Archives of cardiovascular diseases
Volume 107, n° 5
pages 308-318 (mai 2014)
Doi : 10.1016/j.acvd.2014.03.006
Received : 27 February 2014 ;  accepted : 28 Mars 2014
Programming implantable cardioverter-defibrillators in primary prevention: Higher or later
Programmation des défibrillateurs implantables en prévention primaire : plus haut ou plus tard
 

Nicolas Clementy , Bertrand Pierre, Edouard Simeon, Bénédicte Lallemand, Laurent Fauchier, Dominique Babuty
 Cardiology Department, Trousseau Hospital, François-Rabelais University, Tours, France 

Corresponding author. Service de Cardiologie B, Hôpital Trousseau, 37044 Tours, France.
Summary

Defibrillator shocks, appropriate or not, are associated with significant morbidity, as they decrease quality of life, can be involved in depression and anxiety, and are known to be proarrhythmic. Most recent data have even shown an association between shocks and overall mortality. As opposed to other defibrillator-related complications, the rate of inappropriate and unnecessary shocks can (and should) be decreased with adequate programming. This review focuses on the different programming strategies and tips available to reduce the rate of shocks in primary prevention patients with left ventricular dysfunction implanted with a defibrillator, as well as some of the manufacturers’ device specificities.

The full text of this article is available in PDF format.
Résumé

Les chocs délivrés par les défibrillateurs automatiques implantables sont associés à une baisse de la qualité de vie des patients implantés (dépression, anxiété), et peuvent avoir un effet pro-arythmogène. Les études les plus récentes ont même montré un lien entre les chocs électriques internes et la mortalité totale. Contrairement aux autres complications, la fréquence de ces chocs peut (et doit) être diminuée grâce à une programmation adaptée. Cette revue de la littérature se focalise plus particulièrement sur les stratégies disponibles pour réduire la fréquence des chocs chez les patients implantés en prévention primaire pour une dysfonction systolique ventriculaire gauche, ainsi que sur les spécificités des appareils des différents constructeurs.

The full text of this article is available in PDF format.

Keywords : Defibrillator, Shock, Prevention

Mots clés : Défibrillateur, Choc, Prévention

Abbreviations : AF, ATP, bpm, ICD, LVEF, SVT, VF, VT


Background

Implantable cardioverter-defibrillators (ICDs) significantly reduce all-cause mortality in the primary prevention of sudden cardiac death in patients at increased risk, especially those with a reduced left ventricular ejection fraction (LVEF). The Multicenter Automatic Defibrillator Implantation Trial II (MADIT-II) first established a strong mortality benefit for the ICD in patients with a LVEF30% and a previous myocardial infarction [1]. The reduction in total mortality was 31% in patients implanted with a prophylactic ICD compared with patients randomized to optimal medical therapy. This effect was durable, with a 34% reduction at 8years by multivariable analyses; six patients had to be implanted to prevent one death [2]. The Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) confirmed these results in patients with a LVEF35% and ischaemic or non-ischaemic cardiomyopathy [3], finally leading to Class I recommendations in the current guidelines [4].

For the past 10years, as the number of prophylactic ICD implantations has drastically increased, so have all the related complications: acute implant-related complications in up to 5% of patients (especially pocket haematoma, haemothorax, pneumothorax, cardiac tamponade); an almost 1% rate of 30-day mortality; lead dislodgement; pocket infection and lead-related endocarditis; lead and device failures; and inappropriate or unnecessary shocks [5].

The rate of inappropriate shocks during the first year of ICD implantation is reported to be 6–10%, affecting 11.5–17.4% of patients in major trials [6, 7]. Such shocks are usually classified as supraventricular tachycardia (SVT)-related or oversensing-related inappropriate discharges. The former are the most frequent, due to the absence or failure of device discrimination algorithms, leading to the misdiagnosis of the supraventricular origin of arrhythmia as ventricular, whereas the latter can be ICD-related (QRS, T-wave or P-wave oversensing, inappropriate programming), lead-related (isolation defect, fracture, dislodgement) or environmental (electromagnetic interference, myopotentials).

The prevalence of unnecessary shocks is more difficult to quantify, as one cannot predict what would have happened if a shock had not been delivered to a ventricular arrhythmia. These are mainly represented by shocks to non-sustained ventricular tachycardias (VTs), which represent the majority of detected VTs. They may also be ICD-related, as antitachycardia pacing (ATP) can induce or accelerate ventricular arrhythmias and cause unnecessary shocks. Finally, the question of whether sustained asymptomatic VT should be treated remains controversial.

It is now well established that shocks, appropriate or not, are associated with significant morbidity, as they decrease quality of life, can be involved in depression and anxiety, and are known to be proarrhythmic [8, 9, 10, 11]. Most recent data have even shown an association between shocks and overall mortality [6, 7]. These data are easily explained: patients with more severe cardiomyopathy and a poorer prognosis usually present with a higher ventricular arrhythmia burden and thus may experience more appropriate shocks; most inappropriate shocks are related to atrial fibrillation (AF), which is one of the main factors associated with a poor outcome in patients with heart failure. Shocks were actually thought to reflect a more advanced cardiac disease. The results of the recent MADIT trial to reduce inappropriate therapy (MADIT-RIT) demonstrated that changing the ICD programming to reduce the rate of ICD shocks might lead to a reduction in overall mortality of up to 55% [12]. Even if the underlying mechanisms still remain unclear, the deleterious effects of shocks themselves, which can cause myocardial damage, may explain these findings. Shocks can be associated with an increased adrenergic response, inflammation and oxidative stress, which eventually may lead to an alteration of the patient's haemodynamics [13, 14, 15].

