Suscribirse

Cementless Corailtm femoral stems with laser neck etching: Long-term survival, rupture rate and risk factors in 295 stems - 13/02/16

Doi : 10.1016/j.otsr.2015.10.009 
A. Merini a, A. Viste a, b, c, d, , R. Desmarchelier a, M.-H. Fessy a, b, c, d
a Hospices civils de Lyon, centre hospitalier Lyon-Sud, service de chirurgie orthopedique et traumatologique, 165, chemin Grand-Revoyet, 69495 Pierre-Bénite cedex, France 
b Université de Lyon, 69622 Lyon, France 
c Université Claude-Bernard Lyon 1, 43, boulevard du 11-Novembre, 69100 Villeurbanne, France 
d IFSTTAR, UMRT 9406, laboratoire de biomécanique et mécanique des chocs, 25, avenue Mitterrand, 69500 Bron, France 

Corresponding author. Service de chirurgie orthopédique et traumatologique, centre hospitalier Lyon-Sud, 165, chemin du Grand-Revoyet, 69495 Pierre-Bénite cedex, France. Tel.: +33 4 78 86 37 38; fax: +33 4 78 86 59 34.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Introduction

Implant neck fracture involving a non-modular femoral stem is rare in primary total hip arthroplasty (THA). Occasional cases have been reported following laser etching of the Corailtm stem, but risk factors have not been precisely determined. We therefore performed a retrospective study on a series of Corailtm stems with laser neck etching, in order to: (1) determine the exact implant neck fracture rate at 10 years, and (2) identify associated risk factors.

Hypothesis

Laser etching increases the rate of implant neck fracture.

Materials and methods

Between October 2002 and December 2003, 295 THAs were consecutively performed using the Corailtm stem with laser neck etching, in 286 patients: 151 male (53%), 135 female (47%); mean age, 63 years (range, 18–89 years); mean weight, 73kg (range, 45–120kg). Stems were standard in 240 cases (81%) and lateralized in 55 (19%). The main assessment criterion was stem replacement for implant neck fracture.

Results

At a mean 10 years’ follow-up (range, 1–11 years), 11 patients were lost to follow-up (4%) and 35 had died (12%) (with stem in situ). Overall 10-year stem survival was 91% (95% CI: [87–94%]). Sixteen patients (5.4%) underwent revision surgery for implant neck fracture, 6 (2%) bone and joint infection and in 4 cases (1.3%) the stem was replaced preventively for fracture risk suspected during a revision procedure on the cup. All fractures were of the fatigue type, implicating implant neck laser etching. Mean time to fracture was 4.5 years (range, 1.4–9.8 years). Risk factors comprised: weight>80kg (P=0.002) (OR=5.7; 95% CI: 1.9–17), age<60 years (P=0.02) (OR=3.4; 95% CI: 1.2–9.6), male gender (P=0.01) (OR=14.8; 95% CI: 1.9–113) and lateralized stem (P<0.001) (OR=6.5, 95% CI: 2.3–18).

Conclusion

The present 5.4% fracture rate was higher than in registry data (<1%). Fracture mechanisms involved excessive stress in an area under tension, leading to fatigue fracture. Male gender, high weight and young age were risk factors, as in the literature for fatigue fracture. Location and depth of laser etching induced fatigue fracture. The study demonstrated that laser etching creates an area of weakness in the implant neck and should therefore be eschewed in this part of the femoral stem.

Level of evidence

IV, retrospective study.

El texto completo de este artículo está disponible en PDF.

Keywords : Stem neck fracture, Laser etching, Total hip arthroplasty, Corailtm stem


Esquema


© 2015  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 102 - N° 1

P. 71-76 - février 2016 Regresar al número
Artículo precedente Artículo precedente
  • Hidden blood loss in anterior lumbar interbody fusion (ALIF) surgery
  • H. Ju, R.A. Hart
| Artículo siguiente Artículo siguiente
  • Incidence and characteristics of proximal femoral fractures in an Afro-Caribbean population
  • J.-L. Rouvillain, F. De Cazes, R. Banydeen, K. Rome, P. Numéric, M. De Bandt, C. Derancourt

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.