Is determination of the hip rotation center during computer-assisted surgery influenced by the pinless femoral sensor attachment method? An in vitro preliminary study comparing osseous pin fixation - 25/04/16

Doi : 10.1016/j.otsr.2016.01.014 
E. de Thomasson a, , S. Gouzy b, F. Cloerec b, L. Geais b
a Institut Mutualiste Montsouris, 42, boulevard Jourdan, 75014 Paris, France 
b Société Amplitude, 11, cours Jacques-Offenbach, 26000 Valence, France 

Corresponding author. Tel.: +33 1 56 61 64 09; fax: +33 1 56 61 64 38.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Introduction

Computer-assisted surgery improves the positioning of hip prostheses but requires use of transosseous pins requiring a complementary approach exposing the patient to rare but at times serious complications. The use of sensor arrays attached to the skin could advantageously replace pins provided that comparable results are obtained, but their validity has not yet been assessed. We conducted a prospective in vitro study to: measure the possible error of a cutaneous versus transosseous fixation to determine the hip rotation center (HRC) position and determine the inter- and intraobserver reproducibility of the cutaneous versus the transosseous fixation.

Hypothesis

Use of cutaneous sensor arrays while recording the HRC is sufficiently reliable for its calculation algorithm to provide measurement accuracy within 5mm.

Materials and methods

A rigid array attached with either a silicone strap or an adhesive were compared to a transosseous array. Four series of 96 HRC measurements were collected by four operators on two cadavers, half with an array attached with a strap and half with an adhesive. The results were compared to those obtained by a sensor attached with transosseous pins.

Results

On condition that the hip-knee is mobilized in extension, a sensor array attached with an adhesive gives results with comparable accuracy (standard deviation [SD]: 2.89mm [1.9–4.8]) to the results obtained with a transosseous fixation (SD: 1.2mm [0.9–1.6]), with no significant inter- or intraobserver variation (0.97<ICC<0.99). If the knee was flexed, accuracy was within more than 1cm, or twice the predetermined limit. Only the adhesive array gave measurements systematically within the limits established with the data spread (8.7%) close to that obtained with a pin (6%).

Discussion

An adhesive array seems to be a reliable alternative to transosseous pins to evaluate HRC and provides a three-dimensional landmark, indispensable to total hip arthroplasty navigation. These preliminary results confirm our hypothesis but should be validated in vivo.

Level of evidence

Level III, comparative prospective in vitro study.

El texto completo de este artículo está disponible en PDF.

Keywords : THA, Pinless, Navigation, Hip rotation center


Esquema


© 2016  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 102 - N° 3

P. 375-378 - mai 2016 Regresar al número
Artículo precedente Artículo precedente
  • Sequential therapy with “vacuum sealing drainage-artificial dermis implantation-thin partial thickness skin grafting” for deep and infected wound surfaces in children
  • X.-g. Yuan, X. Zhang, Y.-x. Fu, X.-f. Tian, Y. Liu, J. Xiao, T.-w. Li, L. Qiu
| Artículo siguiente Artículo siguiente
  • No effect of femoral offset on bone implant micromotion in an experimental model
  • F. Amirouche, G. Solitro, A. Walia

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.