Suscribirse

Influence of coronal bowing on the lower alignment and the positioning of component in navigation and conventional total knee arthroplasty - 30/03/17

Doi : 10.1016/j.otsr.2016.11.017 
H. Kobayashi , Y. Akamatsu, K. Kumagai, Y. Kusayama, M. Aratake, T. Saito
 Department of orthopaedic surgery, Yokohama city university, school of medicine, 3-9, Fukuura, Kanazawa, Yokohama, Kanagawa 236-0004, Japan 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Background

Coronal alignment is an important factor for the function and longevity of total knee arthroplasty (TKA). Coronal bowing of the lower extremity is common among Asians and it may pose a risk for malalignment of the lower leg and malposition of component.

Hypothesis

We hypothesized that coronal bowing itself has a risk for malalignment of the lower leg and malposition of femoral/tibial components and that navigation TKA is beneficial for patients with coronal bowing. We investigated the incidence of femoral/tibial bowing in patients treated with TKA and compared the radiographic parameters between the navigation group and the conventional group. Additionally, the influence of coronal bowing on these radiographic parameters was investigated.

Materials and methods

We enrolled 35 patients with knee osteoarthritis and 70 bilateral simultaneous TKAs. The patients underwent TKA with the use of a computer tomography-free navigation in one knee and conventional TKA in the contralateral knee. Preoperative coronal bowing were measured, and the subjects were divided into 2 subgroups, i.e. the bowing group and the non-bowing group. Lateral bowing was expressed as plus (+) and medial bowing was expressed as minus (−). Various radiographic parameters, including coronal bowing, lower leg alignment, component position, and outliers were compared between the navigation group and the conventional group.

Results

Femoral bowing varied from −7.4° to 10.9° with an average of 3.0°. Tibial bowing varied from −4.1° to 4.6° with an average of 0.4°. The femoral component was placed more properly in the navigation group. Number of outlier regarding to the coronal femoral component angle to the femoral mechanical axis was 14 cases (37.8%) in the bowing group and 6 cases (18.2%) in the non-bowing group (P=0.04).

Discussion

In conclusion, coronal femoral bowing has an important effect on femoral bone cut in TKA. The navigated TKA was more consistent than conventional TKA in aiding proper alignments of femoral component.

Level of evidence

Level II, comparative prospective study.

El texto completo de este artículo está disponible en PDF.

Keywords : Total knee arthroplasty (TKA), Coronal bowing, Navigation, Alignment


Esquema


© 2017  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 103 - N° 2

P. 251-256 - avril 2017 Regresar al número
Artículo precedente Artículo precedente
  • Can three-dimensional patient-specific cutting guides be used to achieve optimal correction for high tibial osteotomy? Pilot study
  • M. Munier, M. Donnez, M. Ollivier, X. Flecher, P. Chabrand, J.-N. Argenson, S. Parratte
| Artículo siguiente Artículo siguiente
  • Periprosthetic fractures of the distal femur after total knee arthroplasty : Plate versus nail fixation
  • S.S. Wallace, D. Bechtold, A. Sassoon

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.