest concerning this article.

References

Hoffman J.I., Kaplan S. The incidence of congenital heart disease J Am Coll Cardiol 2002 ;  39 : 1890-1900 [cross-ref]
Ward C. Clinical significance of the bicuspid aortic valve Heart 2000 ;  83 : 81-85 [cross-ref]
Fernandes S.M., Sanders S.P., Khairy P., and al. Morphology of bicuspid aortic valve in children and adolescents J Am Coll Cardiol 2004 ;  44 : 1648-1651 [cross-ref]
Fernandes S.M., Khairy P., Sanders S.P., and al. Bicuspid aortic valve morphology and interventions in the young J Am Coll Cardiol 2007 ;  49 : 2211-2214 [cross-ref]
Espinola-Zavaleta N., Munoz-Castellanos L., Attie F., and al. Anatomic three-dimensional echocardiographic correlation of bicuspid aortic valve J Am Soc Echocardiogr 2003 ;  16 : 46-53 [inter-ref]
Alunni G., Giorgi M., Sartori C., and al. Real time triplane echocardiography in aortic valve stenosis: validation, reliability, and feasibility of a new method for valve area quantification Echocardiography 2010 ;  27 : 644-650 [cross-ref]
Blot-Souletie N., Hebrard A., Acar P., and al. Comparison of accuracy of aortic valve area assessment in aortic stenosis by real time three-dimensional echocardiography in biplane mode versus two-dimensional transthoracic and transesophageal echocardiography Echocardiography 2007 ;  24 : 1065-1072 [cross-ref]
Sievers H.H., Schmidtke C. A classification system for the bicuspid aortic valve from 304 surgical specimens J Thorac Cardiovasc Surg 2007 ;  133 : 1226-1233 [cross-ref]
Kasprzak J.D., Nosir Y.F., Dall’Agata A., and al. Quantification of the aortic valve area in three-dimensional echocardiographic data sets: analysis of orifice overestimation resulting from suboptimal cut-plane selection Am Heart J 1998 ;  135 : 995-1003 [inter-ref]
Suradi H., Byers S., Green-Hess D., and al. Feasibility of using real time “Live 3D” echocardiography to visualize the stenotic aortic valve Echocardiography 2010 ;  27 : 1011-1020 [cross-ref]
Nakai H., Takeuchi M., Yoshitani H., and al. Pitfalls of anatomical aortic valve area measurements using two-dimensional transoesophageal echocardiography and the potential of three-dimensional transoesophageal echocardiography Eur J Echocardiogr 2010 ;  11 : 369-376 [cross-ref]
Goland S., Trento A., Iida K., and al. Assessment of aortic stenosis by three-dimensional echocardiography: an accurate and novel approach Heart 2007 ;  93 : 801-807 [cross-ref]
Handke M., Schafer D.M., Heinrichs G., and al. Quantitative assessment of aortic stenosis by three-dimensional anyplane and three-dimensional volume-rendered echocardiography Echocardiography 2002 ;  19 : 45-53
Bharucha T., Fernandes F., Slorach C., and al. Measurement of effective aortic valve area using three-dimensional echocardiography in children undergoing aortic balloon valvuloplasty for aortic stenosis Echocardiography 2012 ;  29 : 484-491 [cross-ref]
Mahle W.T., Sutherland J.L., Frias P.A. Outcome of isolated bicuspid aortic valve in childhood J Pediatr 2010 ;  157 : 445-449 [cross-ref]
Michelena H.I., Desjardins V.A., Avierinos J.F., and al. Natural history of asymptomatic patients with normally functioning or minimally dysfunctional bicuspid aortic valve in the community Circulation 2008 ;  117 : 2776-2784 [cross-ref]
Acar P., Aggoun Y., Saliba Z., and al. Effect of balloon dilatation on aortic stenosis assessed by 3-dimensional echocardiographic reconstruction Circulation 1999 ;  99 : 2598-2599 [cross-ref]
Sadron Blaye-Felice M.A., Seguela P.E., Arnaudis B., and al. Usefulness of three-dimensional transthoracic echocardiography for the classification of congenital bicuspid aortic valve in children Eur Heart J Cardiovasc Imaging 2012 ;  13 : 1047-1052 [cross-ref]
Singh P., Dutta R., Nanda N.C. Live/real time three-dimensional transthoracic echocardiographic assessment of bicuspid aortic valve morphology Echocardiography 2009 ;  26 : 478-480 [cross-ref]
Richards K.E., Deserranno D., Donal E., and al. Influence of structural geometry on the severity of bicuspid aortic stenosis Am J Physiol Heart Circ Physiol 2004 ;  287 : H1410-H1416
Hope M.D., Hope T.A., Meadows A.K., and al. Bicuspid aortic valve: four-dimensional MR evaluation of ascending aortic systolic flow patterns Radiology 2010 ;  255 : 53-61
Hope M.D., Meadows A.K., Hope T.A., and al. Images in cardiovascular medicine. Evaluation of bicuspid aortic valve and aortic coarctation with 4D flow magnetic resonance imaging Circulation 2008 ;  117 : 2818-2819 [cross-ref]
Weigang E., Kari F.A., Beyersdorf F., and al. Flow-sensitive four-dimensional magnetic resonance imaging: flow patterns in ascending aortic aneurysms Eur J Cardiothorac Surg 2008 ;  34 : 11-16 [cross-ref]
Buchner S., Hulsmann M., Poschenrieder F., and al. Variable phenotypes of bicuspid aortic valve disease: classification by cardiovascular magnetic resonance Heart 2010 ;  96 : 1233-1240 [cross-ref]

1  Thomas Cognet and Pierre-Emmanuel Séguéla contributed equally.


© 2012  Elsevier Masson SAS. All Rights Reserved.
EM-CONSULTE.COM is registrered at the CNIL, déclaration n° 1286925.
As per the Law relating to information storage and personal integrity, you have the right to oppose (art 26 of that law), access (art 34 of that law) and rectify (art 36 of that law) your personal data. You may thus request that your data, should it be inaccurate, incomplete, unclear, outdated, not be used or stored, be corrected, clarified, updated or deleted.
Personal information regarding our website's visitors, including their identity, is confidential.
The owners of this website hereby guarantee to respect the legal confidentiality conditions, applicable in France, and not to disclose this data to third parties.
Close
Article Outline