Lower-limb lengths and angles in children older than six years: Reliability and reference values by EOS® stereoradiography - 24/04/18

Doi : 10.1016/j.otsr.2017.10.007 
V. Rampal a, b, , P.-Y. Rohan a, A. Assi c, I. Ghanem c, O. Rosello b, A.-L. Simon d, E. Gaumetou d, V. Merzoug e, f, W. Skalli a, P. Wicart g
a Institut de biomécanique humaine Georges-Charpak, arts et métiers ParisTech, 151, boulevard de l’Hôpital, 75013 Paris, France 
b Service d’orthopédie pédiatrique, hôpitaux pédiatriques de Nice, CHU Lenval, 57, avenue de la Californie, 06000 Nice, France 
c Laboratoire de biomécanique, faculté de médecine, université de Saint-Joseph, Beyrouth, Lebanon 
d Service d’orthopédie pédiatrique, hôpital Robert-Debré, 48, boulevard Serurier, 75019 Paris, France 
e Service de radiologie, hôpital Cochin, 27, rue du Faubourg-Saint-Jacques, 75014 Paris, France 
f Hôpital Bicètre, 78, rue du Général-Leclerc, 94270 Le-Kremlin-Bicêtre, France 
g Service d’orthopédie pédiatrique, hôpital Necker–Enfants-Malades, 149, rue de Sèvres, 75015 Paris, France 

Corresponding author. Institut de biomécanique humaine Georges-Charpak, arts et métiers ParisTech, 151, boulevard de l’Hôpital, 75013 Paris, France.Institut de biomécanique humaine Georges-Charpak, arts et métiers ParisTech, 151, boulevard de l’Hôpital, 75013 Paris, France.

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.
Artículo gratuito.

Conéctese para beneficiarse!

Abstract

Background

Lower-limb alignment in children is classically assessed clinically or based on conventional radiography, which is associated with projection bias. Low-dose biplanar radiography was described recently as an alternative to conventional imaging. The primary objective of this study was to assess the reliability of length and angle values inferred from 3D reconstructions in children seen in everyday practice. The secondary objective was to obtain reference values for goniometry parameters in children.

Hypothesis

3D reconstructions can be used to assess the lower limbs in children.

Material and methods

The paediatric reliability study was done in 18 volunteers who were divided into three groups based on whether they were typically developing (TD) children, had skeletal development abnormalities, or had cerebral palsy. The reference data were obtained in 129 TD children. Each study participant underwent biplanar radiography with 3D reconstruction performed by experts and radiology technicians. Goniometry parameters were computed automatically. Reproducibility was assessed based on the intra-class coefficient (ICC) and the ISO 5725 standard (standard deviation of reproducibility, SDR).

Results

For length parameters, the ICCs ranged from 0.94 to 1.00 and the SDR from 2.1 to 3.5mm. For angle parameters, the ICC and SDR ranges were 0.60–0.95 and 0.9°–4.6°, respectively. No significant differences were found across experts or radiology technicians. Age-specific reference data are reported.

Discussion

These findings confirm the reliability of low-dose biplanar radiography for assessing lower-limb parameters in children seen in clinical practice. In addition, the study provides reference data for commonly measured parameters.

Level of evidence

IV.

El texto completo de este artículo está disponible en PDF.

Keywords : Lower limb, Low-dose biplanar X-ray imaging, Paediatrics


Esquema


© 2017  Elsevier Masson SAS. Reservados todos los derechos.
Añadir a mi biblioteca Eliminar de mi biblioteca Imprimir
Exportación

    Exportación citas

  • Fichero

  • Contenido

Vol 104 - N° 3

P. 389-395 - mai 2018 Regresar al número
Artículo precedente Artículo precedente
  • Anterior and lateral overcoverage after triple pelvic osteotomy in childhood for developmental dislocation of the hip with acetabular dysplasia: Frequency, features, and medium-term clinical impact
  • C. Klein, A. Fontanarosa, N. Khouri, J. Bellity, J.-P. Padovani, C. Glorion, P. Wicart
| Artículo siguiente Artículo siguiente
  • Primary pyomyositis in children
  • M.M. Elzohairy

Bienvenido a EM-consulte, la referencia de los profesionales de la salud.

Mi cuenta


Declaración CNIL

EM-CONSULTE.COM se declara a la CNIL, la declaración N º 1286925.

En virtud de la Ley N º 78-17 del 6 de enero de 1978, relativa a las computadoras, archivos y libertades, usted tiene el derecho de oposición (art.26 de la ley), el acceso (art.34 a 38 Ley), y correcta (artículo 36 de la ley) los datos que le conciernen. Por lo tanto, usted puede pedir que se corrija, complementado, clarificado, actualizado o suprimido información sobre usted que son inexactos, incompletos, engañosos, obsoletos o cuya recogida o de conservación o uso está prohibido.
La información personal sobre los visitantes de nuestro sitio, incluyendo su identidad, son confidenciales.
El jefe del sitio en el honor se compromete a respetar la confidencialidad de los requisitos legales aplicables en Francia y no de revelar dicha información a terceros.


Todo el contenido en este sitio: Copyright © 2024 Elsevier, sus licenciantes y colaboradores. Se reservan todos los derechos, incluidos los de minería de texto y datos, entrenamiento de IA y tecnologías similares. Para todo el contenido de acceso abierto, se aplican los términos de licencia de Creative Commons.