As opposed to other defibrillator-related complications, the rate of inappropriate and unnecessary shocks can be decreased with adequate programming. This review focuses on the different programming strategies available to reduce the rate of shocks in primary prevention patients with reduced LVEF, implanted with an ICD.

Supraventricular tachycardia discrimination

Discrimination is used to define the ability of the algorithms for differentiating non-ventricular-detected events from true ventricular tachyarrhythmias. The early generations of ICDs had only rate detection. Once the ventricular rate was detected above a preprogrammed rate during a minimal prespecified duration, therapies were automatically delivered. The sensitivity (percentage of clinical ventricular arrhythmias accurately diagnosed as ventricular episodes) was high, but the specificity (percentage of accurately diagnosed ventricular arrhythmias among all ventricular episodes) was rather low, with a high-rate of inappropriate therapies. Algorithms were then progressively implemented in the software to improve specificity, while remaining safe (high sensitivity); they should be used routinely, but in patients with permanent complete atrioventricular block in which rapid ventricular rates can only have a ventricular origin.

Single-chamber algorithms

Three criteria are mainly used: the sudden onset criterion is able to discriminate sinus tachycardia from VT, as the former usually displays a gradual rate increase; the stability criterion can discriminate AF from VT, as the former usually displays irregular RR intervals; the morphology criterion is used to discriminate supraventricular from ventricular arrhythmia by comparing the similarity of the ventricular electrogram during tachycardia with a reference electrogram template acquired during a rhythm in which the supraventricular origin is certain.

These three main algorithms can usually be used in combination to improve discrimination accuracy. In a tiered-therapy cardioverter-defibrillator study, a programmed stability of 40ms (interval maximum variation between detected interval and any of the three previous intervals) decreased detection of AF by 95–99%, with a sensitivity for true VT episodes of 100% [16]. In the same study, a programmed sudden onset ratio of 87% (detected interval divided by the mean of the four preceding intervals) rejected 98% of sinus tachycardia episodes, while 0.5% of true VT episodes were misdiagnosed. A strategy of programming both stability and onset criteria has proven effective in decreasing the rate of inappropriate therapies by>50% [17]. Some manufacturers have since implemented morphology algorithms (Table 1).

Dual-chamber algorithms

Addition of an atrial lead allows analysis of the atrioventricular relationship during arrhythmia. Each manufacturer uses its own approach, but the main principles are similar:

if the ventricular electrograms are more frequent than the atrial electrograms (V>A), which is actually the case in>90% of VTs, the ventricular origin is certain;
if the number of detected ventricular electrograms equals those from the atrium (V=A), the supraventricular origin is more probable, although VT with a 1:1 retrograde conduction cannot be ruled out;
if the atrial electrograms are more frequent (A>V), the supraventricular origin is also more probable, although dual-tachycardia should also not be ruled out (VT in a patient with AF, for instance).

Single-chamber algorithms and PR association analyses are then implemented to achieve a final diagnosis (Figure 1, Figure 2, Figure 3, Figure 4, Figure 5).



Figure 1


Figure 1. 

Biotronik SMART detection algorithm for dual-chamber implantable cardioverter-defibrillators. Atrial (PP intervals) and ventricular (RR intervals) cycle lengths are first compared; the rhythm is then checked for stability (RR, PP and PR intervals) in the RR=PP diagnostic branch, and for multiplicity (N:1 atrioventricular association) in the RR>PP branch. The RR<PP branch is diagnostic for ventricular tachycardia (VT). SVT: supraventricular tachycardia.

Zoom



Figure 2


Figure 2. 

Boston Scientific Rhythm ID detection algorithm for dual-chamber implantable cardioverter-defibrillators. Diagnosis of ventricular tachycardia (VT) requires either a higher ventricular rate (V rate) than atrial rate (A rate) or a non-correlated morphology associated with stable RR intervals. SVT: supraventricular tachycardia; VTC: vector timing and correlation.

Zoom



Figure 3


Figure 3. 

Medtronic enhanced PR Logic/Wavelet detection algorithm for dual-chamber implantable cardioverter-defibrillators. Tachyarrhythmias are classified according to prespecified patterns, according to atrioventricular (AV) relationship, morphology and regularity. A far-field R-wave (FFRW) algorithm is also used to discriminate ventriculoatrial crosstalk oversensing. A: atrial; AF: atrial fibrillation; ST: sinus tachycardia; SVT: supraventricular tachycardia; V: ventricular; VT: ventricular tachycardia.

Zoom



Figure 4


Figure 4. 

