Subacromial temperature profile during bipolar radiofrequency use in shoulder arthroscopy. Comparison of Coblation® vs. VAPR® - 29/11/17

Doi : 10.1016/j.otsr.2017.02.011 
V. Huynh , O. Barbier, X. Bajard, A. Bouchard, D. Ollat, G. Versier
 Service de chirurgie orthopédique et traumatologique, hôpital d’instruction des armées Bégin, 69, avenue de Paris, 94160 Saint-Mandé, France 

Corresponding author.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Background

The use of bipolar electrodes for arthroscopic procedures carries a theoretical ex vivo risk of inducing burn injuries. Few studies have measured the in vivo temperatures produced by bipolar electrodes during arthroscopy, and their results are conflicting. The objective of this study was to evaluate the temperature profile within the subacromial space during shoulder arthroscopy with two different electrode systems.

Hypothesis

The primary hypothesis was that the two electrode systems produced similar temperature variations and peak temperatures. The secondary hypothesis was that neither electrode system produced irrigation-fluid temperatures above the tissue-damage threshold.

Material and methods

A comparative, prospective, single-centre, single-surgeon, single-blind study was conducted to compare the Coblation® system (Smith&Nephew, Andover, MA, USA) and the VAPR® system (DePuy Synthes Mitek Sports Medicine, Raynham, MA, USA) in 13 patients undergoing shoulder arthroscopy. A temperature probe inserted into the subacromial space was used to record temperatures at 10-second intervals for 60seconds during continuous radiofrequency application.

Results

Mean baseline temperature was 21.4±0.7°C with VAPR® and 23.0±2.2°C with Coblation®. No significant between-group differences were found during the first 40seconds. The mean peak temperature reached after 60seconds was 25.0±1.9°C with VAPR® and 27.9±2.8°C with Coblation® (P<0.05).

Discussion

Few studies have compared the in vivo temperatures produced during arthroscopy by different electrode systems. In vivo studies have established that temperature increases can cause tissue damage, particularly to chondrocytes, and that the irrigation flow rate plays a key role in lowering the in vivo temperatures. Our study showed a significant difference between the two electrode systems after 50seconds of use, with lower temperatures with the VAPR®. Nevertheless, neither system increased the irrigation-fluid temperatures above the tissue-damage threshold. Both systems can be used safely, provided the manufacturer instructions are followed and the irrigation system is effective.

Level of evidence

II (prospective randomized trial).

El texto completo de este artículo está disponible en PDF.

Keywords : Temperature, Coblation, VAPR, Shoulder, Arthroscopy


Esquema


© 2017  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 103 - N° 4

P. 489-491 - juin 2017 Regresar al número
Artículo precedente Artículo precedente
  • Learning the skills needed to perform shoulder arthroscopy by simulation
  • P.-S. Marcheix, G. Vergnenegre, F. Dalmay, C. Mabit, J.-L. Charissoux
| Artículo siguiente Artículo siguiente
  • Changes of the bone structure after cap resurfacing arthroplasty of the humeral head
  • L.V. von Engelhardt, M. Hahn, T. Schulz, K. Peikenkamp, J. Jerosch

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.