St. Jude Medical Discrimination algorithm for dual-chamber implantable cardioverter-defibrillators. Three decision branches according to atrioventricular relationship are initially used; discrimination is then performed using programmable single-chamber algorithms (morphology, sudden onset and stability). Initial cavity (chamber onset, atrial [A] or ventricular [V]) helps to discriminate 1:1 atrial tachycardia in the A=V branch. AVA: atrial ventricular association; MD match: morphology similar to template; SVT: supraventricular tachycardia; VT: ventricular tachycardia.

Zoom



Figure 5


Figure 5. 

Sorin Group PARAD+ detection algorithm for dual-chamber implantable cardioverter-defibrillators. Discrimination is based on analysis of stability, PR association (N:1 atrioventricular association), long cycle occurrence (VTLC), sudden onset and chamber onset (atrial [A] or ventricular [V]), in this specific order. AF: atrial fibrillation; AFl: atrial flutter; AT: atrial tachycardia ST: sinus tachycardia; VT: ventricular tachycardia.

Zoom

Single or dual-chamber?

If the Detect Supraventricular Tachycardia (DETECT SVT) study showed better specificity in dual-chamber patients, the rate of inappropriate therapies was significantly higher than in single-chamber patients [18]; this might be related to the prerequisite of reliable atrial sensing for accurate performance of dual-chamber algorithms. Atrial undersensing, especially frequent when AF occurs, atrial undercounting due to atrial blanking periods, and atrial oversensing (often due to far-field R-wave sensing), may all impair dual-chamber discrimination and lead to inappropriate therapies. Moreover, it has been demonstrated that dual-chamber ICD implantations are associated with a 50% increase in perioperative adverse events and a two-fold increase in the perioperative mortality rate compared with single-chamber IDC implantations [19]. It is now accepted that dual-chamber ICDs should not be implanted in order to improve discrimination, but should be implanted in patients without permanent AF and needing to be paced.

Specificities of biventricular devices

Discrimination during cardiac resynchronization therapy usually uses the same dual-chamber algorithms, but in patients with chronic AF. As patients with biventricular devices are supposed to be paced permanently, a morphology algorithm may appear less relevant, as a template of the spontaneous QRS might not be obtained. In Boston Scientific devices, for instance, a stability/onset algorithm is favoured over morphology.

Subcutaneous defibrillator

Cameron Health (Boston Scientific) has developed a specific discrimination algorithm for the S-ICD System, which uses, in a so-called “conditional zone”, three criteria for analysis of morphology:

comparison of the QRS morphology with a memorized template (static morphology);
comparison of the QRS morphology with the morphology of the previous complex (dynamic morphology);
change in QRS width.

Three far-field sensing vectors are available for the analysis of morphology. A sensitivity for appropriate detection of ventricular tachyarrhythmias of 100% and a specificity for supraventricular arrhythmias of 98% were shown in the START study, in which S-ICD discrimination was even better than some conventional transvenous systems [20]. A standard shock zone is also programmable with rate-only detection.

Discrimination algorithms have become mandatory to reduce the rate of inappropriate detection of supraventricular arrhythmias and their subsequent treatment. The algorithms are rather diverse because of proprietary patents from the different manufacturers. A precise understanding of these algorithms is useful to appropriately reprogramme the device after a therapy is delivered.

Oversensing discrimination
Device-related oversensing

Around 10% of inappropriate therapies result from noise/artefact and oversensing [21]. Specific algorithms have been developed to avoid signal-related oversensing, and T-wave oversensing has become rare (2%), although it might still occur in specific situations (electrolyte abnormalities, drugs) or patients (low ventricular electrogram amplitude, hypertrophic cardiomyopathy, short/long QT and Brugada syndromes), usually causing ventricular double counting. All latest generation ICDs allow automatic adjustment of the sensitivity threshold on the detected R-wave amplitude, with smooth decay with time, usually preventing T-wave detection. Some manufacturers, such as Medtronic, Boston Scientific and St. Jude Medical, rely on algorithms to improve signal processing (frequency analysis with band-pass filters) and thus noise-to-signal ratio [22, 23]. Biotronik and St. Jude Medical enable the adjustment of detection algorithms (threshold start value and decay delay), but these must be used with caution as they might compromise sensing of ventricular tachyarrhythmias.

Lead-related oversensing

Inappropriate shocks on oversensing are mainly related to lead fracture (up to 1.1% per year) [21]. Kleemann et al. reported a 40% rate of ICD lead dysfunction at 8years [24]. These failures are often lead material and design specific, such as in the Medtronic Fidelis and St. Jude Medical Riata leads. Some algorithms have been developed based on signal processing, with analysis of both near-field and far-field electrograms (noise usually manifests on near-field signals), avoiding 83.7% of inappropriate shocks in a computer model study [23]. Coupled with variation in lead impedance measurements, the Medtronic algorithm, for instance, automatically prolongs detection duration and alerts the patient; this allowed a drastic reduction in the rate of inappropriate shocks in patients implanted specifically with Sprint Fidelis leads [25], but may also be useful with non-Medtronic leads [26]. Detection of ventricular oversensing/noise events then withholds therapy delivery, without compromising sensitivity to ventricular arrhythmias. Combined with home monitoring, this allows rapid and adequate management of ICD lead defects in these patients (device deactivation and surgical revision) [27, 28].

Other detection algorithms
Redetection

Occurring after therapy delivery, redetection determines whether an episode has terminated. Typically, three to eight intervals classified as sinus are required to terminate the treatment sequence. It is important to point out that during redetection, the discrimination algorithm uses fewer algorithms (or sometimes none) than are necessary during initial detection, depending on the manufacturer, to avoid delaying the next therapy. As stated by Koneru et al., reducing the number of redetection intervals might be interesting in some patients with incessant VT episodes associated with temporary efficacy of ATP, permitting repetition of the first sequence without moving towards more aggressive therapies [29]. Conversely, redetection duration may be increased if VT termination by ATP is systematically preceded by a transient acceleration.

Reconfirmation

These algorithms ensure whether an arrhythmia has self-terminated after capacitor charging. After charge completion, a window of a few detected beats allows shock abortion whenever detected intervals are longer than the maximum-programmed interval in the therapy zone. Inappropriate shocks may still be delivered if the therapy zone (including the monitoring zone in some devices) is programmed to a long cycle length when a sinus tachycardia follows ventricular arrhythmia spontaneous resumption. A new shock algorithm from Medtronic addresses this concern by aborting therapy whenever redetected intervals are60ms longer than the detected VT intervals.

Timers

These algorithms (called Sustained Rate Duration, High-Rate Time Out, SVT Time Out or Sustained VT) induce therapy delivery whenever an event (supraventricular or ventricular) lasts longer than a programmable duration value. There is a consensus that these algorithms should be turned to OFF to avoid inappropriate therapies for sustained SVT events.

Antitachycardia pacing

This mode of therapy consists of an overdrive pacing that employs ventricular stimulation at intervals that are shorter than the tachycardia cycle length. ATP offers several advantages: it has been proven to be a painless but effective therapy, reducing the necessity of shock delivery and hospitalization, improving patient quality of life, and increasing battery longevity.

Slow ventricular tachycardia

Below 180bpm, ATP has been proven efficient in treating VT episodes [30], but therapies for low ventricular rates should not be initially programmed in primary prevention patients. Conversely, adding a monitoring zone for slow VT may be interesting to document symptomatic events.

Fast ventricular tachycardia

Above 180bpm, ATP has been proven to efficiently treat most monomorphic VT episodes. Grimm et al. reported 86% of episodes terminated with ATP (mean VT rate: 204bpm) [31]. In the PROVE trial, ATP (two bursts, eight stimuli, 88% of tachycardia cycle length) terminated 92% of VT episodes [32]. The PITAGORA (Project for the Investigation and Treatment of ventricular Arrhythmias: a General Observational Registry on Antitachycardia pacing efficacy) ICD trial randomly compared two ATP strategies for fast VT episodes>187bpm (one sequence of 88% coupling interval, eight-pulse burst versus one sequence of 91% coupling interval, eight-pulse ramp) and demonstrated a 75% efficacy for burst versus 54% for ramp [33]. A 15-pulse ATP strategy did not show any superiority over an eight-pulse strategy [34]. More recently, Martins et al. showed that one or two sequences of ATP terminated>98% of fast VT episodes (average of 1.1 ATP sequences), but programming up to 10 attempts (five bursts then five ramps) remained safe, while avoiding shocks in 12.4% of patients [35].

If ATP programming is selected, there is enough evidence to suggest that at least two sequences of ATP should be programmed, with eight stimuli bursts at 88% of the tachycardia cycle length.

Ventricular fibrillation zone

In the ventricular fibrillation (VF) zone, ATP should be programmed before or during charging, when available, as it has proven safe and efficient. Schoels et al. reported efficacies for monomorphic and polymorphic VT episodes of 77% and 44%, respectively [36].

Shocks

Although the subject remains controversial, current data do not support the systematic testing of the defibrillation threshold or safety margin [37]. Therefore, until further data are available, programming maximum energy shocks routinely can be recommended. Beyond better efficacy, they have not been proven to be more painful than lower energy shocks and the slightly longer capacitor charging time may allow arrhythmias (appropriately or inappropriately detected) to self terminate.

Zone programming

In primary prevention patients, two programming options may appear relevant: single zone programming with a VF zone (discrimination OFF) and a monitoring-only zone below; dual-zone programming with a VT zone (discrimination ON) with ATP followed by shocks, and a VF zone above.

The latter strategy may not necessarily be more efficient than programming for a single (VF) zone. Duncan et al., in a retrospective study, compared patients programmed for a single VF zone (mean>193bpm) with patients programmed for two-zones (mean VT zone>171bpm, mean VF zone>205bpm) [38]. The two-zone group actually received more appropriate shocks than the group with a single VF zone (22% and 12%, respectively). As efficient as it can be, ATP is only useful if a true SVT event is not misdiagnosed as VT. Moreover, ATP programming may not always prevent symptoms, as it may only delay the occurrence of an efficient therapy; it may also accelerate non-sustained VT and provoke unnecessary ICD shock. In the study by Grimm et al., this was the case in up to 14% of VT episodes and 48% of patients [31]. More recently, Schukro et al. showed that 8.5% of patients presented with accelerated ventricular tachyarrhythmias within a mean follow-up of>5years, mainly caused by appropriate ATP, and that this was associated with a higher mortality rate [39].

The same concerns can be applied to inappropriate therapies. In MADIT-II, the incidence of inappropriate shocks was 11.5% [6]. Similarly, in the Pacing Fast VT Reduces Shock Therapies (PainFREE Rx II) trial, 15% of primary prevention patients had inappropriate therapies, despite the systematic use of ATP for fast VT episodes (>188bpm) [40]. In the Detect Supraventricular Tachycardia Study, inappropriate therapies remained very frequent and 31% of SVT episodes were inappropriately detected [18]. In a “real-life” prospective study of 1544 patients (mean follow-up 41months) at least one inappropriate shock occurred in 13% of ICD recipients [41]. Devices were programmed with a monitoring zone of 150–188bpm, a VT zone with two bursts of ATP of 188–210bpm and a shock-only zone above. Inappropriate shocks were related to SVT or sinus tachycardia episodes in 76% of cases (11.5% of patients). In the study by Duncan et al., inappropriate shocks were also more frequent in the two-zone programming group (10% vs 2%) [38]. There was no SVT-related inappropriate shock in the single VF zone group.

Detection duration

Programming long detection intervals is effective in reducing both appropriate and inappropriate therapies. The EMPIRIC (Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-Defibrillators) trial has already shown that standardized programming with longer intervals (18 of 24) in the VT zone is associated with a significant reduction in shocks of any cause [42]. The Primary Prevention Parameters Evaluation (PREPARE) study compared a long-delay strategy (30 of 40 detection intervals for fast VT>182bpm, 30 of 40 for VF zone>250bpm) with a standard programming cohort [43]. After 12months of follow-up, the rate of appropriate shocks in the long-delay group was 8.5% (43% reduction) and the rate of inappropriate shocks was 3.6% (a 52% reduction). The RELEVANT (Role of long dEtection window programming in patients with LEft VentriculAr dysfunction, Non-ischemic eTiology in primary prevention treated with a biventricular ICD) study used 30 of 40 detection intervals in non-ischaemic patients, with an 84% reduction in inappropriate shocks [44]. Interestingly,>90% of VT episodes detected after 12 intervals spontaneously resumed before reaching 30 intervals. Most VT episodes are non-sustained, and if ATP is very efficient, it is also probably useless for the majority of episodes and may lead to a significant amount of unnecessary shocks. Indeed, an 88% reduction in appropriate therapies was observed in the long-delay group. There was also a trend towards reduction in appropriate shocks in the long-delay group, although this was not significant. The usefulness of a strategy of programming a 30 of 40 intervals detection window was recently confirmed in the randomized ADVANCE III trial [45].

The MADIT-RIT trial randomized 1500 primary prevention patients into three groups:

a standard ICD programming group (170–199bpm VT zone with a 2.5s delay, ≥200bpm VF zone with a 1.0s delay);
a high-rate cut-off programming group (170–199bpm monitoring zone, ≥200bpm VF zone with a 2.5s delay);
a prolonged-delay programming group (170–199bpm VT1 zone with a 60s delay, 200–249bpm VT2 zone with a 12s delay, ≥250bpm VF zone with a 2.5s delay) [12].

After a mean follow-up of 1.4years, the reduction in occurrence of first inappropriate therapy in the long-delay group compared with in the standard programming group was 76%. The reduction in all-cause mortality was 44%.

When treating in the VT zone is decided upon, it is safe to say that detection duration should exceed 30 of 40 intervals in Medtronic devices before ATP, and eventually shocks, are delivered. Appropriate detection timings with the (very) different classification and binning algorithms of other major manufacturers still wait to be assessed, but detection should be programmed in order to exceed 10s between tachycardia onset and therapy delivery [46, 47]. However, despite a drastic improvement in discrimination algorithms, SVT episodes may still be inappropriately detected and then treated if sustained. Thus, in the RELEVANT study, around 90% of inappropriate episodes were related to SVT, both in the group programmed with longer detection intervals and in the control group.

Detection rate

In the SCD-HeFT trial, a single VF zone at 188bpm was programmed (18 of 24 beats), leading to a high incidence of shocks for VT episodes that may have been asymptomatic and non-sustained [7]. Some of these appropriate shocks may then have been unnecessary therapies for appropriately detected episodes. Indeed, the PainFREE Rx II trial found that approximately 34% of the fast VT episodes terminated spontaneously during capacitor charging [40]. Inappropriate shocks are also very frequent with a low rate single zone (up to 17.4% in SCD-HeFT), as SVT episodes are not discriminated.

In a recent study, we showed that high-rate cut-off programming (single VF zone>220bpm) appeared to be safe and effective in maintaining a low rate of inappropriate shocks (6.6% of patients at 40months) [48]. SVT-related inappropriate shocks were only present in 2.7% of the patients after a mean follow-up of 3.5years. Most inappropriate shocks were related to oversensing (58%). The maximum ventricular rate during SVT should rarely exceed 220bpm, as confirmed by the results of the Altitude Reduces Study, which showed a majority of inappropriate shocks for an event rate<200bpm and a majority of appropriate shocks for an event rate200bpm [49]. The authors concluded that increasing rate detection to200bpm resulted in a four-fold reduction in overall shock risk of both appropriate and inappropriate therapies and was not associated with excess mortality.

The MADIT-RIT trial confirmed these results, with a 79% reduction in first inappropriate therapy occurrence and even a 55% reduction in all-cause mortality in the high-rate group (cut-off at 200bpm), without difference in risk of syncope [12].

Programming a single-therapy zone without discrimination (VF zone)<200bpm should not be performed in primary prevention patients, as this strategy is related to a high-rate of both appropriate and inappropriate shocks.

Higher or later?

No direct comparison has been made between high-rate and long-delay strategies, as it was not prespecified in the MADIT-RIT study design [50]. Both strategies are succinctly compared in Table 2.

Overall, if these strategies can be used safely, as discussed previously, programming one single VF zone should be associated with a high-rate cut-off (≥200bpm), while programming therapies in the VT zone should be associated with rather long detection duration (≥10s) (Figure 6).



Figure 6


Figure 6. 

Examples of programming strategies in primary prevention patients. A. Long-delay programming requires up to three zones: a monitoring zone, a ventricular tachycardia (VT) zone with a number of detection intervals (NDI) of 30 or more below 200beats per minute (bpm) (≥10s detection time), which might even be extended up to 60s as in MADIT-RIT, and a ventricular fibrillation (VF) zone. Antitachycardia pacing (ATP) should also be prioritized before shocks below 200bpm. Over 200bpm, programming an FVT zone with ATP up to 250bpm is also possible. B. High-rate cut-off programming uses a monitoring zone below 200bpm, and a shock zone above. ATP burst should be used before or during charging in the shock zone.

Zoom

Conclusions

Careful programming of ICDs in primary prevention should remain a major concern for physicians after implantation (Table 3). Several studies have recently raised the importance of reducing both inappropriate and unnecessary therapies, and manufacturers have made several tools available in order to achieve this goal. The recent results from the MADIT-RIT trial, showing the effect of adequate programming in reducing overall mortality in these patients, have made this programming phase even more crucial.

Disclosure of interest

Nicolas Clementy received consulting fees and/or honoraria from Boston Scientific, Medtronic and St. Jude Medical.

Bertrand Pierre received consulting fees and/or honoraria from Biotronik, Boston Scientific, Medtronic, St. Jude Medical and Sorin Group.

Laurent Fauchier received consulting fees and/or honoraria from Boston Scientific, Medtronic and Sorin Group.

Dominique Babuty received consulting fees and/or honoraria from Biotronik, Boston Scientific, Medtronic, St. Jude Medical and Sorin Group.


Acknowledgements

The authors especially want to thank the following people for their expert technical assistance: Vincent Alexandre and Anthony Rabret, Biotronik; Cédric Durand and Pascal Guesdon, Boston Scientific; Jacques Raad and Fabrice Chomienne, Medtronic; Jean-François Babouhot, St. Jude Medical; Adonis Kobeissi, Sorin Group.

References

Moss A.J., Zareba W., Hall W.J., and al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction N Engl J Med 2002 ;  346 : 877-883 [cross-ref]
Goldenberg I., Gillespie J., Moss A.J., and al. Long-term benefit of primary prevention with an implantable cardioverter-defibrillator: an extended 8-year follow-up study of the Multicenter Automatic Defibrillator Implantation Trial II Circulation 2010 ;  122 : 1265-1271 [cross-ref]
Bardy G.H., Lee K.L., Mark D.B., and al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure N Engl J Med 2005 ;  352 : 225-237 [cross-ref]
Zipes D.P., Camm A.J., Borggrefe M., and al. ACC/AHA/ESC 2006 guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: a report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death) developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society Europace 2006 ;  8 : 746-837
Atwater B.D., Daubert J.P. Implantable cardioverter defibrillators: risks accompany the life-saving benefits Heart 2012 ;  98 : 764-772 [cross-ref]
Daubert J.P., Zareba W., Cannom D.S., and al. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact J Am Coll Cardiol 2008 ;  51 : 1357-1365 [cross-ref]
Poole J.E., Johnson G.W., Hellkamp A.S., and al. Prognostic importance of defibrillator shocks in patients with heart failure N Engl J Med 2008 ;  359 : 1009-1017 [cross-ref]
Francis J., Johnson B., Niehaus M. Quality of life in patients with implantable cardioverter defibrillators Indian Pacing Electrophysiol J 2006 ;  6 : 173-181
Messali A., Thomas O., Chauvin M., and al. Death due to an implantable cardioverter defibrillator J Cardiovasc Electrophysiol 2004 ;  15 : 953-956 [cross-ref]
Mitchell L.B., Pineda E.A., Titus J.L., and al. Sudden death in patients with implantable cardioverter defibrillators: the importance of post-shock electromechanical dissociation J Am Coll Cardiol 2002 ;  39 : 1323-1328 [cross-ref]
Sears S.F., Sowell L.D., Kuhl E.A., and al. The ICD shock and stress management program: a randomized trial of psychosocial treatment to optimize quality of life in ICD patients Pacing Clin Electrophysiol 2007 ;  30 : 858-864 [cross-ref]
Moss A.J., Schuger C., Beck C.A., and al. Reduction in inappropriate therapy and mortality through ICD programming N Engl J Med 2012 ;  367 : 2275-2283 [cross-ref]
Bode F., Wiegand U., Raasch W., and al. Differential effects of defibrillation on systemic and cardiac sympathetic activity Heart 1998 ;  79 : 560-567
Schluter T., Baum H., Plewan A., and al. Effects of implantable cardioverter defibrillator implantation and shock application on biochemical markers of myocardial damage Clin Chem 2001 ;  47 : 459-463
Tokano T., Bach D., Chang J., and al. Effect of ventricular shock strength on cardiac hemodynamics J Cardiovasc Electrophysiol 1998 ;  9 : 791-797 [cross-ref]
Swerdlow C.D., Chen P.S., Kass R.M., and al. Discrimination of ventricular tachycardia from sinus tachycardia and atrial fibrillation in a tiered-therapy cardioverter-defibrillator J Am Coll Cardiol 1994 ;  23 : 1342-1355 [cross-ref]
Weber M., Bocker D., Bansch D., and al. Efficacy and safety of the initial use of stability and onset criteria in implantable cardioverter defibrillators J Cardiovasc Electrophysiol 1999 ;  10 : 145-153 [cross-ref]
Friedman P.A., McClelland R.L., Bamlet W.R., and al. Dual-chamber versus single-chamber detection enhancements for implantable defibrillator rhythm diagnosis: the detect supraventricular tachycardia study Circulation 2006 ;  113 : 2871-2879 [cross-ref]
Dewland T.A., Pellegrini C.N., Wang Y., and al. Dual-chamber implantable cardioverter-defibrillator selection is associated with increased complication rates and mortality among patients enrolled in the NCDR implantable cardioverter-defibrillator registry J Am Coll Cardiol 2011 ;  58 : 1007-1013 [cross-ref]
Gold M.R., Theuns D.A., Knight B.P., and al. Head-to-head comparison of arrhythmia discrimination performance of subcutaneous and transvenous ICD arrhythmia detection algorithms: the START study J Cardiovasc Electrophysiol 2012 ;  23 : 359-366 [cross-ref]
Powell B.D., Asirvatham S.J., Perschbacher D.L., and al. Noise, artifact, and oversensing related inappropriate ICD shock evaluation: ALTITUDE noise study Pacing Clin Electrophysiol 2012 ;  35 : 863-869 [cross-ref]
Cao J., Gillberg J.M., Swerdlow C.D. A fully automatic, implantable cardioverter-defibrillator algorithm to prevent inappropriate detection of ventricular tachycardia or fibrillation due to T-wave oversensing in spontaneous rhythm Heart Rhythm 2012 ;  9 : 522-530 [cross-ref]
Volosin K.J., Exner D.V., Wathen M.S., and al. Combining shock reduction strategies to enhance ICD therapy: a role for computer modeling J Cardiovasc Electrophysiol 2011 ;  22 : 280-289 [cross-ref]
Kleemann T., Becker T., Doenges K., and al. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of>10years Circulation 2007 ;  115 : 2474-2480 [cross-ref]
Swerdlow C.D., Gunderson B.D., Ousdigian K.T., and al. Downloadable software algorithm reduces inappropriate shocks caused by implantable cardioverter-defibrillator lead fractures: a prospective study Circulation 2010 ;  122 : 1449-1455 [cross-ref]
Kneller J., Delacey W., Wood M.A., and al. Detection of a Riata insulation failure by the Medtronic Lead Integrity Alert Europace 2012 ;  14 : 1215-1216 [cross-ref]
Crossley G.H., Boyle A., Vitense H., and al. The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) trial: the value of wireless remote monitoring with automatic clinician alerts J Am Coll Cardiol 2011 ;  57 : 1181-1189 [cross-ref]
Varma N., Epstein A.E., Irimpen A., and al. Efficacy and safety of automatic remote monitoring for implantable cardioverter-defibrillator follow-up: the Lumos-T Safely Reduces Routine Office Device Follow-up (TRUST) trial Circulation 2010 ;  122 : 325-332 [cross-ref]
Koneru J.N., Swerdlow C.D., Wood M.A., and al. Minimizing inappropriate or “unnecessary” implantable cardioverter-defibrillator shocks: appropriate programming Circ Arrhythm Electrophysiol 2011 ;  4 : 778-790 [cross-ref]
Sadoul N., Mletzko R., Anselme F., and al. Incidence and clinical relevance of slow ventricular tachycardia in implantable cardioverter-defibrillator recipients: an international multicenter prospective study Circulation 2005 ;  112 : 946-953 [cross-ref]
Grimm W., Plachta E., Maisch B. Antitachycardia pacing for spontaneous rapid ventricular tachycardia in patients with prophylactic cardioverter-defibrillator therapy Pacing Clin Electrophysiol 2006 ;  29 : 759-764 [cross-ref]
Saeed M., Neason C.G., Razavi M., and al. Programming antitachycardia pacing for primary prevention in patients with implantable cardioverter defibrillators: results from the PROVE trial J Cardiovasc Electrophysiol 2010 ;  21 : 1349-1354 [cross-ref]
Gulizia M.M., Piraino L., Scherillo M., and al. A randomized study to compare ramp versus burst antitachycardia pacing therapies to treat fast ventricular tachyarrhythmias in patients with implantable cardioverter defibrillators: the PITAGORA ICD trial Circ Arrhythm Electrophysiol 2009 ;  2 : 146-153 [cross-ref]
Santini M., Lunati M., Defaye P., and al. Prospective multicenter randomized trial of fast ventricular tachycardia termination by prolonged versus conventional anti-tachyarrhythmia burst pacing in implantable cardioverter-defibrillator patients-Atp DeliVery for pAiNless ICD thErapy (ADVANCE-D) Trial results J Interv Card Electrophysiol 2010 ;  27 : 127-135 [cross-ref]
Martins R.P., Blangy H., Muresan L., and al. Safety and efficacy of programming a high number of antitachycardia pacing attempts for fast ventricular tachycardia: a prospective study Europace 2012 ;  14 : 1457-1464 [cross-ref]
Schoels W., Steinhaus D., Johnson W.B., and al. Optimizing implantable cardioverter-defibrillator treatment of rapid ventricular tachycardia: antitachycardia pacing therapy during charging Heart Rhythm 2007 ;  4 : 879-885 [cross-ref]
Stavrakis S., Patel N.H., Reynolds D.W. Defibrillation threshold testing does not predict clinical outcomes during long-term follow-up: a meta-analysis Pacing Clin Electrophysiol 2013 ;  36 : 1402-1408 [cross-ref]
Duncan E., Thomas G., Johns N., and al. Do traditional VT zones improve outcome in primary prevention ICD patients? Pacing Clin Electrophysiol 2010 ;  33 : 1353-1358 [cross-ref]
Schukro C., Leitner L., Siebermair J., and al. Impact of accelerated ventricular tachyarrhythmias on mortality in patients with implantable cardioverter-defibrillator therapy Int J Cardiol 2013 ;  167 : 3006-3010 [cross-ref]
Wathen M.S., DeGroot P.J., Sweeney M.O., and al. Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) trial results Circulation 2004 ;  110 : 2591-2596 [cross-ref]
van Rees J.B., Borleffs C.J., de Bie M.K., and al. Inappropriate implantable cardioverter-defibrillator shocks: incidence, predictors, and impact on mortality J Am Coll Cardiol 2011 ;  57 : 556-562 [cross-ref]
Wilkoff B.L., Ousdigian K.T., Sterns L.D., and al. A comparison of empiric to physician-tailored programming of implantable cardioverter-defibrillators: results from the prospective randomized multicenter EMPIRIC trial J Am Coll Cardiol 2006 ;  48 : 330-339 [cross-ref]
Wilkoff B.L., Williamson B.D., Stern R.S., and al. Strategic programming of detection and therapy parameters in implantable cardioverter-defibrillators reduces shocks in primary prevention patients: results from the PREPARE (Primary Prevention Parameters Evaluation) study J Am Coll Cardiol 2008 ;  52 : 541-550 [cross-ref]
Gasparini M., Menozzi C., Proclemer A., and al. A simplified biventricular defibrillator with fixed long detection intervals reduces implantable cardioverter defibrillator (ICD) interventions and heart failure hospitalizations in patients with non-ischaemic cardiomyopathy implanted for primary prevention: the RELEVANT [Role of long dEtection window programming in patients with LEft VentriculAr dysfunction Non-ischemic eTiology in primary prevention treated with a biventricular ICD] study Eur Heart J 2009 ;  30 : 2758-2767 [cross-ref]
Gasparini M., Proclemer A., Klersy C., and al. Effect of long-detection interval vs standard-detection interval for implantable cardioverter-defibrillators on antitachycardia pacing and shock delivery: the ADVANCE III randomized clinical trial JAMA 2013 ;  309 : 1903-1911 [cross-ref]
Khairy P., Mansour F. Counting down before blasting off Heart Rhythm 2011 ;  8 : e2-e4
Lim H.S. To the editor – programming tips and practical programming Heart Rhythm 2011 ;  8 : e1-e2 [cross-ref]
Clementy N., Pierre B., Lallemand B., and al. Long-term follow-up on high-rate cut-off programming for implantable cardioverter defibrillators in primary prevention patients with left ventricular systolic dysfunction Europace 2012 ;  14 : 968-974 [cross-ref]
Gilliam F.R., Hayes D.L., Boehmer J.P., and al. Real world evaluation of dual-zone ICD and CRT-D programming compared to single-zone programming: the ALTITUDE REDUCES study J Cardiovasc Electrophysiol 2011 ;  22 : 1023-1029 [cross-ref]
Clementy N., Fauchier L. A clinical trial of ICD programming N Engl J Med 2013 ;  368 : 964-965



© 2014  Elsevier Masson SAS. All Rights Reserved.
EM-CONSULTE.COM is registrered at the CNIL, déclaration n° 1286925.
As per the Law relating to information storage and personal integrity, you have the right to oppose (art 26 of that law), access (art 34 of that law) and rectify (art 36 of that law) your personal data. You may thus request that your data, should it be inaccurate, incomplete, unclear, outdated, not be used or stored, be corrected, clarified, updated or deleted.
Personal information regarding our website's visitors, including their identity, is confidential.
The owners of this website hereby guarantee to respect the legal confidentiality conditions, applicable in France, and not to disclose this data to third parties.
Close
Article Outline